{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Reading",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81375",
      "recalling_firm": "Arrow International Inc",
      "address_1": "2400 Bernville Rd",
      "address_2": "",
      "postal_code": "19605-9607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution.",
      "recall_number": "Z-0940-2019",
      "product_description": "AF w/AS 550 x 1-14ML/HR CPNB, Product code MVBXL- CPNB, Arrow AutoFuser Disposable Pain Control Pump",
      "product_quantity": "38",
      "reason_for_recall": "Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay  while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug  delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This  may result in the need for other, additional analgesic drug administration.",
      "recall_initiation_date": "20181002",
      "center_classification_date": "20190225",
      "termination_date": "20211112",
      "report_date": "20190306",
      "code_info": "74A1801376 74G1702799  74A1801900 74H1700167  74A1802544 74H1701111  74B1700925 74H1701796  74C1701765 74H1702406  74D1700392 74J1700193  74D1701253 74J1700680  74D1701539 74J1701583  74D1702241 74J1702180  74E1700485 74J1702742  74E1701026 74K1700635  74E1701692 74K1701435  74E1702410 74K1702040  74F1700147 74K1702604  74F1701966 74L1700203  74F1702681 74L1700932  74F1703240 74L1702112  74G1700539 74M1700833  74G1701226 74M1701615",
      "more_code_info": ""
    }
  ]
}