  "results": [
    {
      "status": "Terminated",
      "city": "Best",
      "state": "",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82038",
      "recalling_firm": "Philips Medical Systems Nederlands",
      "address_1": "Veenpluis 4-6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. Nationwide distribution.",
      "recall_number": "Z-0930-2019",
      "product_description": "MARLIN 1.5T, Model Number 781474.   Nuclear Magnetic Resonance Imaging System",
      "product_quantity": "11226 total",
      "reason_for_recall": "There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.",
      "recall_initiation_date": "20181221",
      "center_classification_date": "20190222",
      "termination_date": "20201119",
      "report_date": "20190306",
      "code_info": "All systems",
      "more_code_info": ""
    }
  ]
}