  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64274",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "Worldwide Distribution-USA including the states of FL, IL, IN, NJ, NC, OK, TX, and WI, and the countries of UNITED KINGDOM, TUNISIA, SERBIA, RUSSIAN FEDERATION, NORWAY, NIGERIA, LITHUANIA, LEBANON, KOREA, REPUBLIC OF KOREA, ITALY, INDONESIA, INDIA, FRANCE, FINLAND, COLOMBIA, CHINA, and BRAZIL.",
      "recall_number": "Z-0934-2013",
      "product_description": "GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging  systems.     For use in generating fluoroscopic images of human anatomy.",
      "product_quantity": "48",
      "reason_for_recall": "Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.",
      "recall_initiation_date": "20110204",
      "center_classification_date": "20130311",
      "termination_date": "20140127",
      "report_date": "20130320",
      "code_info": "SERIAL NUMBER  592043BU4  592676BU1  593459BU1  593465BU8  592458BU4  592673BU8  592682BU9  593059BU9  593051BU6  593475BU7  593476BU5  592049BU1  593468BU2  593054BU0  592047BU5  592042BU6  592046BU7  592459BU2  593048BU2  593056BU5  593049BU0  593063BU1  593062BU3  593464BU1  594762BU7  593473BU2  592451BU9  593460BU9  597954BU7  592456BU8  592674BU6  593052BU4  593058BU1  590867BU8  592050BU9  593064BU9  592460BU0  592457BU6  592680BU3  592455BU0  593462BU5  592678BU7  594763BU5  593055BU7  592675BU3  592452BU7  593463BU3  594385BU7",