  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Indianapolis",
      "address_1": "7301 Georgetown Rd.",
      "reason_for_recall": "Received 2 complaints out of 458 distributed pumps in this lot, where flow restrictor bead became displaced, which were reported to permit fast flow of contents.",
      "address_2": "Suite 150",
      "product_quantity": "458",
      "code_info": "Kit Part Number 510545-BP, lot 12-100983, 12-100984, 12-101096, 12-101125   Kit Part Number 510112-BP, lot 12-101110, 12-101103   Kit:  510401-BP lot 12-101045   Kit: 510644, lot 12-101028, 12-101097, 12-101130 GOPump Dual Infusion Pump Kit    All Kits contain Pump Part Number: 510076 Lot Number: 12-100601",
      "center_classification_date": "20130313",
      "distribution_pattern": "Nationwide Distribution including PA, CA, OH, AL, FL, WV, TN, NJ, NC, TX, CO, NY, TX, MO, MS, SC, and GA.",
      "state": "IN",
      "product_description": "GOPump Elastometric infusion PumpKit with/ and without accesssories all containing Dual GOPump, part number 510076: One infusion pump comprising two,  150 mL elastomeric reservoirs, each of which connects to its own catheter. Liquid flow rate is labeled at 2 mL/hr per elastomer.    The Symbios Disposable Infusion Pump Kit is a disposable, self-contained infusion system utilizing an inflatable elastomeric reservoir to mechanically provide percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management",
      "report_date": "20130320",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Symbios Medical Products, LLC",
      "recall_number": "Z-0940-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64423",
      "termination_date": "20140213",
      "more_code_info": "",
      "recall_initiation_date": "20130225",
      "postal_code": "46268",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}