{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hartland",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81869",
      "recalling_firm": "Merge Healthcare, Inc.",
      "address_1": "900 Walnut Ridge Dr",
      "address_2": "",
      "postal_code": "53029-8347",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution nationwide to Alabama, Alaska, Arizona, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, , Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West , Virginia, Wisconsin.      International distribution to Algeria, Austria, Belgium, Canada, Czech Republic, Denmark, Germany, Israel, Netherlands, Puerto Rico, United Kingdom.",
      "recall_number": "Z-0939-2019",
      "product_description": "The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digital Imaging System    Product Usage:  The WinStation Retinal Imager has the same intended use as other fundus cameras. It is specifically used to produce color or black and white images of the retina (fundus imaging) and anterior segement (slit lamp imaging), and fluorescein and ICG angiographic images in a non-invasive manner.",
      "product_quantity": "1,252 Units",
      "reason_for_recall": "Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).",
      "recall_initiation_date": "20181220",
      "center_classification_date": "20190225",
      "termination_date": "20230228",
      "report_date": "20190306",
      "code_info": "UDI: 00842000100140. Serial Numbers: a0L3200000BVmMo, a0L3200000DyTSA, a0L3200000AiuLX, a0L3200000BUou2, a0L0e00000EPuiz, a0L3200000BVZ7i, a0L3200000Ajd0n, a0L3200000C04Ol, a0L3200000Aj6EX, a0L32000009gkeu, a0L3200000BUOXv, a0L3200000Aj6TP, a0L3200000BTnqr, a0L3200000AjTBg, a0L32000008q1mk, a0L0e00000DzQ3a, a0L3200000B2Csy, a0L32000009hJZk, a0L32000009n0mi, a0L0e00000E0aFN, a0L3200000DyFdA, a0L0e00000EPw0a, a0L3200000C03vm, a0L3200000DyF4P, a0L3200000AjY5P, a0L32000009n1Mk, a0L3200000B1MV5, a0L0e00000EPw6s, a0L32000008qtIF, a0L3200000BVzZ6, a0L3200000BTLiZ, a0L3200000Aj6bv, a0L0e00000E0C04, a0L3200000AihSR, a0L3200000AihUh, a0L3200000BUc9x, a0L3200000BUcAH, a0L32000009hLmm, a0L3200000AivBM, a0L3200000AjYPs, a0L32000009h8o4, a0L0e00000E07hv, a0L32000008quuk, a0L3200000DyNRS, a0L3200000B19Jt, a0L3200000B0uso, a0L0e00000E0CSc, a0L3200000B09PT, a0L3200000B09PJ, a0L3200000B09Pd, a0L3200000BTlfb, a0L3200000DiqQx, a0L32000009hJrd, a0L32000009mlt9, a0L32000009md53, a0L32000009nTWm, a0L0e00000EPxQj, a0L3200000BWKuB, a0L3200000B1N0q, a0L3200000Dyg4I, a0L0e00000Dzwru, a0L3200000BzVK2, a0L3200000AjnnW, a0L3200000DySwx, a0L3200000BgdOF, a0L32000009h6WM, a0L32000009h6WR, a0L3200000B0CXr, a0L3200000Aitxy, a0L32000009hJ6P, a0L3200000DyGJu, a0L60000007LRa6, a0L3200000DyFXC, a0L60000007OcXf, a0L3200000AjkWQ, a0L3200000Di2CQ, a0L3200000BWZGK, a0L0e00000Dkz6R, a0L3200000AjbF3, a0L3200000DyGL3, a0L3200000DyGKd, a0L3200000DyGKs, a0L3200000BTYBz, a0L3200000AjSo7, a0L3200000AjmbA, a0L0e00000DxTju, a0L32000009eaXj, a0L3200000AivsT, a0L32000009hA2T, a0L32000009hZDo, a0L3200000B0vwx, a0L3200000DiGem, a0L32000008qVOb, a0L3200000AjlpT, a0L0e00000DzgWg, a0L32000009mzAO, a0L3200000DiF29, a0L3200000DiF1z, a0L3200000AjYmf, a0L3200000DiSYx, a0L0e00000E0TgS, a0L3200000BzJCA, a0L3200000AjbIq, a0L3200000B0A6r, a0L3200000Aiiis, a0L0e00000DxSED, a0L0e00000EPw0k, a0L3200000DyGKE, a0L3200000DyGKT, a0L3200000DyGKY, a0L0e00000DzNNe, a0L32000009h8mS, a0L3200000B0Un4, a0L3200000B0Un9, a0L3200000B0Umz, a0L3200000B0UnE, a0L0e00000EPvHa, a0L32000009mmxx, a0L3200000AjVHx, a0L3200000DyFYj, a0L0e00000E0CWU, a0L3200000Diqej, a0L3200000AjVby, a0L32000009mpFG, a0L32000009mqPj, a0L3200000BWZRT, a0L0e00000EPtf1, a0L32000009h5NQ, a0L0e00000DkuPo, a0L3200000AifYF, a0L0e00000EPwfs, a0L3200000AjTri, a0L0e00000EPvs7, a0L32000009hJJp, a0L0e00000DzNNy, a0L3200000DhRjV, a0L0e00000DzXBa, a0L3200000BzIUD, a0L3200000AihtL, a0L32000009hLHD, a0L3200000DycpT, a0L3200000Dyd75, a0L60000007OvfM, a0L32000009hXaG, a0L3200000B1z3N, a0L3200000Ajq8e, a0L3200000Bzsim, a0L3200000Bzsjf, a0L32000009g59e, a0L3200000BToIj, a0L32000009nGPH, a0L3200000DiGFZ, a0L0e00000EPqr6, a0L3200000Bg6tH, a0L0e00000DzNPz, a0L3200000AjMEo, a0L3200000BV8Dk, a0L3200000AjMbE, a0L0e00000DzQdh, a0L3200000AijMO, a0L3200000B1ZXY, a0L0e00000EPwCg, a0L3200000AjPKO, a0L32000009h6Sx, a0L3200000B0CEa, a0L3200000B0CF9, a0L3200000BgJfJ, a0L3200000Ajp2G, a0L32000009nGVt, a0L3200000B0S7X, a0L3200000B0Cwo, a0L0e00000EPtfV, a0L0e00000DzV9w, a0L0e00000DzgiM, a0L32000009hKk6, a0L0e00000E0lM5, a0L3200000DhOvR, a0L32000009hVqm, a0L3200000Ajfc7, a0L0e00000EPwYN, a0L0e00000DzNSU, a0L3200000BVTjP, a0L3200000B1O4R, a0L32000009nWFW, a0L3200000B2Rmk, a0L0e00000DzVCW, a0L3200000AiiWw, a0L0e00000DqRSC, a0L32000009eHT9, a0L32000009hKk1, a0L32000009nVXX, a0L0e00000E0CMe, a0L3200000Aiwq1, a0L32000009h9dO, a0L32000009nGjy, a0L3200000Di1aa, a0L0e00000Dl0rE, a0L0e00000Dl1Jr, a0L3200000B0hDN, a0L3200000DiqN0, a0L32000009haQu, a0L0e00000DkuPK, a0L3200000B0CZd, a0L32000009hYJt, a0L3200000BWlgS, a0L0e00000E07gx, a0L0e00000EPuG2, a0L0e00000EPuFx, a0L32000009hMxY, a0L3200000B0u8J, a0L3200000BVUTo, a0L32000009hJuN, a0L3200000B0UNR, a0L3200000AjAkv, a0L3200000Ajnb9, a0L3200000BVmPE, a0L3200000DiTgA, a0L3200000BWkbl, a0L0e00000EfMBR, a0L32000009hKc9, a0L0e00000DzO9k, a0L3200000DhguX, a0L3200000Ajp39, a0L32000009h59x, a0L0e00000E0Sd7, a0L3200000AisHf, a0L32000009eHg3, a0L0e00000DzPyQ, a0L0e00000DzPzJ, a0L3200000DigHt, a0L0e00000DzQ2I, a0L3200000AjkZC, a0L0e00000DqRTF, a0L0e00000DqRTe, a0L3200000BVmIC, a0L3200000AjPzb, a0L0e00000DqRTU, a0L32000008qpH3, a0L32000008qpHD, a0L32000008qpGy, a0L0e00000Dzwoq, a0L0e00000E0kP3, a0L3200000BVcwJ, a0L3200000DiRkQ, a0L3200000DiRkG, a0L0e00000EPv0F, a0L0e00000EPv0A, a0L3200000BWNOD, a0L3200000AjPGP, a0L0e00000EfMDc, a0L3200000AjV66, a0L3200000AjPD0, a0L3200000B1KJW, a0L3200000B1KJM, a0L32000009nUI7, a0L0e00000EPrer, a0L3200000BVynO, a0L3200000DyNVK, a0L32000008qV53, a0L0e00000E0ltx, a0L3200000BVmij, a0L3200000BUmKq, a0L32000009hXyD, a0L32000009hXxo, a0L3200000BWkCU, a0L3200000BVVac, a0L3200000B1KJl, a0L3200000BV8DV, a0L3200000B1KJR, a0L3200000BUo9h, a0L0e00000EPt0y, a0L3200000Ajkr2, a0L3200000AjTRX, a0L3200000B0jAs, a0L0e00000DxY42, a0L3200000DhDXm, a0L3200000DhsCj, a0L3200000BWl3a, a0L32000009moWy, a0L3200000BVKpV, a0L3200000BUx34, a0L3200000BUx2z, a0L3200000DyF7d, a0L3200000BUx2a, a0L3200000Bgbdg, a0L3200000BUYqX, a0L3200000BWNQG, a0L3200000Bzsb1, a0L32000009ngLZ, a0L32000009ngPx, a0L32000009ngQl, a0L3200000BUx3E, a0L3200000DhqKP, a0L32000009hY1q, a0L3200000AjMVk, a0L3200000AisQc, a0L32000009hW71, a0L32000009hKtX, a0L3200000BWky0, a0L3200000BWkxg, a0L3200000BWkwa, a0L3200000BWkxq, a0L3200000BWkxv, a0L32000008rXRq, a0L32000009hXIW, a0L0e00000DxSLJ, a0L32000009mdAh, a0L3200000B21Uq, a0L3200000B1nIO, a0L3200000BVXjc, a0L3200000BVXjm, a0L0e00000EPwYr, a0L3200000Di3H1, a0L32000009hMo2, a0L3200000B0UB0, a0L3200000BWBJP, a0L0e00000EPsXc, a0L32000009nFoF, a0L0e00000E0Bpu, a0L3200000DisEg, a0L3200000AjZUl, a0L32000009hZn7, a0L3200000AjegQ, a0L3200000Dhryh, a0L3200000B0vDR, a0L32000009hJPR, a0L32000009h6dj, a0L3200000DhggD, a0L0e00000E0C3r, a0L0e00000DzxLG, a0L3200000Bz6LS, a0L32000009hZh6, a0L0e00000Dkz05, a0L0e00000DzQHF, a0L3200000B0D1f, a0L0e00000DzVCC, a0L3200000B19uW, a0L32000009n0w5, a0L0e00000EPwmF, a0L32000009nDM1, a0L3200000BTOG3, a0L0e00000Dkyte, a0L3200000BTOGD, a0L3200000BgdIO, a0L3200000B1apU, a0L3200000BTOFy, a0L32000009gtEq, a0L3200000BUwZ9, a0L3200000AjehT, a0L0e00000E0mzD, a0L3200000DhDh3, a0L32000009h9mG, a0L3200000AjN8S, a0L32000009hLSL, a0L32000009hLSa, a0L3200000BUlrM, a0L32000009hYik, a0L32000009hZnH, a0L0e00000DzQeV, a0L3200000AixBW, a0L3200000DiGzi, a0L3200000DiGzx, a0L3200000DiH1q, a0L3200000Dighg, a0L3200000Dighb, a0L3200000BVxRI, a0L3200000BVxQt, a0L3200000BVxRS, a0L3200000BVxRN, a0L3200000DiGyQ, a0L3200000DiIEq, a0L3200000DiIBF, a0L3200000DiIAW, a0L3200000DiIEK, a0L3200000DiIFR, a0L3200000DiIFF, a0L3200000DiI8u, a0L3200000DiI8a, a0L3200000DiIE5, a0L3200000DiIB0, a0L3200000DiIEg, a0L3200000BVn2X, a0L3200000Di2do, a0L3200000Di2dZ, a0L3200000Di2eh, a0L3200000Di2dF, a0L3200000Di2ie, a0L3200000Di2j8, a0L0e00000DqRRO, a0L0e00000DqRRJ, a0L3200000DyaP8, a0L3200000DyaPr, a0L3200000DipDP, a0L3200000DipDU, a0L32000009hNAh, a0L3200000AjBEA, a0L3200000B2RoH, a0L3200000BzLg5, a0L3200000BzKLe, a0L3200000DhgNk, a0L3200000DhEbU, a0L0e00000EfLoI, a0L3200000B0tAC, a0L3200000Dpw2q, a0L3200000Dykp5, a0L0e00000DzxL1, a0L0e00000DxV8c, a0L0e00000DzNRl, a0L32000009nGRc, a0L32000009hIx7, a0L3200000BUcNx, a0L32000009hZnM, a0L3200000DyNth, a0L3200000BWBPm, a0L3200000B0SF2, a0L3200000DyPqR, a0L0e00000EfKJE, a0L3200000B2Eu9, a0L0e00000DzO7Z, a0L32000009hIxW, a0L32000009hYBL, a0L0e00000E0Bij, a0L32000009hK2S, a0L32000008pHlh, a0L60000008pCXi, a0L32000008pO3Z, a0L3200000DyIQ0, a0L3200000Bg4iC, a0L0e00000EPu4L, a0L3200000BVTqg, a0L0e00000EPs5i, a0L0e00000DzO8I, a0L3200000Aj9Af, a0L60000008p87v, a0L0e00000DzXFI, a0L3200000BVTsX, a0L0e00000DzXFX, a0L3200000B0vFN, a0L3200000B1kSl, a0L3200000Ajpyn, a0L3200000BUlH5, a0L3200000BV8dg, a0L3200000DyNrB, a0L0e00000Dzboz, a0L3200000Ajgrh, a0L3200000AiuS8, a0L3200000AiuyH, a0L3200000AiwYK, a0L3200000Aj5z5, a0L3200000Aj5zA, a0L3200000Aj61f, a0L3200000Aj61a, a0L3200000Aji1r, a0L0e00000DzQRd, a0L3200000B1KIY, a0L3200000BVTsh, a0L0e00000E0CCo,   a0L0e00000DzQGH, a0L3200000Aj8aE, a0L3200000AjA3K, a0L3200000AjBG6, a0L3200000DzDpv, a0L0e00000DzNCl, a0L0e00000DzQO0, a0L3200000AjozO, a0L0e00000EPtrW, a0L3200000DykPy, a0L0e00000EPpTN, a0L0e00000EPpi8, a0L0e00000EPtrM, a0L0e00000E0Bpp, a0L3200000Aj8Qg, a0L3200000B1KIn, a0L0e00000E0mlz, a0L0e00000E0CXD, a0L3200000AjoCZ, a0L3200000BVmV0, a0L3200000BWlDd, a0L0e00000DzOB2, a0L3200000Dy3n9, a0L0e00000DxPD2, a0L0e00000DzXEF, a0L3200000DyVe9, a0L32000009hJNF, a0L0e00000DqRbP, a0L0e00000E0hr9, a0L3200000B186W, a0L3200000BWlCT, a0L3200000BVmTO, a0L32000009hJ2E, a0L32000009hYSp, a0L0e00000EPuII, a0L3200000Bgb9h, a0L3200000DpwMM, a0L3200000BziPr, a0L0e00000DzQ5z, a0L0e00000EPtgs, a0L0e00000EPtgn, a0L3200000BTZVI, a0L0e00000DzNNA, a0L0e00000DkuTX, a0L3200000AjegV, a0L0e00000Dzgk3, a0L3200000B1KIE, a0L3200000B1KI9, a0L3200000B1KIJ, a0L0e00000DqRZs, a0L32000009gkit, a0L3200000Dj3TN, a0L3200000DpkOH, a0L3200000C03yW, a0L32000009mmzZ, a0L32000009nUtq, a0L0e00000Dl5zm, a0L3200000AjSK9, a0L3200000AjSM7, a0L0e00000DzQZ7, a0L0e00000EPvdq, a0L32000009mq3U, a0L0e00000E0CAx, a0L0e00000EPrKD, a0L0e00000EPrK8, a0L32000009hIvw, a0L3200000BWYT3, a0L0e00000DzQfd, a0L32000008peLC, a0L3200000Aj624, a0L0e00000EPpSt, a0L0e00000DxPtS, a0L0e00000Dl5zr, a0L3200000B0wf2, a0L0e00000DzQ3f, a0L3200000AjhJF, a0L0e00000DzcfG, a0L0e00000EPsXh, a0L3200000Bz7pS, a0L3200000Ajkmq, a0L3200000AjXH2, a0L0e00000E0BcB, a0L0e00000EPrx5,   a0L3200000B0woQ, a0L0e00000EPqIL, a0L0e00000EPqMN, a0L32000009n2qT, a0L3200000B0f1C, a0L3200000B0f1b, a0L3200000B0S8a, a0L3200000B0T7a, a0L3200000BWl28, a0L0e00000EPu3r, a0L0e00000DzbyW, a0L3200000B1Zcv, a0L32000009hIf1,   a0L3200000Dyke1, a0L3200000DhDCk, a0L3200000AjeiM, a0L3200000AjTtA, a0L0e00000DqRbK, a0L0e00000Dl5zh, a0L0e00000DzQ3x, a0L0e00000E0C0d, a0L32000009hIfi, a0L3200000B0tqM, a0L3200000B18Ce, a0L3200000AifGK, a0L3200000AifGF, a0L32000009h8kM, a0L3200000DhFZ4, a0L3200000BTOG8, a0L3200000B2Fgn, a0L0e00000DxS0B, a0L32000009ekM1, a0L0e00000DxPfB, a0L3200000B19N2, a0L3200000DyP9F, a0L0e00000E0btG, a0L3200000Aihc2, a0L32000009hY8b, a0L0e00000E0h97, a0L3200000AjYzA, a0L3200000BVSms, a0L32000009nTPM, a0L32000009n1Qw, a0L32000009nExk, a0L3200000BVYoq, a0L32000009nTzg, a0L3200000BVc7t, a0L3200000BVc8I, a0L3200000BVc8S, a0L3200000BVc5M, a0L3200000BVc7y, a0L3200000BVc83, a0L3200000DhCjz, a0L3200000BUNbm, a0L3200000BUNe7, a0L3200000BUNd4, a0L3200000BUNbI, a0L3200000BVYzj, a0L32000009hMk1, a0L32000009hZSD, a0L3200000AifcR, a0L3200000BTnsU, a0L3200000BTnso, a0L3200000DyFko, a0L3200000DyFmG, a0L0e00000DxPlY, a0L32000008pd76, a0L3200000B16Cm, a0L3200000BUxUl, a0L3200000BzUMv, a0L32000009hJ6o, a0L3200000AjASC, a0L0e00000DzNFu, a0L3200000AiggP, a0L3200000Aiipi, a0L32000009hNOH, a0L3200000Aige9, a0L3200000DhrGg, a0L3200000Dhbt5, a0L3200000AjAJn, a0L3200000AjAKR, a0L3200000AjAK2, a0L3200000Aigda, a0L3200000Aifnw, a0L3200000Aiip4, a0L3200000DpljY, a0L3200000B1zaL, a0L3200000B16MP, a0L3200000DyGJz, a0L3200000DisvM, a0L3200000Aiwl6, a0L32000009ha9q, a0L32000009nE1z, a0L3200000BUl3E, a0L32000009hZ7R, a0L3200000AjVLG, a0L3200000BUm77, a0L32000009nEr6, a0L0e00000EPxLF, a0L3200000BVahZ, a0L32000009hX70, a0L32000009hJGH, a0L3200000Bgb9r, a0L32000008pHtq, a0L3200000B0veu, a0L0e00000E0CWK, a0L3200000BUm7M, a0L3200000DidmF, a0L3200000Didll, a0L3200000Didnc, a0L3200000BVW8d, a0L3200000BVW97, a0L0e00000EPrRY, a0L3200000B0OJX, a0L0e00000EPrFh, a0L0e00000DzgfI, a0L0e00000EfL3R, a0L0e00000EfL3W, a0L0e00000EPwre, a0L3200000BWaHH, a0L3200000Ajkq4, a0L32000009hMSr, a0L3200000AiijM, a0L3200000B0tTZ, a0L32000008qpGo, a0L32000008qpGt, a0L32000008qaaX, a0L32000008pZJi, a0L3200000AjkaK, a0L0e00000EPx2S, a0L3200000BVmNb, a0L3200000BVKme, a0L0e00000EPx2c, a0L0e00000DkuT3, a0L3200000Aigf2, a0L32000009haOZ, a0L32000009mp2z, a0L0e00000DxQEa, a0L0e00000E0lKY, a0L0e00000E0lKE, a0L3200000B1a5Z, a0L3200000BVbN4, a0L3200000AjhGF, a0L3200000AiiLy, a0L0e00000DzQ2h, a0L3200000Ajh6K, a0L3200000AisNY, a0L3200000Bgbi2, a0L32000009hVxe, a0L32000009hVy3, a0L32000009hVyS, a0L3200000AiimE, a0L32000009gwY2, a0L0e00000EPqDG, a0L3200000BVXWN, a0L3200000BVXWI, a0L3200000B1nQd, a0L3200000BzW13, a0L3200000B2CsP, a0L3200000BTyHe, a0L0e00000EPwIj, a0L3200000AjhCX, a0L3200000BTbQq, a0L3200000Aig8T, a0L3200000Aiw33, a0L3200000DynjL, a0L3200000AjTxC, a0L3200000Aivsn, a0L32000009hJOY, a0L3200000BWYIG, a0L3200000B22dy, a0L3200000BTybd, a0L3200000DhDen, a0L3200000AiiA3, a0L0e00000E0kSM, a0L0e00000DqRT0, a0L3200000AjnIa, a0L0e00000EfMTG, a0L3200000BTmsS, a0L3200000B0h3V, a0L32000009hLiy, a0L0e00000Dkz4u, a0L0e00000Dkyyn, a0L3200000BgLqL, a0L3200000AjjTm, a0L3200000AjS6i, a0L3200000AjOLD, a0L3200000BgJkx, a0L0e00000EfKmz, a0L3200000DyIxV, a0L3200000AjPaG, a0L3200000AjmXw, a0L3200000B1kbk, a0L3200000B0B6s, a0L3200000BUOzK, a0L3200000BUm48, a0L32000008qpGP, a0L32000009h60O, a0L3200000B0i5q, a0L3200000BWN3n, a0L3200000BWN5B, a0L3200000AjLAG, a0L3200000DyU4c, a0L0e00000Dl1Cl, a0L0e00000E0BxK, a0L32000009hXeU, a0L3200000Aj629, a0L3200000DzIIq, a0L3200000DzIIl, a0L32000009hXeK, a0L32000009n1pe, a0L0e00000DzQep, a0L3200000BUcIY, a0L3200000DiseL, a0L0e00000DzwyH, a0L32000009hX7A, a0L3200000BVWmP, a0L0e00000DzVCv, a0L3200000Dynno, a0L3200000DiqbW, a0L32000009eZco, a0L3200000DyItp, a0L3200000Aj6oe, a0L3200000AjM8N, a0L0e00000Dl1RC, a0L3200000B0f6G, a0L0e00000EPqwa, a0L32000009hVu6, a0L32000008qpGj, a0L3200000B09OB, a0L3200000AjnNY, a0L3200000B20ld, a0L3200000DyHC3, a0L3200000AiwN7, a0L3200000BWmpV, a0L3200000B0w8C, a0L3200000B0Sr6, a0L3200000B1KiP, a0L32000009hapG, a0L0e00000EfKhB, a0L3200000DyOJH, a0L3200000AjqBE, a0L0e00000DzgkI, a0L3200000BV6eb, a0L3200000BV8cJ, a0L3200000AiiYx, a0L3200000AjhWA, a0L3200000Aj87N, a0L3200000BgdHL, a0L3200000BWa9s, a0L3200000BgM7L, a0L3200000BgbBT, a0L3200000Bg6XB, a0L0e00000DqRU8, a0L0e00000DzQEz, a0L3200000AjpGS, a0L0e00000EPu4z, a0L3200000B1Y8Q, a0L3200000Ajk9l, a0L3200000Di3cq, a0L3200000AjeEV, a0L3200000DyJIr, a0L3200000BVJwX, a0L32000009hIst, a0L0e00000Dl0cx, a0L3200000BVHcC, a0L32000009hYJj, a0L3200000Ajm62, a0L32000008qtHv, a0L3200000Di4RX, a0L32000009hKvB, a0L3200000B0Ayn, a0L3200000Aj914, a0L3200000AjZVc, a0L0e00000DxZBY, a0L3200000Di3Fo, a0L3200000B218m, a0L3200000BWBlr, a0L3200000DySF8, a0L32000009hImk, a0L3200000BVmUW, a0L3200000Dyi3I, a0L32000009hA37, a0L3200000BzLRT, a0L3200000C02sQ, a0L32000009n2tX, a0L3200000DyPps, a0L3200000AiwEu, a0L0e00000EPsWo, a0L32000009mchG, a0L0e00000EPphA, a0L0e00000Dl1I5, a0L3200000AigoM, a0L3200000BVU1o, a0L0e00000Dkz6M, a0L3200000BVZuY, a0L32000009hNLb, a0L32000009n0iW, a0L3200000BUxSF, a0L3200000BV7Wv, a0L32000008qvEm, a0L3200000B2GbM, a0L3200000AjZJo, a0L0e00000EfJmj, a0L3200000B16Bo, a0L3200000DzHbs, a0L0e00000DzPwy, a0L0e00000DzPwj, a0L0e00000DzPwt, a0L3200000Ajjba, a0L0e00000DxR4b, a0L3200000AjdH9, a0L0e00000DzgiW, a0L3200000AjMQ6, a0L3200000AjMR4, a0L3200000BTluh, a0L3200000BTNp3, a0L3200000B20Wy, a0L3200000Dykyy, a0L3200000BVa5h, a0L3200000BWlT8, a0L3200000AjYa4, a0L3200000Ajaxs, a0L3200000AigET, a0L0e00000Dl0Bc, a0L3200000AjkPm, a0L0e00000Dl0Dn, a0L3200000DirIM, a0L0e00000EPu3X, a0L32000009hYBQ, a0L3200000B0ihC, a0L3200000AjNFq, a0L3200000DyWZW, a0L3200000B0eZl, a0L0e00000EPraL, a0L0e00000DxRZA, a0L32000009hM6S, a0L3200000BWkUy, a0L3200000Aiufr, a0L32000009mz67, a0L32000009hZGi, a0L0e00000DzgWq, a0L0e00000DzxMx, a0L0e00000EPr0D, a0L0e00000DxPpp, a0L3200000DhPHT, a0L3200000BTmZD, a0L0e00000DzlGS, a0L0e00000Dzbou, a0L3200000Dj4dg, a0L3200000BWkVI, a0L3200000BUNqA, a0L3200000DhpuB, a0L32000009ngPJ, a0L0e00000Dzwsn, a0L3200000DyqkE, a0L0e00000E0khR, a0L3200000Aitp5, a0L3200000BVYs4, a0L32000009ha19, a0L0e00000Dl5tt, a0L0e00000EPtpa, a0L3200000Ajbnf, a0L3200000Ajbs6, a0L3200000Ajbs1, a0L3200000B0fFK, a0L3200000Aje6m, a0L32000009hM0V, a0L3200000B0SmJ, a0L3200000AjVgP, a0L0e00000E0By8, a0L0e00000E0Bvi, a0L32000008qv9I, a0L32000008qv4h, a0L32000009hX7j, a0L32000009h9aE, a0L3200000BVn5C, a0L32000009e66V, a0L3200000Aj9ri, a0L3200000BVo5n, a0L3200000BVn4x, a0L3200000Aj9t5, a0L3200000Ajbaa, a0L3200000Aj9g1, a0L3200000Aj9gB, a0L3200000BVo5J, a0L0e00000EPsYQ, a0L32000009n1hH, a0L3200000BVo6C, a0L0e00000DxS0z, a0L3200000B0PYk, a0L0e00000EPtgE, a0L3200000BWbMz, a0L32000009mnxA, a0L0e00000Dl0VN, a0L0e00000Dl19c, a0L32000009nVre, a0L0e00000Dl1Az, a0L3200000Ajkg3, a0L32000009n30A, a0L3200000B0QR9, a0L3200000BTyCS, a0L3200000BgdVb, a0L0e00000E0kJP, a0L32000008rNGA, a0L3200000AihQV, a0L3200000Bz8Ab, a0L3200000Ajm9f, a0L32000009ng6Y, a0L3200000AiiNp, a0L0e00000EfL8R, a0L32000009h528, a0L3200000AjZ6p, a0L32000009nGFd, a0L32000009n0DG, a0L3200000AjkVw, a0L3200000BUl2k, a0L3200000B0Np6, a0L0e00000DkymE, a0L3200000DhS47, a0L32000009hIxq, a0L32000009md4Z, a0L0e00000Dl0Hz, a0L0e00000DkyrT, a0L0e00000Dl0Hp, a0L0e00000Dl0Hu, a0L0e00000Dl0I9, a0L3200000AifR8, a0L3200000Dhqyz, a0L3200000B21lP, a0L0e00000DzOBC, a0L3200000BTmHc, a0L0e00000E0BLA, a0L3200000B0U1R, a0L3200000B0U0n, a0L3200000B0U17, a0L3200000DyFXq, a0L3200000DyKil, a0L0e00000Dzbun, a0L3200000BWLpj, a0L32000009mdB6, a0L32000009mdAr, a0L32000009mdBB, a0L3200000DyFX2, a0L3200000DyFXl, a0L0e00000EPtxe, a0L0e00000EPwI5, a0L3200000BWO6L, a0L0e00000E0CYL, a0L3200000B1pSc, a0L0e00000EPrIg, a0L32000008pw81, a0L3200000BVTgG, a0L3200000B0ifE, a0L3200000B18AJ, a0L3200000AjogU, a0L3200000Bg7Nm, a0L3200000BzuJl, a0L60000007OXJQ, a0L0e00000EPutT, a0L32000009hK2H, a0L3200000Ajk9q, a0L3200000AjqD6, a0L3200000B0Ql9, a0L3200000DyHLS, a0L3200000Aj8Rt, a0L32000009hYBp, a0L3200000DpkqA, a0L3200000DyFYo, a0L3200000B0Ulh, a0L32000009h5Rw, a0L3200000B1Y5g, a0L0e00000DxQ38, a0L3200000BVIfK, a0L3200000DhCAh, a0L0e00000Dkz70, a0L32000009h5VN, a0L0e00000EPwlM, a0L3200000BVTfX, a0L0e00000EPr9K, a0L3200000B1KGw, a0L0e00000EPrAI, a0L0e00000E0BkL, a0L3200000BWBnO, a0L0e00000Dzbta, a0L32000009hJnR, a0L0e00000DqRZE, a0L32000009hX3w, a0L3200000AjfOU, a0L3200000AjjLs, a0L32000009nUHn, a0L3200000AjpNF, a0L3200000BUZR4, a0L3200000AjaSq, a0L0e00000EPrfp, a0L3200000BgR0I, a0L3200000BgR0S, a0L3200000BgR0D, a0L3200000BgR0X, a0L32000009nGQU, a0L3200000B19GF, a0L3200000BzVBr, a0L3200000BVY9F, a0L32000008qvLb, a0L32000008qpGZ, a0L0e00000EPq9Y, a0L0e00000EPq9O, a0L0e00000EPq9T, a0L0e00000EPq9J, a0L0e00000EPq9E, a0L0e00000EPto8, a0L0e00000EPscc, a0L0e00000Dkz1H, a0L0e00000Dkz2F, a0L0e00000Dl10f, a0L3200000Ajhtn, a0L3200000B2RzI, a0L32000009hNXQ, a0L0e00000DxY55, a0L3200000B0OEO, a0L3200000DyNJa, a0L32000008pH9X, a0L3200000Dhr5z, a0L3200000BVJFh, a0L32000009h9e2, a0L3200000BVKnN, a0L32000009hX7t, a0L3200000BTNBM, a0L3200000AjPyi, a0L32000009hA1d, a0L3200000AiuMk, a0L32000009hKKE, a0L0e00000EPsYp, a0L3200000Aitep, a0L3200000B1Klq, a0L0e00000DxP45, a0L3200000B2CZF, a0L3200000DpkhX, a0L32000008pbCK, a0L3200000DiFTa, a0L3200000BVUm7, a0L3200000BVXWm, a0L3200000BVXWh, a0L3200000AijvM, a0L3200000AijI0, a0L3200000AixJx, a0L32000009gYXr, a0L0e00000EPuoJ, a0L3200000DyPo6, a0L3200000B2QVC, a0L60000005LLQf, a0L3200000DyGTf, a0L3200000DyGS3, a0L32000009hYsJ, a0L3200000DyGTL, a0L3200000DyFYF, a0L3200000DyFYA, a0L0e00000DxT8L, a0L32000009gkfs, a0L3200000BUG6D, a0L3200000B0CZ9, a0L3200000Ajdak, a0L3200000Aj6i9, a0L3200000BVyAO, a0L32000008pZIe, a0L3200000AiwEa, a0L3200000AiwHp, a0L3200000DhPIC, a0L32000008qrs1, a0L0e00000Dl0gu, a0L3200000BWLFQ, a0L3200000BWLFB, a0L3200000BWAI4, a0L32000008pw9T, a0L32000009md4o, a0L3200000DyNPR, a0L3200000DhbAm, a0L3200000DhbAN, a0L3200000DyPro, a0L0e00000DzNIP, a0L3200000BWA1i, a0L0e00000Dzgf3, a0L3200000DyFiW, a0L3200000DyFfw, a0L0e00000E0C41, a0L3200000B0Umk, a0L32000009n3K7, a0L3200000B0U6m, a0L3200000B0U7B, a0L3200000DyLQp, a0L3200000B1my2, a0L32000009mcLl, a0L3200000Dpw1O, a0L32000009nUrR, a0L32000009hM9C, a0L32000009hKZs, a0L0e00000Dl0Bh, a0L3200000Aiv0I, a0L3200000BV9Vc, a0L3200000B0iGa, a0L0e00000E0gyT, a0L0e00000EfLJ0, a0L3200000Aj6ep, a0L3200000B2GLq, a0L32000009hX9V, a0L3200000BTbl9, a0L3200000BTMhy, a0L32000009n0Qa, a0L3200000B17w6, a0L3200000DyOtf, a0L3200000DyOu9, a0L32000008pKtJ, a0L0e00000DzQFx, a0L0e00000DzQFs, a0L3200000DyOux, a0L3200000DyOv7, a0L3200000DyOvq, a0L3200000Ajm8w, a0L0e00000E0SdW, a0L32000009hMXd, a0L3200000BVoob, a0L3200000BTad3, a0L3200000AjYAK, a0L0e00000DzNIj, a0L32000009hKwx, a0L3200000BWZQz, a0L3200000BUak3, a0L3200000DykWQ, a0L3200000DykVw, a0L3200000AjViM, a0L32000009nTTt, a0L3200000DhFbZ, a0L3200000DhF2F, a0L3200000DhNew, a0L3200000DhNer, a0L3200000DhNf1, a0L32000009mmXo, a0L32000009nGQy, a0L3200000BTaTO, a0L3200000B1Xxt, a0L3200000DhdZN, a0L3200000DhdZr, a0L3200000DhdYU, a0L3200000DhdZh, a0L3200000DhdZw, a0L3200000DhdZD, a0L3200000BgNch, a0L3200000AjnHD, a0L3200000BWZQu, a0L3200000BW02U, a0L3200000BzLLd, a0L32000009hX2y, a0L0e00000DzQEk, a0L3200000BzJOt, a0L0e00000DzVBi, a0L32000009nU2R, a0L3200000B1Znb, a0L3200000AjVxg, a0L32000009hJDr, a0L3200000BzhZ6, a0L3200000B09U9, a0L0e00000Dl0mx, a0L0e00000Dkyl1, a0L0e00000EfKgh, a0L3200000Bg6f0, a0L32000009hZ7v, a0L3200000Bzu7S, a0L32000008pUze, a0L32000008pUzy, a0L0e00000DxYph, a0L32000009mpoW, a0L3200000DyT5R, a0L0e00000E0M5O, a0L32000009n0Yo, a0L3200000DyDHm, a0L32000009mplR, a0L0e00000DxTU2, a0L0e00000E0CSX, a0L3200000BTZcU, a0L0e00000EPpDo, a0L3200000DhSCV, a0L3200000Ajd0d, a0L3200000BTlbK, a0L3200000B0jiZ, a0L0e00000EPqsd, a0L3200000BTyEe, a0L32000009hWVI, a0L3200000Dy3oR, a0L0e00000E0mcT, a0L0e00000E0mlk, a0L32000009nUQB, a0L3200000Ajhe7, a0L3200000Ajj12, a0L0e00000DxU41, a0L3200000AiimO, a0L3200000BVW8E a0L3200000Aihb4, a0L32000008rXSF, a0L32000008rXRx, a0L32000008rXSP, a0L3200000AihZr, a0L3200000B1zXx, a0L3200000BVZpa, a0L3200000AifYA, a0L60000008p4aW, a0L32000009hJo0, a0L3200000BUoJ3, a0L32000009hKnF",
      "more_code_info": ""
    }
  ]
}