{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Largo",
      "state": "FL",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64367",
      "recalling_firm": "Linvatec Corp. dba ConMed Linvatec",
      "address_1": "11311 Concept Blvd",
      "address_2": "N/A",
      "postal_code": "33773-4908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the states of: CA, FL, GA, IL, KS, MA, MD, MO, NE, NY, and TX and the countries of: Canada, Germany, Romania, Spain, Palestine, South Africa, France, Sweden and Singapore",
      "recall_number": "Z-0916-2013",
      "product_description": "D3000 Advantage Drive System, 115v     Product Usage:  The Advantage Drive Systems function as a powered instrument system consisting of handpieces and accessories to perform cutting of soft tissue and bone",
      "product_quantity": "31 units",
      "reason_for_recall": "The Membrane Switch Panel (referred to as \"front panel\" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment: General requirements for basic safety and essential performance.",
      "recall_initiation_date": "20130124",
      "center_classification_date": "20130306",
      "termination_date": "20130705",
      "report_date": "20130313",
      "code_info": "Model #D3000"
    }
  ]
}