  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75860",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to: IL  IL  GA  SC  SD  OK  NY  MI  IN  MD  AR",
      "recall_number": "Z-0920-2017",
      "product_description": "SOMATOM Scope Power, Computed tomography x-ray system",
      "product_quantity": "N/A",
      "reason_for_recall": "Siemens is providing a new installation of the   Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.",
      "recall_initiation_date": "20161122",
      "center_classification_date": "20161223",
      "termination_date": "20170530",
      "report_date": "20170104",
      "code_info": "Model # 10967888  Serial Number:59376  91766  91767  115456  115454  59380  77560  98840  77557  91758  59377  59378  115451"
    }
  ]
}