      "country": "United States",
      "city": "Draper",
      "address_1": "12050 Lone Peak Pkwy",
      "reason_for_recall": "Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is  rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.",
      "address_2": "",
      "product_quantity": "422",
      "code_info": "lot 59617777 and 59658495.",
      "center_classification_date": "20140203",
      "distribution_pattern": "Nationwide Distribution",
      "state": "UT",
      "product_description": "Crimper Model 9100CR26    The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation.",
      "report_date": "20140212",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Edwards Lifesciences, LLC",
      "recall_number": "Z-0918-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67362",
      "termination_date": "20140324",
      "more_code_info": "",
      "recall_initiation_date": "20140120",
      "postal_code": "84020-9414",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}