{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Louvain La Neuve",
      "state": "",
      "country": "Belgium",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79231",
      "recalling_firm": "Ion Beam Applications S.A.",
      "address_1": "Chemin du Cyclotron, 3",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to the states of : MA, FL, VA, PA, OK, TN",
      "recall_number": "Z-0934-2018",
      "product_description": "Proteus 235, Uniform and Pencil Beam Scanning    The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.",
      "product_quantity": "14",
      "reason_for_recall": "An accessory (range shifter, ridge filter&) can be improperly inserted in  snout 300x400 with one rail only and the proton therapy system can still  operate.",
      "recall_initiation_date": "20171205",
      "center_classification_date": "20180309",
      "termination_date": "20190329",
      "report_date": "20180321",
      "code_info": "PAT.000 (US), PAT.003 (KR), PAT.006 (US), PAT.107 (EU), PAT.108  (US), PAT.109 (US), PAT.110 (US), SAT.119 (US).",
      "more_code_info": ""
    }
  ]
}