      "country": "United States",
      "city": "Portage",
      "address_1": "3800 E Centre Ave",
      "reason_for_recall": "Stryker Medical has identified that on some units of the MTA7900,  there is a lack of a water temperature display while the unit is in Auto mode. While in Auto mode, only the patient temperature is displayed along with the set point. The water display will be blank during Auto mode. If the user wants to see the water temperature displayed and switches out of Auto mode to Manual mode, then back to",
      "address_2": "",
      "product_quantity": "1002 units",
      "code_info": "Domestic Serial Numbers   MTA7900C01023              MTA7900E01043             MTA7900F01021              MTA7900F90020             MTA7900G90007              MTA7900B90022             MTA7900B90023              MTA7900D80022              MTA7900B01006             MTA7900B01009             MTA7900B01012              MTA7900H90007             MTA7900H90008              MTA7900A01019             MTA7900A01020             MTA7900A01024              MTA7900D01028             MTA7900F01002             MTA7900H80009             MTA7900H80021              MTA7900G90016             MTA7900H90006              MTA7900K90056             MTA7900K90061             MTA7900K90063             MTA7900L90003             MTA7900L90006              MTA7900L90020              MTA7900H80057              MTA7900E01016              MTA7900I90006              MTA7900I90038              MTA7900B90020              MTA7900E80014             MTA7900E80020             MTA7900E80027              MTA7900I90023             MTA7900I90027              MTA7900E90031             MTA7900I90031             MTA7900I90041              MTA7900H90004              MTA7900B80004              MTA7900A80028             MTA7900E80054              MTA7900E01042              MTA7900D80001              MTA7900C01012              MTA7900J90018              MTA7900A01028              MTA7900E80046              MTA7900J70023             MTA7900J70024             MTA7900J70025             MTA7900J70026             MTA7900J70027             MTA7900J70031             MTA7900J70032              MTA7900E80011             MTA7900E80035              MTA7900H80014             MTA7900H80018             MTA7900H80019              MTA7900A90017             MTA7900K80010             MTA7900K80011             MTA7900K80012             MTA7900K80013             MTA7900L80001             MTA7900L80004             MTA7900L80013              MTA7900E80059              MTA7900A01014              MTA7900H90019             MTA7900H90020             MTA7900H90021             MTA7900H90022             MTA7900H90023              MTA7900D80055              MTA7900J90035              MTA7900J70028              MTA7900H80023              MTA7900A80032              MTA7900J70029              MTA7900I80029              MTA7900C80001             MTA7900C80002             MTA7900C80003              MTA7900J90027              MTA7900D80006              MTA7900E80029             MTA7900E80040             MTA7900E80065              MTA7900D01022             MTA7900E01022              MTA7900B01036              MTA7900F01010              MTA7900D90030             MTA7900K90003              MTA7900C01009              MTA7900C01025              MTA7900J60008              MTA7900A01032             MTA7900K90004              MTA7900E01029             MTA7900E90017              MTA7900J70022              MTA7900K90019              MTA7900B01040             MTA7900B01043              MTA7900B01024              MTA7900A90001             MTA7900A90005             MTA7900A90008             MTA7900E80022             MTA7900G80009             MTA7900G80011             MTA7900H80002             MTA7900H80003             MTA7900H80006             MTA7900H80007             MTA7900H80008             MTA7900H80010             MTA7900H80011             MTA7900H80012             MTA7900H80013              MTA7900L70008             MTA7900L70009              MTA7900G60063             MTA7900K40071              MTA7900B01019             MTA7900F01014              MTA7900C90010             MTA7900J80039              MTA7900K90053             MTA7900K90054              MTA7900A90012             MTA7900A90016             MTA7900A90020              MTA7900J90036              MTA7900K80005              MTA7900H80026              MTA7900L90017              MTA7900I90003             MTA7900I90008              MTA7900F01051              MTA7900I90043             MTA7900I90044             MTA7900I90046             MTA7900I90048              MTA7900G90012             MTA7900K90018              MTA7900J80051              MTA7900J80054              MTA7900A90030             MTA7900C90001              MTA7900F90018             MTA7900F90022             MTA7900L50006              MTA7900E01010              MTA7900H90009             MTA7900H90010             MTA7900H90014             MTA7900H90015             MTA7900H90016              MTA7900D90006             MTA7900D90009             MTA7900D90012             MTA7900D90013             MTA7900D90014             MTA7900D90015              MTA7900B90033             MTA7900I80025              MTA7900B01041             MTA7900C01002              MTA7900L90001              MTA7900B90017              MTA7900C01013             MTA7900C01018              MTA7900J80028              MTA7900A01045             MTA7900F90035              MTA7900E80058             MTA7900E80070              MTA7900K90001              MTA7900D80011              MTA7900I90022             MTA7900I90028              MTA7900A90034             MTA7900A90035              MTA7900G80007              MTA7900K90015              MTA7900J90011              MTA7900E80002             MTA7900E80021             MTA7900E80023             MTA7900E80025             MTA7900E80037             MTA7900E80044             MTA7900E80050             MTA7900E80055             MTA7900E80057             MTA7900E80060             MTA7900E80061             MTA7900E80062             MTA7900E80066             MTA7900E80067             MTA7900E80069             MTA7900E80071             MTA7900E80073             MTA7900E80074             MTA7900E80076             MTA7900E80078             MTA7900E80079             MTA7900H80039              MTA7900J80022              MTA7900A01005             MTA7900C01003             MTA7900C01011              MTA7900J90023              MTA7900D80027              MTA7900B01022             MTA7900H80032              MTA7900A01018             MTA7900A01027             MTA7900A01040             MTA7900B01016             MTA7900B01017             MTA7900B01021             MTA7900B01025              MTA7900L90011             MTA7900L90014              MTA7900D01024             MTA7900E01025             MTA7900I90040              MTA7900F01006              MTA7900G90006             MTA7900H80017              MTA7900B90014              MTA7900C90021              MTA7900I90032              MTA7900C90003              MTA7900A01001             MTA7900A01003             MTA7900A01046             MTA7900A70012             MTA7900A70013             MTA7900A70016             MTA7900A70019             MTA7900J70019             MTA7900J80027             MTA7900J80030              MTA7900E80017             MTA7900E80030             MTA7900E80034              MTA7900L90004              MTA7900A01034              MTA7900J80021             MTA7900J80041              MTA7900B01010             MTA7900F01011              MTA7900B80001              MTA7900A01030              MTA7900J90034              MTA7900D80031              MTA7900E90021              MTA7900D90026             MTA7900D90027              MTA7900L70007              MTA7900K90031              MTA7900F01005             MTA7900F90012             MTA7900F90014              MTA7900L90018              MTA7900E90020              MTA7900J90016              MTA7900F01056              MTA7900D80043              MTA7900J90005              MTA7900L90007              MTA7900E90004              MTA7900B01031              MTA7900C01020              MTA7900A90007              MTA7900A80041             MTA7900A80042             MTA7900A80043             MTA7900A80046              MTA7900F01039              MTA7900C90007              MTA7900J90026              MTA7900H80055             MTA7900I80009             MTA7900I80010             MTA7900I80011             MTA7900I80012              MTA7900L90015              MTA7900I90020             MTA7900I90021              MTA7900E80026              MTA7900E90026             MTA7900F90002              MTA7900L80003             MTA7900L80010              MTA7900A80031              MTA7900A80017              MTA7900D80053             MTA7900J80023              MTA7900J80024              MTA7900F01029              MTA7900F90031             MTA7900F90032             MTA7900G90009             MTA7900G90010             MTA7900G90020              MTA7900H80029             MTA7900J80038              MTA7900B90032             MTA7900J90012              MTA7900A80019              MTA7900D90024              MTA7900J90024              MTA7900I80014             MTA7900I80015             MTA7900I80016              MTA7900C90017             MTA7900H90002              MTA7900J90028              MTA7900F90015             MTA7900L80002             MTA7900L80011              MTA7900B01035             MTA7900B01038             MTA7900B01039             MTA7900C01001              MTA7900E01014              MTA7900A80002              MTA7900A80025              MTA7900D01013              MTA7900E01038              MTA7900K90058             MTA7900K90060              MTA7900C01010              MTA7900C90008              MTA7900D80008             MTA7900D80009             MTA7900D80010              MTA7900G01010              MTA7900A80037              MTA7900D01023              MTA7900E01021              MTA7900J80018              MTA7900K80002              MTA7900H80027              MTA7900A01013              MTA7900A90003             MTA7900A90010             MTA7900J80026              MTA7900E80024             MTA7900E80039              MTA7900E80008             MTA7900H80046             MTA7900J80025             MTA7900K80003             MTA7900L80005              MTA7900F90021              MTA7900E80041             MTA7900E80051              MTA7900H80028             MTA7900H80042             MTA7900I80017             MTA7900I80018             MTA7900I80020             MTA7900I80022              MTA7900D80029              MTA7900E80048             MTA7900E80052             MTA7900E80053             MTA7900E80063             MTA7900E80064              MTA7900C90002             MTA7900C90016             MTA7900D80012             MTA7900D90003             MTA7900E90002             MTA7900E90005             MTA7900E90011              MTA7900J80029             MTA7900L80009              MTA7900B90031             MTA7900D90025             MTA7900F90001             MTA7900F90003             MTA7900F90007             MTA7900F90009             MTA7900F90011             MTA7900F90013              MTA7900E01035              MTA7900L80017              MTA7900J80036              MTA7900L80007             MTA7900L80012              MTA7900I90039             MTA7900I90042              MTA7900A80039              MTA7900L80014             MTA7900L80018              MTA7900E90023              MTA7900I90005             MTA7900I90007             MTA7900I90009              MTA7900D01008              MTA7900C01005             MTA7900C01006              MTA7900I90010              MTA7900C90018             MTA7900C90020              MTA7900A80038             MTA7900A90025              MTA7900A80012              MTA7900B01030              MTA7900K80008              MTA7900F01041             MTA7900F01055              MTA7900F01024              MTA7900E90025              MTA7900C90014             MTA7900E01005              MTA7900A01008             MTA7900B01014              MTA7900I80021             MTA7900I80024              MTA7900B01002             MTA7900B01026             MTA7900B01037              MTA7900F01022              MTA7900C90022             MTA7900D90005             MTA7900E80045             MTA7900E80075             MTA7900J80031             MTA7900J80032             MTA7900J80033             MTA7900J80035              MTA7900A70014             MTA7900A70015             MTA7900A90028             MTA7900A90031             MTA7900A90032              MTA7900D80028             MTA7900K90024             MTA7900K90027              MTA7900J90007             MTA7900J90020              MTA7900A01025             MTA7900A01026              MTA7900C01026              MTA7900E80031              MTA7900C01019              MTA7900A80036              MTA7900B01023             MTA7900F01037              MTA7900E01001              MTA7900B01001              MTA7900B01003              MTA7900B01004             MTA7900B01005             MTA7900B01007             MTA7900K90002              MTA7900A01015             MTA7900D80002             MTA7900F90034              MTA7900G80012             MTA7900G80014              MTA7900I90026             MTA7900I90029              MTA7900H80043              MTA7900A80015             MTA7900E80032             MTA7900E80036              MTA7900L90009              MTA7900B90009             MTA7900D90001             MTA7900D90004             MTA7900F01004             MTA7900J80043              MTA7900J80046             MTA7900K80001              MTA7900B90029             MTA7900C90013              MTA7900D90002             MTA7900D90007             MTA7900D90008             MTA7900D90010              MTA7900G01004             MTA7900G01005             MTA7900G01013             MTA7900G01014             MTA7900G01015              MTA7900D60059             MTA7900D60061             MTA7900E60002             MTA7900J70014             MTA7900J70020              MTA7900I90012              MTA7900B90027             MTA7900B90035              MTA7900B01027             MTA7900G90008              MTA7900I90036             MTA7900I90037              MTA7900J90022              MTA7900D90029              MTA7900A01042              MTA7900G80013              MTA7900K90057              MTA7900J90025              MTA7900B01033              MTA7900I90016             MTA7900I90017              MTA7900A80008             MTA7900A80009             MTA7900A80010             MTA7900A80014             MTA7900E80012             MTA7900E80015             MTA7900F90008              MTA7900E01037              MTA7900D80020             MTA7900D80021              MTA7900B01032              MTA7900A90004              MTA7900J80019              MTA7900G01011              MTA7900E01019              MTA7900A90027             MTA7900C90009             MTA7900C90011              MTA7900I90011             MTA7900I90014              MTA7900I80019              MTA7900J90017              MTA7900E90014             MTA7900E90016              MTA7900C01024              MTA7900D80013             MTA7900D80016              MTA7900D90011              MTA7900D80046              MTA7900E01004              MTA7900C01021              MTA7900E01033             MTA7900F01018             MTA7900F01023             MTA7900F01025             MTA7900F01028             MTA7900G90005              MTA7900F90006              MTA7900I90024             MTA7900I90025              MTA7900F01008             MTA7900F01031             MTA7900F01033             MTA7900F01040              MTA7900A80013             MTA7900A80024             MTA7900K90009             MTA7900K90013              MTA7900E01002             MTA7900E01015              MTA7900A01002             MTA7900L90023             MTA7900L90024             MTA7900L90025              MTA7900F01034              MTA7900B90018              MTA7900E90015              MTA7900J80017             MTA7900J80020              MTA7900J80040              MTA7900E80042             MTA7900K90014              MTA7900C80007              MTA7900A01016             MTA7900A01017              MTA7900H80016             MTA7900K80004              MTA7900D80007             MTA7900J70002              MTA7900G90015              MTA7900F01045              MTA7900A80020              MTA7900L90002             MTA7900L90005              MTA7900K90055              MTA7900L90012              MTA7900J90014             MTA7900J90015              MTA7900A80040             MTA7900J80045              MTA7900A80006             MTA7900A80044              MTA7900J90031             MTA7900J90032             MTA7900J90033              MTA7900D80048             MTA7900E01032             MTA7900F01027             MTA7900F01044             MTA7900F01054              MTA7900F01032             MTA7900F01050              MTA7900H80051              MTA7900E90022             MTA7900E90024             MTA7900E90029             MTA7900F90004              MTA7900B01042             MTA7900C01008              MTA7900G80002             MTA7900G80003             MTA7900G80005             MTA7900G80008              MTA7900H80038              MTA7900H90024             MTA7900I90018              MTA7900H80036             MTA7900J80034             MTA7900J80048              MTA7900B01013             MTA7900B01015             MTA7900K90030              MTA7900A80018             MTA7900F90029              MTA7900F90024              MTA7900K90039              MTA7900D80014             MTA7900D80015              MTA7900I80006              MTA7900L70002             MTA7900L70004              MTA7900F01012              MTA7900B90008             MTA7900B90028             MTA7900B90030              MTA7900A90029              MTA7900J90037             MTA7900L90010              MTA7900B90038              MTA7900H80044              MTA7900D01002             MTA7900D01004              MTA7900C90004             MTA7900D90016             MTA7900D90017              MTA7900C90012              MTA7900I90015             MTA7900I90019              MTA7900F90028              MTA7900J80042              MTA7900K90026             MTA7900K90029             MTA7900K90032             MTA7900K90033             MTA7900K90034             MTA7900K90036             MTA7900K90037             MTA7900K90040             MTA7900K90043              MTA7900K90006             MTA7900K90008              MTA7900E01017              MTA7900D80036              MTA7900A01035             MTA7900C01027              MTA7900C90015             MTA7900E90001             MTA7900E90006             MTA7900E90007             MTA7900E90008             MTA7900E90009             MTA7900E90012             MTA7900E90027              MTA7900A01006             MTA7900A01007             MTA7900A01009             MTA7900A01010             MTA7900A01011             MTA7900A01012              MTA7900C80004             MTA7900C80005              MTA7900F90023             MTA7900J90021              MTA7900E01027              MTA7900L70005              MTA7900C90006              MTA7900F90005              MTA7900D80003             MTA7900D80023             MTA7900D80024              MTA7900B90021              MTA7900E80038              MTA7900E90010              MTA7900A01033              MTA7900D01014             MTA7900D01019              MTA7900C01014              MTA7900A90002              MTA7900K90022              MTA7900D01001             MTA7900D01003              MTA7900E90013              MTA7900G90018             MTA7900G90019             MTA7900G90024              MTA7900H90018              MTA7900B90016              MTA7900H90005              MTA7900I90033             MTA7900I90034             MTA7900I90035              MTA7900A80048             MTA7900D80005              MTA7900D90020             MTA7900D90028             MTA7900E90003              MTA7900E80009              MTA7900H90001             MTA7900H90011              MTA7900J90029              MTA7900B01011             MTA7900B01020              MTA7900K90059              MTA7900B90007             MTA7900B90010             MTA7900E01040             MTA7900F01009             MTA7900G90013             MTA7900G90022              MTA7900A80047             MTA7900D80044              MTA7900D80045              MTA7900H80061             MTA7900I80002              MTA7900I90013              MTA7900H80056             MTA7900H80058             MTA7900H80059             MTA7900H80060              MTA7900E90030              MTA7900C80006             MTA7900C80008             MTA7900C80009              MTA7900A01041             MTA7900A01043              MTA7900B90037             MTA7900B90040             MTA7900B90042              MTA7900L90021             MTA7900L90022              MTA7900A90013              MTA7900J90013             MTA7900J90019              MTA7900E70005             MTA7900E70006             MTA7900F01046             MTA7900F01048             MTA7900G80004             MTA7900G80010              MTA7900G80006              MTA7900E01024              MTA7900H80037             MTA7900H80040              MTA7900D01005             MTA7900D01006             MTA7900D01007             MTA7900D01009             MTA7900D01010             MTA7900D01011              MTA7900H80022             MTA7900H80024             MTA7900H80031             MTA7900H80033             MTA7900H80035             MTA7900H80047             MTA7900H80048             MTA7900H80050             MTA7900H80052             MTA7900H80053             MTA7900H80054              MTA7900A80001             MTA7900L70010             MTA7900L70011             MTA7900L70012             MTA7900L70013             MTA7900L70015             MTA7900L70016             MTA7900L70017             MTA7900L70018             MTA7900L70019             MTA7900L70020             MTA7900L70021             MTA7900L70022             MTA7900L70023             MTA7900L70024              MTA7900A90011              MTA7900E80005              MTA7900E80072             MTA7900I80030             MTA7900I80031             MTA7900I80032             MTA7900I80033             MTA7900I80036             MTA7900I80037             MTA7900I80038             MTA7900I80039             MTA7900I80040             MTA7900I80041             MTA7900I80042             MTA7900I80043             MTA7900I80044             MTA7900I80045             MTA7900I80046             MTA7900I80047             MTA7900I80048             MTA7900I80049             MTA7900J80001             MTA7900J80002             MTA7900J80003             MTA7900J80004             MTA7900J80005             MTA7900J80006             MTA7900J80007             MTA7900J80008             MTA7900J80009             MTA7900J80010             MTA7900J80011             MTA7900J80012             MTA7900J80013              MTA7900D80050              MTA7900B80003             MTA7900D80032             MTA7900D80038             MTA7900D80039             MTA7900D80041             MTA7900D80042              MTA7900B01029             MTA7900B01034              MTA7900D90018             MTA7900H90017             MTA7900H90025             MTA7900H90026             MTA7900H90027             MTA7900H90028             MTA7900H90029             MTA7900H90030             MTA7900I90001             MTA7900I90002             MTA7900I90004              MTA7900I80027             MTA7900I80028             MTA7900J80014             MTA7900J80016              MTA7900C01004             MTA7900C01007              MTA7900C01022              MTA7900D01012              MTA7900J70015              MTA7900H80020              MTA7900B90004             MTA7900B90015              MTA7900I80026              MTA7900E01006             MTA7900E01009             MTA7900E01011             MTA7900E01012             MTA7900E01013              MTA7900D01026             MTA7900D01027             MTA7900F90026              MTA7900F01017             MTA7900F01030             MTA7900F01053              MTA7900D01021             MTA7900E01007             MTA7900E01008              MTA7900L80015             MTA7900L80016              MTA7900B80002              MTA7900E01039              MTA7900C90019             MTA7900D90022              MTA7900I80005              MTA7900J80037             MTA7900J80050             MTA7900J80052              MTA7900K90007             MTA7900K90010              MTA7900A90019             MTA7900A90021             MTA7900A90024             MTA7900A90026             MTA7900A90033              MTA7900A90022              MTA7900A01038             MTA7900A01039              MTA7900B90043             MTA7900L90008              MTA7900K90038             MTA7900K90041              MTA7900D01016             MTA7900D01017             MTA7900D01018             MTA7900D01020              MTA7900E80001             MTA7900I80023              MTA7900E90019             MTA7900K90025             MTA7900K90035             MTA7900K90044             MTA7900K90045             MTA7900K90046             MTA7900K90047             MTA7900K90048             MTA7900K90049             MTA7900K90050             MTA7900K90051             MTA7900K90052              MTA7900B90003             MTA7900B90005              MTA7900D90019              MTA7900D90023              MTA7900K90016             MTA7900K90017             MTA7900K90021             MTA7900K90023              MTA7900G90011             MTA7900G90014             MTA7900H80034              MTA7900I80001             MTA7900I80034              MTA7900B90024             MTA7900B90025     International Serial Numbers     MTA7900E01003              MTA7900H90003              MTA7900A90014              MTA7900A90015              MTA7900I90047              MTA7900I90049              MTA7900I90050              MTA7900I90051              MTA7900I90052              MTA7900J90001              MTA7900J90002              MTA7900J90003              MTA7900J90004              MTA7900J90006              MTA7900J90008              MTA7900J90009              MTA7900J90010              MTA7900A01022              MTA7900E01031              MTA7900E80003              MTA7900E80019              MTA7900F01047              MTA7900H80025              MTA7900H80041              MTA7900H80045              MTA7900I80004              MTA7900I80007              MTA7900I80008              MTA7900L70006              MTA7900L90019              MTA7900F01043",
      "center_classification_date": "20140205",
      "distribution_pattern": "nationwide and worldwide",
      "state": "MI",
      "product_description": "The Gaymar Medi-Therm device provides a means of regulating patient temperature by supplying temperature-controlled water through a connector hose to accessory Gaymar Hyper/Hypothermia blanket(s)/body wrap(s). The blanket/body wrap provides an interface for heating or cooling the patient. Accessory YSI 400 series patient probe interfaces between the  Medi-Therm and patient to sense patient temperature, which is displayed on the device's control panel. The Medi-Therm device controls output water temperature by mixing hot and cold water using hot and cold solenoid valves under microprocessor control. The device includes a circulating pump, heater and refrigeration system.  The Medi-Therm controller, connector hose, blanket/body wrap and patient probe comprise the Medi-Therm Hyper/Hypothermia System.",
      "report_date": "20140212",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Medical Division of Stryker Corporation",
      "recall_number": "Z-0933-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66861",
      "termination_date": "20141104",
      "more_code_info": "",
      "recall_initiation_date": "20131227",
      "postal_code": "49002-5826",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}