  "results": [
    {
      "country": "United States",
      "city": "Carlsbad",
      "address_1": "2271 Cosmos Ct",
      "reason_for_recall": "Potential issues with ventilator navigation ring interface and liquid ingress:  1) The navigation ring may become inoperative or operate intermittently;  2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments.",
      "address_2": "",
      "product_quantity": "3,453 systems",
      "code_info": "Model: V680  Part Number: 850011  V680 Ventilator, Finished    Serial Numbers:  201009427  201009428  201009429  201009493  201009494  201009495  201009496  201009498  201009499  201009660  201009661  201009662  201009663  201009664  201009665  201009666  201009667  201009668  201009669  201009670  201009671  201009672  201009673  201009675  201009676  201009677  201009678  201009679  201009674  201008743  201008765  201008766  201008770  201008771  201008776  201008785  201008786  201008805  201009497  201008801  201008741  201008744  201008760  201008764  201008767  201008768  201008772  201008775  201008783  201008808  201008762  201008773  201008781  201008809  201008807  100124645  100124646  100125009  100125011  100125012  100125013  100125014  100125148  100125149  100125150  100125152  100125153  100125154  100125155  201008740  201008742  201008774  201008777  201008802  201008804  100125541  100125544  201008780  201008800  201008803  100125504  100125505  100125509  100125512  100125542  100125545  100125550  100125551  100125552  201008784  100125508  100125546  100125549  100125010  100125516  100125517  100125151  100125515  100125547  201008782  100125513  100125510  100129145  201010488  201010551  201010627  100129141  100129142  100129143  201010254  201010535  201010546  201010674  100126254  100129144  100134076  201010428  201010482  201010490  201010536  201010548  201010620  201010622  201010675  100125518  100129147  100134077  201010255  201010360  201010434  201010485  201010538  201010623  201011218  100125553  100129146  100129148  100134073  100134074  100134075  100134078  100134079  100134180  100134181  100134182  100134183  201010251  201010252  201010256  201010257  201010258  201010259  201010426  201010427  201010429  201010430  201010438  201010484  201010491  201010537  201010539  201010540  201010544  201010545  201010547  201010550  201010552  201010553  201010624  201010625  201010626  201011216  201011217  201011220  100134184  201010439  201010483  100134337  100135409  100125543  100134335  100135396  100135398  100135400  100135401  100135402  100135404  100135405  100135406  100135407  100135631  100135633  100135634  100135635  100135636  100135638  100135639  100135630  100135637  100135642  100135644  100135650  201011219  100135640  100135641  100135643  100135647  100135652  100135397  100135645  100135646  100135648  100135649  100135651  100135653  100135654  100135655  100137426  100137427  100137428  100137429  100137430  100137431  100137433  100137434  100137436  100137437  100137438  100137439  100137440  100137444  100137446  100137447  100137448  100137449  100137450  100138009  100138010  100138011  100138012  100138016  100138017  100138018  100138019  100138508  100138510  100135403  100135632  100137435  100137442  100137443  100137445  100138013    100138080  100138082  100138507  100137432  100137441  100138008  100138081  100138083  100138086  100138087  100138503  100138512  100138513  100138514  100138515  100138522  100138084  100138085  100138509  100138511  100138517  100138518  100138519  100138520  100138521  100138705  100138707  100138814  100138816  100138818  100138943  100139067    100138506  100138516        100138819  100138940  100138941  100138942  100139064  100139066  100139068  100139069  100139070  100139071  100139074  100139075  100139077  100139078  100139079  100139080  100139082  100139083  100139084  100139087  100140238  100140239  100140300  100140301  100140302  100140303  100140304  100140305  100140310  100140311  201012739  201013230  201013231  100139076  100139081  100139088  100140308  100140315  100140317  201012733  201012735    201013213  201013220  201013221  201013224  201013225  201013226  201013228  201013229  100138504  100139072  100139085  100140312  100140314  100140316  201012732  201012734  201012736  201013209  201013210  201013211  201013212  201013215  201013223  201013227  201013233  201013234  201013235  100139065  100139073  100140309  201012738  201013214  201013236  100139086  100140306  100140313  100142426  100142428  100142429  100142431  100142432  100142434  100142437  100142460  201013217  201013218  201013219  201013222  100135399  100135408          100142637  100142427  100142436  100142438  201013216  201013232  100142634  100142435  100142638  100142640  100142641  100142635  100142636  100142642  201013271  201013272  201013275  201013273  201013276  201013279  201013720  201013725  201013747  100142639  201013274  201013277  201014176  201014178  201014179  201014182  201014183  201014211  201014213  201013713  201013717  201013749  201013887  201014185  201014214  201013890  201014177  201014181  201014184  201014196  201014198  201014199  201014215  201014216  201014217  201014218  201014219  201013709  201013710  201013711  201013712  201013714  201013715  201013716  201013724  201013744  201013748  201013750  201013753  201013888  201013889  201013891  201013893  201014197  201014340  201014180  201014324  201014342  201014388  201014195  100146372  100146374  201014322  201014323  201014326  201014328  201014386  100146369  100146370  100146371  100146373  100146873  201014321  201014325  201014327  201014341  201014343  201014344  100147209  100146442  100146875  100146375  100146876  100147212  201014544  100147217  201014534  201014536  201014542  201014600  201014538  100147215  100146376  100146377  100146440  100146443  100146872  100147218  100147300  201014329  100146441  100146869  100147219  201014543  100146871  100147211  100147301  201014537  201014539  100146867  201014384  201014385  201014535  201014545  100146378  100146379  100146868  100146870  100146874  100147210  100147216  201014330  201014387  201014541  201010431  201010489  100147214  201010486  201013746  201010487  100147208  100147213  201010432  201010549  201010621  201011215  100129149  100147302  100148513  100148515  201010433  100154345    100153251  100153256  100154348  201015021  100153252  100154346  100154349  100154353  100156200  100156201  100160490  201015022  100154721  100154722  100153451  100154696  100154698    100153257  100153448  100153449  100154352  201015023  100153447  100154697  100154699  100154724  100156105  100156202  100156204  100156205  100160493  100160494  100160495  100160496  100160962  100160975  100160663  100153450  100154347  100154351  100154354  100156206  100160489  100160492  100160498  100160664  100160665  100160666  100160668  100160918  100160919  100160960  100160964  100160965  100160966  100160968  100160969  100160970  100160971  100160972  100160973  100160974  100160976  100160977  100160978  100160979  100160980  100160981  100160982  100160983  100160984  100161100  100161101  100161102  100161103  100161107  100161109  100161110  100161112  100154350  100160667  100160961  100160967  100161105  100161108  100161111  100161113  100161114  100161115  100161116  100161117  100161119  100161120  100160963  100161104  100161106  100161121  100161122  100161845  100161846  100161847  100161848  100162258  100162300  100162303  100162259  100162257  100162301  100161849  100162256  100162311  100162312  100154725  100162255  100162302  100162305  100162306  100162307  100162308  100163248  100162314  100162304  100162313  100163045  100163249  100162310  100163049  100163050  100163051  100163054  100163064  100163066  100163067  100163068  100163242  100163245  100163250  100163357  100163358  100163359  100163420  100163421  100163782  100163784  100163787  100163791  100162309  100163065  100163244  100163246  100163247  100163792  100163047  100163052  100163241  100163786  100163789  100163783  100163048  100163785  100163790  100163793  100163794  100163795  201016701  201016702  201016705  100163788  100165980  100165981  100165982  100165983  100165984  100165985  100165986  100165987  100165988  100165989  100165990  100165991  100165992  201016703  201016704  201016719  201016920  201016922  201016923  100163053          201016926  201017222    100168214  100168219  100168263  100168274  100168282  100168817  100168867  100168868  100168869  100168870  100168874  100168818  100169142  100168814  100168872  100168875  100169144  100169146  100169148  100169380  100168812  100168813  100168815  100168816  100168819  100168866  100168873  100169141  100169143  100169145  100169147  100169149  100169150  100169151  100169152  100169153  100169154  100169155  100169156  100169157  100169158  100169159  100170298  100170301  100170302  100170379  100170440  100170441  100170442  100170447  100169381  100170300  100170444  100169500  100170299  100170443  100170445  100170446  100170449  100170450  100170451  100170452  100170453  100170454  100170455  100170457  100170448  100170456  100170461  100174095  100174096  100174097  100174098  100174099  100174261  100174262  100170460  100174260  100174263  100174392  100174395  100170458  100170459  100174264  100174393  100174394  100174396  100174397  100174399  100174503  100174504  100174506  100175290  100174398  100174505    100175288  100175361  100175362  100175363  100175364  100175365  100175367  100175369  100175371  100175334  100175366  100174507  100174510  100175284  100175286  100175289  100175291  100175292  100175325  100175326  100175327  100175330  100175331  100175332  100175333  100175335  100175337  100175339  100175368  100175372  100175373  100175773  100175774  100175776  100175778  100174502  100175293  100175338  100175360  100175374  100175775  100175777  100175779  100176180  100176181  100176441  100176442  100176443  100175329  100176444  100176445  100176451  100176657  100176800  100176802  100176803  100176805  100176807  100176811  100176825  100176826  100176827  100174501  100176182  100176812  100176813  100176821  100175328  100175336  100176446  100176447  100176448  100176449  100176658  100176759  100176808  100176809  100176817  100176822  100176824  100176450  100176814  100176820    100160661  100177146  100179320  100179996  100179998  100179999  100180003  100180004  100180005  100180001  100180006  100180007  100180011  100180021  100180321  100180325  100180012  100180013  100180013  100180008  100180009  100179819  100180000  100180002  100180010  100180014  100180015  100180016  100180017  100180018  100180019  100180020  100180024  100180323  100180324  100180326  100177145  100179997  100180328  100176275  100176276  100176277  100176278  100176279  100176810  100177140  100177141  100177142  100177143  100177144  100177147  100177148  100177149  100179310  100179311  100179312  100179313  100179314  100179315  100179316  100179317  100179318  100179319  100179324  100161118  100180022  100180320  100180322  100180327  100180916  100180917  100180918  100180919  100181640  100181641  100181642  100181644  100181645  100181646  100181647  100181649  100181652  100181653  100182076  100182074  100182078    100180915  100181643            100182289        100181648  100181650  100181651  100182073  100182075  100182077  100182081  100182290  100182292  100182294  100182295  100182297  100182298  100182299  100182300  100182301  100182302  100182303  100182304  100182305  100182306  100182307  100183045  100183046  100183047  100183049  100183050  100183057  100183059  100183117  100183304  100183355  100183048  100183051  100183054  100183055  100183056  100183114  100183119  100183303  100183305  100183310  100183311  100183312  100183313  100183317  100183318  100183354  100183356  100183359  100183360  100183362  100183053  100183115  100183302  100183307  100183309  100183319  100183353  100183400  100183401  100183403  100183705  100183612  100183713  100183118  100183308  100183538  100183553  100183556  100183557  100183559  100183566  100183604  100183605  100183606  100183607  100183611  100183700  100183702  100183703  100183704  100183706  100183707  100183708  100183709  100183710  100183711  100183714  100183715  100183716  100183717  100183718  100183721  100183724  100183841  100183306  100183527  100183529  100183530  100183532  100183535  100183572  100183608  100183610  100183842  100183843  100183847  100183849  100183850  100183052  100183058  100183358  100183361  100183402  100183524  100183525  100183526  100183528  100183531  100183533  100183534  100183536  100183539  100183554  100183555  100183558  100183560  100183565  100183570  100183603  100183701  100183720  100183844  100183848  100183116  100183537  100183561  100183562  100183563  100183564  100183569  100183571  100183601  100183609  100183722  100183846  100183856  100183845  100183851  100183854  100183567  100183568  100183602  100183719  100183723  100183840  100183852  100183853  100183301  100183357  100183855  100183858  100183857  201020755  201020752  201020756  201020758  100183859  100183860  201020751  201020753  201020754  201021222  201021223  201021224  201021225  201021226  201021227  201021228  201021229  201021230  201021231  201021232  201021233  201021234  201021235  201021236  201021237  201021238  201021239  201021260  201021261  201021262  201021263  201021265  201021267  201021268  201021269  201021271  201020757  201021264  201021272  100190530  100190531  100190532  100190537  100190786  201021266  201021475  201021478  201021479  201021480  100190785  100190529  100190533  100191248  100191320  100192523  201021473  201021474  201021487  100191242  100191243  100191244  100191245  100191246  100191247  100191249  100191250  100191251  100191252  100191253  100191254  100191255  100191256  100191257  100191259  100191321  100191324  100191325  100191326  100192522  100190534  100190535  100190536  100190780  100190782  100190783  100191323  100192526  100192524  100192525  100192531  100193335  100193336  100190538  100190781  100190784  100190787  100191322  100192527  100192528  100192529  100192530  100192533  100192535  100193320  100193321  100193323  100193326  100193327  100193329  100193330  100193331  100193332  100193333  100193334  100191258  100192532  100193322  100193324  100193325  100193337  100193339  100193341  100193342  100193343  100194369  100193340  100193344  201021476  201021477  201021481  201021482  201021483  201021484  201021485  201021486  100190539  201021270  100195367  100195384  100194425  100195300  100195370  100195372  100195373  100195378  100195379  100195387  201022774  201022779  201023060  201023063  100194376  100195309  100195375  201023061  100190788  100194372  100194375  100194377  100194420  100194421  100195303  100195376  100195382  100195385  201023062  201023069  100192534  100194374  100194378  100194422  100194423  100194427  100194428  100194432  100195301  100195302  100195304  100195307  100195368  100195369  100195371  100195374  100195377  100195380  100195381  201022776  100194368  100194373  100194379  100194424  100194426  100194429  100195308  201023066  201023070  201023071  201023072  100194431  100195383  201022773  201022778  201023065  201023074  201023076  201023440  201023447  201023449  100194371  100194430  201023064  201023067  201023068  201023075  201023077  201023439  201023443  201023448  100194370  201023073  201023445  201023450  201023451  201023441  201023454  201023457  201022775  201022777  201023442  201023444  201023452  201023453  201023455  201023459  201023460  100193338  100195306  201023446  201023456  201023458  201023461  201024840  201024841  201024842  201024845  201024846  201024847  201024848  201024849  201024850  201024851  201024844  201024852  201024853  201025424    201026081  201026084  201024843  201024854  201024855  201024856  201024857  201024858  201024859  201025408  201025409  201025410  201025411  201025412  201025413  201025415  201025416  201025425  201025429  201025430  201025431  201025433  201025434  201025435  201025436  201025437  201025839  201026082  201026083  201026085  201026086  201026087  201026088  201026089  201026090  201026093  201025427  201026080  201026092  201026094  201026095  201026097  201026098  201026099  201026100  201026484  201026482  201026483  201026487  201025414  201026091  201026102  201026480  201026481  201026491  201026492  201026494  201026103  201026485  201026488  201026490  201026101  201025426  201026500  201026677  201025417  201025418  201025419  201025422  201025423  201025428  201026489  201026495  201026497  201026498  201026502  201026674  201026675  201026676  201026678  201027007  201027009  201027011  201027012  201027014  201027015  201027016  201027017  201027023  100155777  100155778  100155779  100156203  100160499  100155776  100160660  100176804  100176806  100176815  100176816  100176818  100176819  201025420  201026486  201026493  201026499  201026501  201026503  201026504  201026505  201026506  201026507  201026508  201026509  201027004  201027005  201027006  201027008  201027010  201027018  201027459  201027460  201027461  201027462  201027022  201027028  201027724  201027727  201026496  201027013  201027020  201027025  201027027  201027456  201027457  201027458  201027464  201027721  201027723  201027725  201027726  201027728  201027729  201027720  201026096  201027024  201027465  201027854  201028145  201027679  201027732  201027853  201027858  201027859  201027861  201027862  201028156  201028157  201027463  201027846  201027852  201028142  201025421  201028147  201028440  201027849  201028161  201028443  201028447  201027733  201027848  201028144  201027860  201027026  201027842  201027730  201027019  201027731  201027734  201027739  201027840  201027843  201027844  201027845  201027847  201027850  201027856  201027857  201028148  201028149  201028150  201028154  201028155  201028159  201028450  201028454  201028457  201028461  201028462  201028463  201028588  201027021  201028446  201028458  201028460  201027865  201028146  201028151  201028152  201028153  201028162  201028163  201028439  201028442  201028444  201028445  201028451  201028453  201028455  201028456  201028141  201028158  201028448  201028449  201027867  201028140  201028143  201028160  201028441  201028459  201028586  201028587  201028591  201027738  201027851  201028452  201028592  201028595  201028596  201028597  201028598  201028747  201028599  201028744  201028745  201030354  201030355  201028742  201030358  201030600  201030602  201028741    201028746      201028750  201030357  201030359  201030601  201030604  201030606  201030608  201030611  201030607  201030610  201028740  201030614  201031225  201031229  201031230  201031231  201030603  201030605  201030609  201030615  201031224  201031228  201031233  201031234  201031236  201031443  201031227  201031232  201031235  201031239  201032045  201031237  201031238  201031441  201031442  201032046  201032047  201032048  201031440  201032044  201032291  201032293  201032442  201032443  201028589  201028590  201028593  201028594  201032151  201032152  201032290  201032309  201032440  100218839  100218840  100218841  201032153  201032294  201032298  201032311  201032441  201032445  201032447  100218818  100218820  100218822  100218823  100218824  100218825  100218826  100218828  100218829  100219117  100219119  100219120  100219121  201031226  201032049  201032050  201032051  201032052  201032053  201032147  201032148  201032149  201032150  201032289  201032296  201032297  201032310  201032312  201032313  201032315  201032316  201032444  100218821  100218836  100218817  100219037  100219322  100219325  100219332  100220764  100220765  100220768  201032777  201032778  201032844  100219036  201032779  201032780  100218837  100219038  100219327  100219330  100219331  100220766  100220767  100220769  100220770  100220771  201032446  201032781  201032782  100218827  100219031  100219034  100219320  100219321  100220775  100220780  100220784  100220787  201032314  100218835  100219030  100219033  100219122  100219329  100220773  100220781  100221889  100221899  100218833  100220788  100221892  100222822  100218834  100219028  100221900  100222838  100221908  100222836  100222841  100224287  100224288  100222833  100222840  100225758  100224294  100224293  100224296  100225757  100221890  100221896  100221898  100222821  100222831    100227938  100219029  100220779  100221895  100221902  100221910  100222638  100219323  100220772  100221893  100218831  100218832  100218838  100219032  100219035  100219039  100219326  100220776  100220777  100220778  100220782  100220783  100220785  100221891  100221894  100221897  100221901  100221903  100221909  100221911  100221912  100222639  100222817  100222818  100222819  100222826  100222832  100224295  100225761  100226491  100226495  100226496  100226499  100226520    201032950  201032951  201032952  201032953  100226494  100227972  100227974  100226493  100220786  100229271  100229272  100229273  100229274  100229275  100229277  100229278  100229279  201032295  100229276  100229280  100229281  100229312  100229313  100229314  100229315  100229316  100229317  100229318  100229319  100229441  100229442  100229444  100229445  100229440  100229443  100229446  100230807  100230808  100230809  100230811  100230812  100230814  100230815  100230816  100230818  100230856  100231124  100230810  100230817  100230821  100230859  100231121  100231122  100231123  100231126  100231417  100231134  100231416  100231419  100230820  100231128  100231132  100231133  100231462  100231463  100231664  100231665  100231666  100231667  100231668  100231669  100230813  100230819  100230857  100231127  100231129  100231130  100231131  100229266  100231125  100231415  100229267  100231671  100231784  100227960  100231697  100231698  100231780  100232322  201033110  100222823  100229268  100224291  201033135  100222637  100227970  100229269  201033118  100222824  100222834  100224290  100226498  100227964  100227973  201033117  100225756  100225762  201033115  100221913  100222820  100224297  100227965  100227977  100219328  100222635  100222825  100222828  100222829  100222837  100224292  100225755  100225759  100225763  100226492  100227937  100227961  100227963  100227966  100227967  100230855  100230858  100231418  100231460  100231461  100231464  100231670  201033106  201033107  201033108  201033114  201033116  100222830  100222835  100227962  100227978  100222634  100222636  100227976  100229270  100227975  201033109  100221906  100231785  100232723  100231699  100232329  100232721  100232735  201033908  201034043  100231783  100231786  100232321  100232324  100232325  100232326  100232327  100232328  100232330  100232720  100232722  100232727  100232728  100232729  100232733  100232736  201033796  201033797  201033911  201033912  201033913  201033914  201034049  201034050  201034052  100232323  100232724  100232725  100232726  100232730  100232731  100232734  201033794  100232732  100238558  100238559  100238588  100238593  100238761  201033795  201033798  201033799  201033907  201033909  201033910  201033915  201033916  201034040  201034041  201034042  201034051  201034053  201034092  201034093  201034094  201034095  201034097  201034098  201034099  100221904  100231672  100238580  100238581  100238582  100238583  100238584  100238585  100238586  100238587  100238589  100238590  100238591  100238592  100238594  100238595  100238596  100238597  100238763  201034096  201034160  201034161  100225764  100238765  100238768  100238762  100238766  100238767  100238769  100238770  100238931  100238932  100238933  100238934  100238935  100238936  100238937  100238938  100238939  100239262  100239264  100239265  100239266  100239270  100239272  100239263  100239269  100239271  100239274  100239519  100239795  100239796  100239798  100239799  100240120  100240121  100238764  100239260  100239261  100239275  100239517  100240140  100240141  100239797  100240122  100240123  100240143  100239267  100240079  100240144  100239268  100239273  100239518  100240073  100240074  100240075  100240076  100240078  100240142  100241293  100241296  100241292  100241294  100241295  100241297  100241299  100241300  100241301  100241764  100240077  100241298  100241762  100241763  100241769  100241768  100241765  100241767  100241771  100241766  100241770  100242962  100242963  100242969  100242971  100242972  100242973  100243770  100243771  100243774  100243775  100243776  100243778  100244319  100242970  100243779  100244320  100244323  100244325  100244326  100244327  100244328  100244329  100244331  100243777  100243840  100244318  100244321  100244322  100244324  100244333  100244334  100244335  100244337  100244330  100244332  100244914  100244915  100244916  100244917  100244919  100244336  100244918  100244920  100244921  100244922  100244923  100244924  100244925  100244926  100244927  100244928  100244929  100244930  100244932  100244933  100245753  100245754  100245755  100243772  100244931  100245757  100245756  100245758  100245759  100245760  100245761  100245762  100246942  100246944  100246945  100246947  100246946  100246943  100246948  100246949  100246950  100246951  100246952  100246954  100246955  100246956  100246957  100246958  100246959  100246961  100247849  100247850  100247851  100247853  100247855  100246953  100246960  100247852  100247854  100247856  100247857  100247858  100247859  100247861  100247864  100247865  100247866  100247867  100248328  100248331  100248333  100247868  100247863  100248324  100248325  100248327  100248329  100248330  100248332  100248335  100248337  100248338  100248336  100248341  100248342  100248343  100248344  100248345  100248346  100248347  100248348  100249203  100247862  100249206  100249208  100248326  100248340  100249204  100249207  100249210  100249212  100249213  100249802  100248339  100249209  100249211  100249214  100249215  100249216  100249217  100249801  100249803  100249804  100249805  100249806  100249807  100249205  100250065  100250066  100250069  100250070  100250071  100250073  100250067  100250068  100250072  100250074  100250076    100250077  100250079  100250481  100250483  100250486  100250078  100250482  100250484  100250485  100249809  100249815  100249816  100250487  100249800  100249808  100249810  100249811  100249812  100249813  100249814  100249817  100249818  100249819  100250064  100250488  100250489  100250490  100251604  100251607  100251608  100251610  100251611  100251612  100251614  100251603  100251606  100253109  100253110  100253113  100253114  100253124  100251609  100251615  100253115  100253116  100253119  100253126  100253323  100251613  100253111  100253125  100253299  100251605  100253117  100253118  100253320  100253330  100253333  100253336  100253340  100253329  100253321  100253335  100253338  100253128  100253327  100253121  100253112  100253325  100253332  100253341  100253322  100253324        100253123  100253127        100255957  100257488  100257490  100257496  100257500  100258060  100257491  100253337  100255874  100255875  100257487  100257493  100257495  100257497  100257498  100257499  100258061  100258063  100258065  100258066  100258067  100258068  100257494  100257501  100258064  100260223  100260220  100260222  100260616  100260224  100260625  100260221  100260227  100260618  100260619  100260620  100260621  100260622  100260623  100260624  100260626  100260627  100260628  100260629  100260630  100260631  100260632  100260633  100260634  100260635  100261843  100261844  100261845  100261846  100261847  100261848  100261849  100261850  100261851  100261852  100261853  100261854  100261855  100261856  100261857  100261900  100261901  100261902  100260228  100260229  100260617  100263995  100263996  100263997  100264020  100264023  100264024  100264026  100264279  100264022  100264025  100263992  100263993  100263994  100263998  100263999  100264021  100264300  100264301  100264302  100264303  100264561  100264562  100264563  100264564  100264565  100264567  100264571  100264566  100264568  100264569  100264570  100264572  100264573  100264574  100264575  100264576  100264780  100264785  100264577  100264579  100264781  100264578  100265625  100265629  100265630  100265634  100265639  100265642  100264782  100265631  100265635  100265637  201035019  201035334  201035360  201035224  100265626  100265627  100265638  100265641  201034919  201035085  201034950  201034946  201035197  201035222  201035226  201035225  201035324  201035331  201035362  201034917  100265632  100265633  201034945  201035015  201035198  201035199  201035221  201035320  201035321  201035333  201034949  201035330  201035016  201035017  201035081  201035084  201035337  201035082  201035339  201035363  201034918  201035223  201035323  100264784  100265636  201035018  201034951  201034952  201034954  201035322  201035328  201035335  201035336  201035338  201035365  201035329  100265624  100265628  100265640  100265643  100265644  100265645  100276770  100276771  100276772  100276774  100276776  100276777  100276778  100276779  100276780  100276784  100276786  100276788  100276789  100277103  100277104  100277105  100277106  100277107  100277108  100277109  100277111  100277115  100277116  100277118  100277120  201034947  100276773  100276775  100277117  100277121  100276781  100276782  100277122  100277127  100276785  100277102  100277112  100277113  100277114  100277119  100277123  100277124  100277125  100277128  100277129  100277131  100277164  100277165  100277166  100277167  100277168  100277169  100277170  100277171  100277172  100277173  100277174  100277175  100277176  100277178  100277320  100277321  100277322  100277323  100277327  100277328  100277329  100277330  100277613  100277616  100276787  100277110  100277126  100277612  100277614  100277615  100277618  100277619  100277620  100277177  100277325  100277326  100277333  100276783  100277130  100277331  100277332  100277623  100277624  201035083  201035332  201035361  100264783  100279988  100279987  100279992  100281272  100281460  100277179  100277324  100277617  100277622  100277625  100277626  100279971  100279972  100279977  100279978  100279976  100281271  100281273  100281276  201035080  100277628  100277631  100279973  100279975  100279979  100279981  100279982  100279983  100279984  100279985  100279986  100279989  100279990  100279991  100279993  100279996  100279997  100279998  100279999  100281461  201037287  100279994  100281278  100280000  201037311  201037293  201037307  100277627  100281279  201037279  201037286  201037308  201037380  201037292  100281277  201037272  201037273  201037278  201037285  201037288  201037289  201037290  201037291  201037310  201037312  201037381  201037405  201037407  201037409  201037410  201037271  201037404  201037408  100277621  100277629  201037274  201037275  201037276  201037284  201037309  201037382      100277630  201037277  100279974  100288469  100288471  100288472  100288473  100288474  100288475  100288476  100288477  100288478  100288479  100288580  100288582  100288583  100288585  100288586  100288587  100288588  100289770  100289771  100289772  100289773  100289776  100289779  201037313  100288584  100289774  100289777  100289778  100289781  100290714  100290719  100289768  100289775  100289782  100289783  100290710  100290711  100290712  100290713  100290718  100290720  100290721  100290723  100290715  100290716  100290717  100290726  100290724  100290725  100291598  100291627  100289769  100290728  100291621  100291630  100291631  100292461  100290727  100292455  100292456  100292464  100292465  100292466  100292469  100292721  100291620  100291624  100291626  100292462  100292467  100292720  100290729  100291599  100291629  100291633  100291634  100291637  100292458  100292459  100292468  100291622  100291628  100291632  100291623  100291625  100292460  100292463  100292719  100296195        100296194                    100292723  100296198  100296858  100296860  100296864  100297020  100297026  100297028  100297029  100296199  100296197  100296851  100296854  100296855  100297021  100296853  100298489  100298488  100298490  100298491  100296865  100297626  100298500  100298526  100297033  100297024  100297030  100296857  100297022  100298486  100298525  100297630  100298482  100297031  100297628  100297631  100297633  100297634  100297635  100298496  100298527  100298529  100298532  100298533  100298535  100298620  100301143  100298485  100298487  100298530  100298484  100298536  100301086  100298531  100297025  100296196  100301088  100301089  100301090  100301097  100301099  100301093  1003",
      "center_classification_date": "20210122",
      "distribution_pattern": "Worldwide distribution - US Nationwide including Guam and Puerto Rico;  Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia.",
      "state": "CA",
      "product_description": "Philips Respironics V680 Ventilator Part Number 850011 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.",
      "report_date": "20210203",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Respironics California, LLC",
      "recall_number": "Z-0917-2021",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "86697",
      "more_code_info": "02050  100302051  100301095  100302052  100302054  100301092  100301094  100301141  100301142  100297027  100297627  100298534    100301087  100301091  100301096  100301140  100298539  100301144  100302049  100304941  100302057  100302058  100302059  100304602  100305132  100305134  100298481  100301085  100302055    100301098  100302060  100304601  100304942  100304976  100304977  100304978  100304979  100305133  100305340  100305342  100305345  100305346  100305347  100305348  100305351  100305352  100305353  100306068  100305349  100306061  100306064  100306070  100302053  100305341  100305343  100305344  100305350  100305354  100305939  100306060  100306062  100306063  100306065  100306066  100306067  100306069  100306071  100306072  100306073  100306074  100306075  100307211  100307213  100307214  100307215  100307216  100307217  100307218  100307541  100307545  100308426  100308428  100308431  100308432  100308433  100308435  100308436  100308437  100308439  100308440  100308517  100308600  100308602  100308605  100307219  100307540  100307544  100308424  100308429    100308434  100308441  100308518  100308609  100307212  100307542  100307546  100307547  100307557  100308425  100308427  100307210  100307548  100308601  100308607  100308608  100308612  100310181  100310182  100310183  100308604  100289780  100290722  100291635  100292457  100296869  100307543  100308438  100308519  100308603  100308606  100308610  100308611  100308613  100308614  100309856  100309857  100309859  100310180  100310184  100310186  100310187  100310188  100310189  100310190  100310191  100310192  100311610  100311611  100311612  100311615  100311619  100312022  100312023  100312024  100312025  100312196  100312026  100312027  100311614  100312020  100312193  100312195  100312191  100312194  100309858  100312192  100309855  100312021  100312320  100312197  100312198  100311616  100311618  100314606  100314860  100314863  100314864  100314866  100314867  100314870  100314874  100314613  100311613  100314607  100314608  100314609  100314611  100314614  100314617  100314862  100314868  100314872  100314610  100314612  100314615  100311617  100314618  100314865  100315000  100315001  100315002  100315004  100315005  100316027  100316035  100316036  100316037  100314616  100314759  100314861  100314871  100314873  100316029  100316031  100316032  100316033  100314869  100315003  100316209  100316215  100316218  100316221  100316028  100316034  100316205  100316212  100316216  100316217  201038853  201038855  201038858  201038951    201038953  201038950  201038940  201038949  201038854  201038943  201038945  201038946  100316200  100316211  201038850  201038851  201038852  201038859  201038948  100314619  100316030  100316038  100316203  100316204  100316213  100316224  100316202  100316206  100316207  100316208  100316214  100316219  100316220  100316222  100316223  100316225  100316227  100324359  100324460  201038849  201038857  201038941  201038947  201038954  201038955  201038956  201038957  100324351  100324354  100316229  100324353  100324356  100324357  100324461          100324464  100324466  100324467  100324469  100328720  100324470  100328674  100328675  100328676  100328715  100328716  100328717  100328718  100328721  100328761  100328781  100328782  100328783  100329188  100328719  100328780  100328784  100329191  100329192  100329424  100329428  100329437  100329612  100329196  100329433  100324358  100324468  100328677  100328678  100328722  100329187  100329420  100329609  100329614  100329617  100328699  100329434  100329079  100329611  100328679  100329422  100329425  100329431  100328698  100329421  100329423  100329615  100329430  100324352  100324465  100329190  100329194  100329195  100329426  100329435  100329438  100329607  100324463  100329189  100329427  100329613  100329616  100329429  100329436  100329608  100329610  100329618  100329620  100329621  100346820  100346821  100346822  100346823  100346824  100346825  100346826  100346827  100346828  100346829  100346830  100346831  100346832  100346833  100346835  100346836  100346837  100346838  100346839  100346840  100346841  100346842  100346844  100347194  100347195  100347196  100347197  100347198  100347199  100347200  100347204  100347205  100347207  100347208  100347209  100347328  100347329  100347332  100347337  100347338  100347339  100347340  100347342  100347343  100347344  100347347  100347350  100347356  100347359  100347360  100347361  100347362  100347367  100347368  100347369  100347370          100329619  100346834  100346843  100347201  100347202  100347206  100347330  100347331  100347333  100347334  100347341  100347345  100347346  100347348  100347351  100347352  100347353  100347354  100347355  100347357  100347358  100347364  100347365  100347366  100347371  100347203",
      "recall_initiation_date": "20201102",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}