{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Charleston",
      "state": "SC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69927",
      "recalling_firm": "Trumpf Medical Systems, Inc.",
      "address_1": "1046 Legrand Blvd",
      "address_2": "N/A",
      "postal_code": "29492-7672",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS,  NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.",
      "recall_number": "Z-0924-2015",
      "product_description": "(R)IR remote control JUPITER, Material no.1219656, Serial number: to (incl.) 102317869.  Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.",
      "product_quantity": "3303 total all devices",
      "reason_for_recall": "Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes",
      "recall_initiation_date": "20141203",
      "center_classification_date": "20150106",
      "termination_date": "20160823",
      "report_date": "20150114",
      "code_info": "The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.",
      "more_code_info": ""
    }
  ]
}