  "results": [
    {
      "status": "Terminated",
      "city": "Memphis",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87035",
      "recalling_firm": "Smith & Nephew, Inc.",
      "address_1": "1450 E Brooks Rd",
      "address_2": "",
      "postal_code": "38116-1804",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the state of OH and the countries of Dubai, Canada, Korea, Chile, Puerto Rico, Panama, and Costa Rica.",
      "recall_number": "Z-0933-2021",
      "product_description": "VERSABOND AB 40 GRAMS FORMULATION 2, antibiotic infused bone cement, Product Number: 71271440 - Product Usage: VERSABOND is a medium viscosity bone cement which can be used for hip, knee, and shoulder implant fixation. VERSABOND AB includes the antibiotic- gentamycin.",
      "product_quantity": "5190 devices",
      "reason_for_recall": "Potential heat and humidity exposure. The affected lot contained an atypical sandy, flaky or grainy consistency when mixing.",
      "recall_initiation_date": "20201218",
      "center_classification_date": "20210129",
      "termination_date": "20230501",
      "report_date": "20210210",
      "code_info": "Product Number: 71271440; Batch Number: 18LC07540",
      "more_code_info": ""
    }
  ]
}