      "country": "United States",
      "city": "Philadelphia",
      "address_1": "1015 Chestnut St",
      "reason_for_recall": "Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated.",
      "address_2": "",
      "product_quantity": "147 systems",
      "code_info": "Device Reference/Catalog Number (Black) - 1000003, (white) - 1000018. Software version 6.4.3    EUA - Emergency Use Authorization",
      "center_classification_date": "20210131",
      "distribution_pattern": "Domestic: FL, TX, IN, KY, OH, NY, CA, PA, MD.   Foreign: Canada, Zimbabwe.",
      "state": "PA",
      "product_description": "Nucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or diagnosis of 2019-nCoV infection in human clinical specimens. The RT-PCR test runs on a Franklin Real-Time PCR Thermocycler Device (Reference/Catalog Number 1000003,1000018) with the Software Controller version 6.4.3. The cycler has a \"baseboard\" unit with firmware and associated software component that control motorized movements of the optical filter carriage used for PCR detection.",
      "report_date": "20210210",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Biomeme, Inc.",
      "recall_number": "Z-0940-2021",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "87153",
      "termination_date": "20210611",
      "more_code_info": "",
      "recall_initiation_date": "20210105",
      "postal_code": "19107-4316",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}