{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75860",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to: IL  IL  GA  SC  SD  OK  NY  MI  IN  MD  AR",
      "recall_number": "Z-0919-2017",
      "product_description": "SOMATOM Scope, Computed tomography x-ray system",
      "product_quantity": "13 systems",
      "reason_for_recall": "Siemens is providing a new installation of the   Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.",
      "recall_initiation_date": "20161122",
      "center_classification_date": "20161223",
      "termination_date": "20170530",
      "report_date": "20170104",
      "code_info": "Model # 10967666  Serial Number:59376  91766  91767  115456  115454  59380  77560  98840  77557  91758  59377  59378  115451"
    }
  ]
}