  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64159",
      "recalling_firm": "Synthes USA HQ, Inc.",
      "address_1": "1302 Wrights Ln E",
      "address_2": "N/A",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA Nationwide Distribution including the states of  AL, AR, AK, CA, FL,  ID, KS, LA, MD, MA, MO, NJ, NY, OH, PA, TN, TX, and VA.",
      "recall_number": "Z-0941-2013",
      "product_description": "Large Combination Clamp/Large External Fixation System, Part Number 390.005    Product Usage:  Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation.",
      "product_quantity": "136",
      "reason_for_recall": "The L1 hex nuts, used to tighten the Large Combination Clamp, were oversized and the 11 mm sockets and wrenches will not fit on the hex nuts.",
      "recall_initiation_date": "20130111",
      "center_classification_date": "20130311",
      "termination_date": "20150831",
      "report_date": "20130320",
      "code_info": "Lot Numbers 6602070, 6582333, 6581191, and 6581193",
      "more_code_info": ""
    }
  ]
}