  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saint Paul",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75813",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "4100 Hamline Ave N",
      "address_2": "N/A",
      "postal_code": "55112-5700",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Distribution to the states of : IA and NY., and to the countries of : United Kingdom, Netherlands, Italy, Germany and Belgium.",
      "recall_number": "Z-0938-2017",
      "product_description": "Boston Scientific, INGEVITY\" MRI Leads, Sterile EO, RxOnly.",
      "product_quantity": "15 (2 US; 13 OUS)",
      "reason_for_recall": "Boston Scientific is conducting a recall on Boston Scientific INGEVITY\" MRI endocardial pacing leads. There is a possibility that the polyurethane boot at the terminal end of the lead was not securely connected to the lead body.",
      "recall_initiation_date": "20161130",
      "center_classification_date": "20161230",
      "termination_date": "20171213",
      "report_date": "20170111",
      "code_info": "Model 7740 \u0013 S/N 633464, 610662.  Model 7741 \u0013 S/N 657588, 563539, 628198, 636545, 641593, 656713, 656904.  Model 7742 \u0013 S/N  572200, 584869, 631202.  Model 7732 \u0013 S/N 488740.  Model 7736 \u0013 S/N 631322, 632188."
    }
  ]
}