  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84335",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 101",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of Alabama, Alaska, Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Kansas, Kentucky, Louisiana, Massachusetts, Michigan, Minnesota, Missouri, Montana, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Texas, Virginia, Washington, West Virginia, Wisconsin and countries of Belgium, Brazil, France, Germany, India, Ireland, Italy, Japan, Norway, Panama, Poland, South Korea, Spain, Taiwan, and United Kingdom",
      "recall_number": "Z-0926-2020",
      "product_description": "da Vinci X Surgical System Surgical System, IS4200, 00886874114216  Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.",
      "product_quantity": "52 surgical systems",
      "reason_for_recall": "The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.",
      "recall_initiation_date": "20190726",
      "center_classification_date": "20200131",
      "termination_date": "20220810",
      "report_date": "20200212",
      "code_info": "System Serial #  PSC Serial #  VSC Serial #   SL0350  665033  Not Affected      SL0360  292201  Not Affected      SL0361  Not Affected  1000216      SL0362  381021  678362      SL0365  34608  Not Affected      SL0366  Not Affected  1000217      SL0369  Not Affected  679520      SL0370  Not Affected  678840      SL0371  Not Affected  305493      SL0372  415418  678699      SL0373  429524  334972      SL0375  332395  678509      SL0376  Not Affected  680721      SL0377  Not Affected  991369      SL0378  266528  872469      SL0379  311710  676421      SL0380  Not Affected  268852      SL0385  Not Affected  304487      SL0386  62043  681143      SL0387  308179  687814      SL0388  Not Affected  688819      SL0389  71796  308181      SL0391  518359  285588      SL0392  Not Affected  685499      SL0393  295941  687813      SL0394  281121  687293      SL0395  43875  689242   SL0396  Not Affected  685077   SL0397  892731  685500   SL0398  Not Affected  330718   SL0399  Not Affected  685080   SL0400  492489  769792   SL0401  292512  682912   SL0402  Not Affected  682536   SL0403  Not Affected  688818   SL0404  278656  691238   SL0406  680584  324350   SL0407  Not Affected  688038   SL0408  81482  331850   SL0409  Not Affected  691236   SL0410  768663  Not Affected   SL0411  685403  695335   SL0412  Not Affected  692045   SL0413  Not Affected  698304   SL0415  870352  699139   SL0416  Not Affected  343215   SL0417  697081  692043   SL0418  268716  694230   SL0419  264189  697313   SL0420  279659  699148   SL0423  291981  260861   SL0424  Not Affected  683692"
    }
  ]
}