  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Allentown",
      "address_1": "901 Marcon Blvd",
      "reason_for_recall": "Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.",
      "address_2": "",
      "product_quantity": "30, 240 units",
      "code_info": "Catalog #CT-2000-100lots 3H0125F151 and 3H0225F151, expiration 7/2018.",
      "center_classification_date": "20140203",
      "distribution_pattern": "US Distribution one importer/wholesale distributor in MO.",
      "state": "PA",
      "product_description": "B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML    Generally used for delivery of antibiotic and local anesthesia medications.",
      "report_date": "20140212",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "B. Braun Medical, Inc.",
      "recall_number": "Z-0921-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67172",
      "termination_date": "20140903",
      "more_code_info": "",
      "recall_initiation_date": "20131213",
      "postal_code": "18109-9512",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}