      "reason_for_recall": "Medical device adjustment knob may be loose and affect the accuracy of blended Air/Oxygen gas mixture to patients.",
      "address_2": "",
      "product_quantity": "1970 units",
      "code_info": "Serial Numbers:  HDJR52093  HDJQ52878  HDJS51010  HDJP51983  HDJN52273  HDJN52274  HDJN52275  HDJN52276  HDJP50604  HDJP50605  HDJM50388  HDJR56613  HDJQ52689  HDJR50734  HDJR50735  HDJS50028  HDJP52371  HDJP52558  HDJN51693  HDJN51694  HDJR56618  HDJQ51960  HDJS52390  HDJM50693  HDJM51714  HDJM51716  HDJM51717  HDJM51718  HDJM52096  HDJN52193  HDJN52194  HDJN52195  HDJN52196  HDJN52197  HDJP52562  HDJP52563  HDJP52564  HDJP52565  HDJQ52920  HDJQ52917  HDJQ52918  HDJQ52919  HDJQ52921  HDJQ52922  HDJQ52923  HDJQ52924  HDJQ52925  HDJQ52926  HDJQ52927  HDJQ53075  HDJR56481  HDJR56482  HDJR56486  HDJR56487  HDJR56488  HDJR56489  HDJR56490  HDJR56491  HDJR56358  HDJP50531  HDJP50532  HDJR56107  HDJR56370  HDJR56669  HDJR56746  HDJR56747  HDJR56748  HDJM51930  HDJM52089  HDJM51405  HDJM51470  HDJM51471  HDJM51472  HDJM51475  HDJM51847  HDJS51291  HDJS51292  HDJR54229  HDJQ51474  HDJQ51475  HDJQ51476  HDJM52059  HDJM50272  HDJP50634  HDJQ51827  HDJR50477  HDJR57730  HDJM50483  HDJQ52812  HDJS52096  HDJR57172  HDJM50694  HDJN51776  HDJN51777  HDJN51778  HDJQ51257  HDJR56005  HDJR56006  HDJR56007  HDJR56008  HDJQ52597  HDJQ52598  HDJQ52599  HDJQ52600  HDJQ52601  HDJN50850  HDJM50704  HDJM50705  HDJM50706  HDJM50707  HDJM50708  HDJM50709  HDJQ53073  HDJR57718  HDJP50352  HDJM50012  HDJP51078  HDJP51079  HDJP51080  HDJQ51561  HDJQ51562  HDJQ51563  HDJR56749  HDJR56750  HDJR56751  HDJR56752  HDJR56753  HDJS50790  HDJS50791  HDJS50792  HDJR57719  HDJR57720  HDJR57721  HDJQ50915  HDJQ50916  HDJQ50917  HDJQ50918  HDJQ50919  HDJS51692  HDJS51693  HDJM51900  HDJM52091  HDJM52092  HDJM52093  HDJS50046  HDJS50047  HDJS50048  HDJR50230  HDJR50231  HDJS50297  HDJP51783  HDJP51784  HDJP51792  HDJP51793  HDJP51796  HDJQ52082  HDJS50733  HDJS50734  HDJS50735  HDJP52863  HDJR50596  HDJR50597  HDJR50598  HDJN50595  HDJM50200  HDJM50201  HDJP52156  HDJP52071  HDJP52072  HDJQ52813  HDJQ52814  HDJQ52815  HDJQ52895  HDJS51414  HDJS51415  HDJS51417  HDJM51320  HDJM51321  HDJM51883  HDJN50762  HDJN50763  HDJN51327  HDJS50042  HDJS50043  HDJR50133  HDJR50134  HDJR50135  HDJR50136  HDJR50138  HDJR50139  HDJR50140  HDJR50141  HDJR50142  HDJR50144  HDJR50145  HDJR50146  HDJR50147  HDJR50148  HDJR50149  HDJR50150  HDJR50151  HDJR50152  HDJR50153  HDJR50154  HDJR50155  HDJR50156  HDJR50158  HDJR50159  HDJR50161  HDJR50143  HDJR50157  HDJR50160  HDJR50299  HDJL50161  HDJL50162  HDJL50163  HDJL50164  HDJL50165  HDJR50298  HDJS51009  HDJP50561  HDJP50562  HDJP50563  HDJP50564  HDJP50565  HDJP50566  HDJQ52690  HDJS50002  HDJS50003  HDJS50004  HDJS50005  HDJS50006  HDJS50007  HDJL50357  HDJL50358  HDJL50360  HDJL50362  HDJL50359  HDJL50383  HDJR52099  HDJR56206  HDJN51948  HDJN51949  HDJN52121  HDJN52122  HDJP52630  HDJP51794  HDJP51795  HDJP52794  HDJP52795  HDJN52437  HDJN52438  HDJN52439  HDJS52720  HDJS52721  HDJR54274  HDJR54275  HDJR52095  HDJR54268  HDJR54269  HDJR54270  HDJR54271  HDJR54272  HDJM50005  HDJM50006  HDJM50007  HDJS51006  HDJS51002  HDJS51003  HDJS51004  HDJS51005  HDJN52376  HDJP50730  HDJR56133  HDJM51307  HDJS51426  HDJS51427  HDJS51458  HDJP51526  HDJP51527  HDJP51528  HDJP52796  HDJP52797  HDJP52798  HDJR56368  HDJR57717  HDJR56207  HDJL50432  HDJL50433  HDJL50434  HDJL50435  HDJQ51484  HDJQ51485  HDJQ51486  HDJQ51487  HDJQ51488  HDJQ51489  HDJQ51490  HDJM51308  HDJP50177  HDJR56361  HDJP51389  HDJP51390  HDJP51391  HDJN50262  HDJN50263  HDJN50264  HDJN50265  HDJP52079  HDJP52080  HDJP52081  HDJP52082  HDJN50541  HDJN50542  HDJN50543  HDJN50544  HDJN50545  HDJN50546  HDJN50547  HDJN50298  HDJN50299  HDJR54214  HDJR54215  HDJR54216  HDJQ50580  HDJQ50581  HDJQ50582  HDJQ50583  HDJQ50584  HDJQ50585  HDJM50056  HDJM50178  HDJM50583  HDJM51924  HDJM52339  HDJM50651  HDJR52092  HDJR52098  HDJM50058  HDJM50097  HDJN51392  HDJR58028  HDJR58029  HDJR58030  HDJR58031  HDJM50142  HDJQ50388  HDJQ50389  HDJQ53023  HDJQ53024  HDJQ53025  HDJQ53026  HDJQ53027  HDJQ53028  HDJQ53029  HDJQ53030  HDJQ53031  HDJN51333  HDJN51855  HDJN51856  HDJN51857  HDJN51858  HDJN51859  HDJN51692  HDJN50786  HDJN50785  HDJN50787  HDJN50788  HDJQ52080  HDJQ52081  HDJR57435  HDJR57436  HDJR57437  HDJR57438  HDJR57439  HDJR57751  HDJS51416  HDJS51554  HDJS51555  HDJS51556  HDJQ51789  HDJQ51792  HDJM50343  HDJP52606  HDJP52607  HDJM51449  HDJM50903  HDJM50904  HDJM51596  HDJM51597  HDJM51598  HDJM51599  HDJM51759  HDJM51760  HDJM51340  HDJP51984  HDJP51985  HDJP51986  HDJP51987  HDJR54144  HDJQ50029  HDJQ50030  HDJQ50031  HDJQ50032  HDJQ50033  HDJQ50950  HDJR50359  HDJM50199  HDJR56624  HDJR56625  HDJM50867  HDJM50868  HDJM50869  HDJM50870  HDJM50871  HDJM50872  HDJN51772  HDJN51811  HDJN51812  HDJS50702  HDJS51058  HDJS51059  HDJS51060  HDJS51061  HDJS51062  HDJS51063  HDJS51064  HDJR56744  HDJR56745  HDJP52358  HDJM50696  HDJQ52896  HDJS50294  HDJS50295  HDJS50296  HDJN50681  HDJN50680  HDJM50698  HDJN51945  HDJP51921  HDJP51922  HDJP51923  HDJP51924  HDJP51925  HDJN51774  HDJN51775  HDJP51521  HDJP51522  HDJR52096  HDJS52119  HDJS52120  HDJQ51991  HDJM51083  HDJM51896  HDJM52134  HDJM52150  HDJM52151  HDJM52152  HDJM52153  HDJM52154  HDJM52155  HDJP52450  HDJP52451  HDJN50207  HDJN50208  HDJP52083  HDJP52084  HDJP52085  HDJP52086  HDJP52087  HDJP52088  HDJP52089  HDJP52090  HDJP52091  HDJP52092  HDJP52093  HDJP52094  HDJP52095  HDJP52096  HDJP52097  HDJP52098  HDJP52099  HDJP52100  HDJP52101  HDJP52102  HDJR56483  HDJP51566  HDJP50179  HDJQ51046  HDJQ51047  HDJQ50439  HDJQ52816  HDJR50265  HDJR50266  HDJR50267  HDJR50268  HDJR50269  HDJR50270  HDJR50271  HDJR50272  HDJN51147  HDJN51148  HDJN51146  HDJR56262  HDJR56263  HDJM52103  HDJM52104  HDJN51301  HDJM52105  HDJN51526  HDJN51527  HDJN51528  HDJN51529  HDJN51530  HDJN51531  HDJN51532  HDJN51533  HDJN51534  HDJN51535  HDJN51536  HDJN51537  HDJN51538  HDJN51539  HDJN51540  HDJN51541  HDJN51542  HDJN51543  HDJN51544  HDJN51545  HDJN51546  HDJN51547  HDJN51548  HDJN51549  HDJN51550  HDJN51551  HDJN51552  HDJN51553  HDJN51554  HDJN51555  HDJN51556  HDJN51557  HDJN51558  HDJN51559  HDJN51560  HDJN51561  HDJN51562  HDJN51563  HDJN51564  HDJN51565  HDJN51566  HDJN51567  HDJN51568  HDJN51569  HDJN51570  HDJL50363  HDJN52489  HDJN52490  HDJP50858  HDJQ50818  HDJQ52271  HDJS50635  HDJS51324  HDJR52097  HDJM51076  HDJM51077  HDJM51078  HDJS51378  HDJS51379  HDJL50285  HDJL50496  HDJL50497  HDJL50498  HDJL50499  HDJL50500  HDJL50501  HDJL50502  HDJL50503  HDJL50504  HDJM52095  HDJM52097  HDJN50745  HDJN50899  HDJN50900  HDJP51230  HDJP51231  HDJP51232  HDJP51233  HDJP51235  HDJP51236  HDJP51234  HDJP51237  HDJP51238  HDJP51239  HDJP51240  HDJP51241  HDJP52073  HDJP52074  HDJP52075  HDJP52076  HDJP52077  HDJP52078  HDJP52355  HDJP52357  HDJP52734  HDJP52735  HDJN51745  HDJN51746  HDJN51747  HDJS50317  HDJS50318  HDJN50266  HDJP51634  HDJP51635  HDJP51636  HDJP51637  HDJS50536  HDJR57238  HDJN50487  HDJN50488  HDJQ51230  HDJR50707  HDJM50046  HDJM50527  HDJM50955  HDJL50317  HDJL50356  HDJM51066  HDJM51067  HDJM51068  HDJM51070  HDJM51071  HDJM51072  HDJM51073  HDJM51069  HDJM50188  HDJN50597  HDJN50746  HDJN51273  HDJN51274  HDJN51328  HDJR56357  HDJR56379  HDJM51061  HDJM51084  HDJL50278  HDJN52174  HDJN52175  HDJN52176  HDJN52177  HDJN52178  HDJM50673  HDJM50674  HDJR57902  HDJR57903  HDJS51733  HDJS51734  HDJM51406  HDJM51474  HDJQ50384  HDJR57561  HDJR57562  HDJQ51338  HDJQ51339  HDJL50237  HDJL50238  HDJL50239  HDJL50241  HDJL50242  HDJN51779  HDJN51780  HDJN51781  HDJN51782  HDJN51783  HDJR56110  HDJR56111  HDJR56112  HDJR56113  HDJR56114  HDJR56115  HDJR56116  HDJR56117  HDJR56118  HDJR56119  HDJR56120  HDJR56121  HDJR56122  HDJR56123  HDJP52736  HDJP52737  HDJP52738  HDJP52739  HDJP52740  HDJP52741  HDJP52742  HDJP52743  HDJP52744  HDJP52745  HDJQ53048  HDJQ53049  HDJQ53050  HDJL50240  HDJL50243  HDJL50244  HDJL50245  HDJM50063  HDJM51079  HDJM51080  HDJM52050  HDJN52322  HDJN52323  HDJL50521  HDJL50522  HDJL50523  HDJL50491  HDJL50493  HDJL50494  HDJL50517  HDJQ51992  HDJQ52894  HDJQ52903  HDJQ52904  HDJQ52905  HDJQ52907  HDJQ52908  HDJQ52909  HDJQ52910  HDJQ52629  HDJS50899  HDJN51149  HDJN51150  HDJN51151  HDJN51152  HDJR57307  HDJR57308  HDJP50601  HDJP50606  HDJQ52897  HDJQ52108  HDJQ52109  HDJQ52110  HDJQ52111  HDJQ52112  HDJQ52113  HDJQ52877  HDJR56205  HDJR56371  HDJR56372  HDJR56373  HDJR56374  HDJR56667  HDJR56668  HDJR56369  HDJP50216  HDJP50217  HDJP50218  HDJP50219  HDJP50220  HDJP50221  HDJP50222  HDJP50224  HDJP50223  HDJQ51609  HDJQ51610  HDJQ51611  HDJP50635  HDJS50199  HDJS52092  HDJS52093  HDJM51420  HDJM51422  HDJM50436  HDJM50437  HDJM50438  HDJM51438  HDJM51439  HDJM51440  HDJM50580  HDJM50581  HDJM50582  HDJM51682  HDJM51683  HDJM51715  HDJR57968  HDJR57969  HDJR57970  HDJR57971  HDJM50480  HDJM50481  HDJM50482  HDJR57729  HDJQ53076  HDJL50364  HDJN51217  HDJN51218  HDJS52118  HDJQ51343  HDJQ51344  HDJQ51345  HDJQ51346  HDJN51854  HDJP50146  HDJP50178  HDJN51744  HDJM50288  HDJM50289  HDJN51698  HDJR50137  HDJN50593  HDJN50594  HDJR57565  HDJR57566  HDJR57567  HDJR57568  HDJR57569  HDJR57563  HDJR57564  HDJM50695  HDJS50694  HDJS50696  HDJS50736  HDJS50737  HDJM50504  HDJS51325  HDJS51278  HDJS51279  HDJN51210  HDJS52737  HDJS52738  HDJS52739  HDJS52740  HDJM52106  HDJM50536  HDJM50537  HDJQ52617  HDJP50650  HDJP50651  HDJP50652  HDJR50510  HDJR50511  HDJR50512  HDJR50513  HDJR50514  HDJR50515  HDJQ50539  HDJQ52202  HDJQ52203  HDJM51184  HDJM51185  HDJM51186  HDJM51187  HDJM50538  HDJM50956  HDJM50957  HDJM50584  HDJN52351  HDJM51062  HDJM51063  HDJR56626  HDJR56627  HDJR56628  HDJS52715  HDJS52716  HDJQ52906  HDJR57440  HDJP51827  HDJP51828  HDJP51900  HDJM50008  HDJM50009  HDJM50010  HDJM50011  HDJM50421  HDJM50422  HDJP50354  HDJR54218  HDJR58033  HDJR58034  HDJR58035  HDJR58036  HDJM51419  HDJM51421  HDJM51423  HDJM51448  HDJR54219  HDJN50167  HDJN50168  HDJN50169  HDJN50170  HDJR57173  HDJL50195  HDJL50196  HDJL50197  HDJL50198  HDJL50199  HDJL50200  HDJL50201  HDJM51252  HDJM51257  HDJM51259  HDJM51253  HDJM51254  HDJM51255  HDJM51256  HDJM51258  HDJL50422  HDJL50423  HDJN52189  HDJN52190  HDJN52192  HDJN52191  HDJN50237  HDJN50238  HDJN50239  HDJQ50961  HDJQ50962  HDJQ50963  HDJQ50964  HDJQ50965  HDJQ50966  HDJQ50967  HDJQ50968  HDJQ50969  HDJQ50970  HDJQ50971  HDJQ50972  HDJQ50973  HDJQ50974  HDJQ51573  HDJR54164  HDJR54165  HDJR54217  HDJR56714  HDJR52094  HDJM51339  HDJP52018  HDJP52019  HDJP52359  HDJP52363  HDJP52513  HDJP51823  HDJP51824  HDJP51825  HDJP51826  HDJP51829  HDJP51830  HDJP51903  HDJP51904  HDJQ52488  HDJQ52489  HDJS50041  HDJS50616  HDJM50055  HDJM51931  HDJP52809  HDJP52810  HDJQ50153  HDJQ50154  HDJQ50155  HDJP51458  HDJP51459  HDJP51460  HDJQ51306  HDJQ51307  HDJQ51638  HDJQ51639  HDJR52104  HDJR52105  HDJR50765  HDJM51553  HDJM51554  HDJP50353  HDJP50602  HDJS52323  HDJN51946  HDJN51947  HDJQ51634  HDJQ51635  HDJQ51636  HDJQ51637  HDJN51957  HDJP51785  HDJP51786  HDJP51787  HDJP50547  HDJP50548  HDJR50065  HDJR50066  HDJM51794  HDJM52027  HDJM52028  HDJM52029  HDJM52030  HDJM52031  HDJM52032  HDJM52033  HDJM52034  HDJM52035  HDJN51601  HDJR54249  HDJR54250  HDJP51883  HDJP51950  HDJM50059  HDJQ52270  HDJS50159  HDJS50160  HDJN50583  HDJN51435  HDJN51436  HDJP51187  HDJP51188  HDJQ50419  HDJM51884  HDJM51885  HDJM52077  HDJQ51788  HDJQ51790  HDJQ51791  HDJP52233  HDJM51882  HDJS50854  HDJS51825  HDJS50459  HDJS50460  HDJS50461  HDJS50851  HDJS50852  HDJS50853  HDJS50855  HDJS51826  HDJQ50646  HDJQ50642  HDJQ50643  HDJQ50644  HDJQ50645  HDJQ50647  HDJR57725  HDJR57728  HDJP51862  HDJM52098  HDJN51697  HDJS52377  HDJS52378  HDJS52400  HDJN51211  HDJN51212  HDJN51213  HDJN51214  HDJN51215  HDJN51216  HDJM50645  HDJM50646  HDJM50647  HDJM50648  HDJM50649  HDJM50650  HDJN52254  HDJR56484  HDJR56485  HDJQ50385  HDJP51861  HDJP51882  HDJS51724  HDJS51725  HDJS51726  HDJS51727  HDJS51728  HDJS52111  HDJL50489  HDJL50490  HDJL50492  HDJL50518  HDJL50519  HDJL50520  HDJN51816  HDJN51817  HDJN51818  HDJM51848  HDJP52806  HDJP52807  HDJP50233  HDJP50234  HDJQ53173  HDJM50419  HDJM50420  HDJM50697  HDJR56933  HDJR56934  HDJM50386  HDJM50387  HDJM51216  HDJM51217  HDJM51218  HDJM51219  HDJM51220  HDJM51221  HDJM51222  HDJM51223  HDJM51224  HDJM51225  HDJM52221  HDJP50355  HDJP51523  HDJP51524  HDJP51525  HDJQ50951  HDJQ50952  HDJQ50953  HDJR54122  HDJR54123  HDJR56798  HDJP51633  HDJQ51525  HDJR50162  HDJM50187  HDJP50007  HDJN51837  HDJN51838  HDJN51839  HDJN51840  HDJQ50252  HDJQ50253  HDJQ50254  HDJQ50255  HDJQ50256  HDJQ50257  HDJQ50258  HDJQ50259  HDJP51022  HDJP51023  HDJR54166  HDJR54167  HDJR54168  HDJR54169  HDJR54170  HDJR54171  HDJR54172  HDJR54173  HDJN51896  HDJL50505  HDJQ50336  HDJQ50337  HDJQ50338  HDJQ50339  HDJQ50340  HDJQ50341  HDJS50695  HDJS50697  HDJS50738  HDJP51104  HDJP52352  HDJP52353  HDJP52354  HDJP52364  HDJP52626  HDJP52627  HDJP52628  HDJP52629  HDJS51797  HDJM50653  HDJR50225  HDJR50224  HDJP51901  HDJP51902  HDJM52302  HDJM52303  HDJN50376  HDJM52094  HDJM50652  HDJP51190  HDJP51191  HDJP50498  HDJP50499  HDJQ51491  HDJQ51492  HDJQ51493  HDJQ51494  HDJQ51495  HDJR56367  HDJN50815  HDJP51388  HDJS50950  HDJS51732  HDJR58208  HDJR58207  HDJR58209  HDJL50487  HDJN50573  HDJN50767  HDJM50902  HDJQ52058  HDJM50479  HDJP51951  HDJL50229  HDJL50230  HDJL50231  HDJL50232  HDJM50879  HDJM50880  HDJM50881  HDJM50882  HDJM50883  HDJP52655  HDJP52656  HDJR54121  HDJQ50493  HDJQ50494  HDJR50297  HDJN51978  HDJN51979  HDJN51980  HDJN51981  HDJN51982  HDJS50037  HDJM51085  HDJM51473  HDJR50708  HDJS50787  HDJS50788  HDJP50603  HDJR54233  HDJR54234  HDJR54235  HDJM50147  HDJP50000  HDJP50001  HDJQ51342  HDJQ51648  HDJQ51641  HDJM50236  HDJM50237  HDJP52869  HDJR56095  HDJR58243  HDJP51019  HDJP51020  HDJQ50648  HDJS50729  HDJS50730  HDJS50731  HDJS50732  HDJS52425  HDJS52426  HDJS52427  HDJN50888  HDJP50102  HDJQ52272  HDJR57883  HDJL50329  HDJL50330  HDJM50243  HDJM51322  HDJM51653  HDJM51711  HDJM51712  HDJM51947  HDJM51950  HDJM51951  HDJM51954  HDJM51955  HDJM51956  HDJM51958  HDJM51948  HDJM51949  HDJM51952  HDJM51953  HDJM51957  HDJM51959  HDJM51960  HDJM51961  HDJM51962  HDJM51963  HDJM51964  HDJM51965  HDJM51966  HDJM51967  HDJM51968  HDJM51969  HDJM51970  HDJM51971  HDJM51972  HDJM51973  HDJM51974  HDJM51975  HDJM51976  HDJM52069  HDJM52070  HDJN50039  HDJN50040  HDJN50041  HDJN50042  HDJN50043  HDJN50044  HDJN50045  HDJN50046  HDJN50047  HDJN50081  HDJN50082  HDJN50089  HDJN50091  HDJN50092  HDJN50093  HDJN50094  HDJN52329  HDJN52330  HDJP50039  HDJP50905  HDJP50906  HDJP50907  HDJP50908  HDJP51021  HDJP51899  HDJQ50223  HDJQ50222  HDJQ50333  HDJQ50334  HDJQ50335  HDJQ50418  HDJQ50330  HDJQ50332  HDJQ50387  HDJQ50272  HDJQ50273  HDJQ50331  HDJQ50386  HDJQ50229  HDJQ50417  HDJQ50415  HDJQ50440  HDJQ50497  HDJQ50224  HDJQ50225  HDJQ50230  HDJQ50528  HDJQ51101  HDJQ51232  HDJQ51194  HDJQ52627  HDJQ52893  HDJQ52900  HDJQ52901  HDJQ52902  HDJR50456  HDJR50536  HDJR50537  HDJR50538  HDJR54213  HDJR56800  HDJR57313  HDJR57314  HDJR57315  HDJR57316  HDJR57966  HDJR57967  HDJR58130  HDJR58131  HDJR58198  HDJS50158  HDJS50498  HDJS50499  HDJS50500  HDJS51290  HDJS52396  HDJS52397  HDJS52398  HDJS52592  HDJS52593  HDJR56108  HDJR57495  HDJR57615  HDJP52539  HDJM51456  HDJR56203  HDJR57745  HDJR57746  HDJR57747  HDJR57748  HDJQ50607  HDJR54136  HDJM51462  HDJQ50606  HDJS50797  HDJS50798  HDJS50800  HDJS51456  HDJS51457  HDJM52262  HDJN50663  HDJP51291  HDJP51290  HDJP51292  HDJP51293  HDJP51332  HDJR50360  377330  HDJS51054  HDJS51055  HDJS51056  HDJS51057  HDJP52475  HDJQ52820  HDJQ52821  HDJR57170  HDJR57171  HDJM52251  HDJR57741  HDJP51192  HDJR57369  HDJM50306  HDJM52076  HDJM50228  HDJM50229  HDJM50230  HDJM50231  HDJM50232  HDJM50233  HDJM50234  HDJM50235  HDJM50338  HDJM50309  HDJM50688  HDJM50753  HDJM50754  HDJM50755  HDJM52048  HDJM52049  HDJM52075  HDJM52139  HDJM52140  HDJM52166  HDJM52102  HDJM52099  HDJM52101  HDJM52397  HDJM52398  HDJM52399  HDJM52400  HDJM52401  HDJM52402  HDJN50052  HDJN50053  HDJN50054  HDJN50055  HDJN50056  HDJN50057  HDJN50260  HDJS51777  HDJS51778  HDJQ52065  HDJQ52066  HDJR56046  HDJN51587  HDJM50305  HDJM50337  HDJM50660  HDJM50661  HDJM51228  HDJM52162  HDJM52257  HDJM52258  HDJN50294  HDJN50316  HDJN50317  HDJN50531  HDJN50761  HDJP50382  HDJP50383  HDJP50384  HDJP50386  HDJP50388  HDJP50385  HDJP50387  HDJP51036  HDJP51037  HDJP51531  HDJP51532  HDJP51533  HDJP51534  HDJP51721  HDJP51722  HDJP51723  HDJP51724  HDJP51725  HDJP51726  HDJP51727  HDJP51728  HDJP51729  HDJP51730  HDJP51731  HDJP51732  HDJP51733  HDJP51734  HDJP51735  HDJP51736  HDJP51737  HDJP51738  HDJP51739  HDJP51740  HDJP51741  HDJP51742  HDJP51743  HDJP51744  HDJP51745  HDJP51746  HDJP51747  HDJP51748  HDJQ50873  HDJQ52716  HDJR50657  HDJN51886  HDJP50814  HDJQ50156  HDJR56476  HDJM50662  HDJM50663  HDJM52133  HDJM52281  HDJP50478  HDJP52553  HDJP52554  HDJP52555  HDJP52556  HDJP52557  HDJP52559  HDJS52112  HDJR54000  HDJR56047  HDJM50621  HDJM51555  HDJM51556  HDJM51557  HDJN51669  HDJQ52273  HDJQ52278  HDJQ52274  HDJQ52276  HDJQ52275  HDJQ52277  HDJP50457  HDJP50458  HDJS52595  HDJS52596  HDJS52597  HDJP52707  HDJP52708  HDJP52709  HDJP52710  HDJP52711  HDJQ51154  HDJR54230  HDJR54231  HDJR54232  HDJR56102  HDJR56103  HDJR56817  HDJR56809  HDJR56810  HDJS50633  HDJS50634  HDJM50385  HDJM51279  HDJM51280  HDJM51281  HDJM51814  HDJP52650  HDJQ50702  HDJQ50703  HDJN52100  HDJR50526  HDJR56480  HDJS51252  HDJS51552  HDJM51050  HDJM52136  HDJM52135  HDJM52137  HDJM52157  HDJM52158  HDJM52138  HDJM52255  HDJN50080  HDJN52239  HDJN52127  HDJN52240  HDJP51529  HDJP52648  HDJN50214  HDJN50215  HDJN50216  HDJN50217  HDJR50562  HDJS51334  HDJS51335  HDJS51336  HDJS51337  HDJS51338  HDJS51339  HDJS51340  HDJS51341  HDJS51342  HDJS51343  HDJS51344  HDJS51345  HDJS51346  HDJS51347  HDJS51348  HDJS51349  HDJS51350  HDJS51351  HDJS51352  HDJS51353  HDJS51354  HDJS51355  HDJS51356  HDJS51357  HDJS51358  HDJS51359  HDJS51360  HDJS51361  HDJS51362  HDJS51363  HDJS51364  HDJS51365  HDJS52702  HDJS52703  HDJS52704  HDJS52705  HDJS52706  HDJS52707  HDJS52708  HDJS52709  HDJS52710  HDJS52711  HDJN52228  HDJN52229  HDJN52230  HDJN52231  HDJN52232  HDJN52233  HDJN52234  HDJP51298  HDJP51299  HDJP51300  HDJP51301  HDJP51302  HDJP51303  HDJP52462  HDJP52463  HDJP52514  HDJR56928  HDJS51182  HDJS51183  HDJR56930  HDJS50198  HDJS51144  HDJS51248  HDJS51249  HDJS51250  HDJS51251  HDJS52283  HDJS52399  HDJM52090  HDJN51051  HDJN51052  HDJN51053  HDJP51548  HDJP51549  HDJP51550  HDJP51551  HDJP51552  HDJP51553  HDJQ50441  HDJQ50442  HDJQ52863  HDJQ52864  HDJQ52865  HDJQ52866  HDJQ52867  HDJQ52868  HDJQ52869  HDJQ52870  HDJQ52871  HDJQ52872  HDJQ52873  HDJQ52874  HDJQ52875  HDJQ52876  HDJN50095  HDJN50530  HDJN51496  HDJN51497  HDJN51498  HDJN51499  HDJS50799  HDJQ50249  HDJM50689  HDJM51704  HDJM51705  HDJM51706  HDJM51708  HDJM51703  HDJM51707  HDJM51709  HDJM51710  HDJM52042  HDJM52252  HDJN50339  HDJN50340  HDJN50341  HDJN50342  HDJN50343  HDJN51495  HDJP51712  HDJQ51129  HDJQ51328  HDJR56096  HDJR57203  HDJR57727  HDJS50675  HDJS50700  HDJS50701  HDJS51145  HDJS51146  HDJR52023  HDJR56983  HDJR58244  HDJS50036  HDJS50200  HDJM52256  HDJN51002  HDJN51001  HDJP51259  HDJP51260  HDJP51261  HDJP51262  HDJP51263  HDJP51264  HDJP51265  HDJP51266  HDJR50595  HDJR57243  HDJR57239  HDJR57240  HDJR57241  HDJR57242  HDJR57244  HDJR57245  HDJR57370  HDJR58119  HDJR58120  HDJQ51189  HDJQ51190  HDJQ52808  HDJR56799  HDJR50532  HDJR50533  HDJR50534  HDJR50535  HDJR54273  HDJR56100  HDJR56101  HDJR57371  HDJR57372  HDJR57373  HDJR57374  HDJR57571  HDJN52118  HDJN51637  HDJN51636  HDJP51801  HDJP51555  HDJP52109  HDJQ50063  HDJQ50064  HDJQ50342  HDJQ51188  HDJQ51724  HDJQ51726  HDJQ51727  HDJQ51822  HDJQ51725  HDJQ52023  HDJQ52024  HDJS50299  HDJS50326  HDJS50327  HDJS50328  HDJQ52843  HDJM51940  HDJP50462  HDJP50463  HDJP50464  HDJQ50126  HDJQ50127  HDJQ50128  HDJQ50129  HDJQ50130  HDJQ52400  HDJQ52401  HDJQ52402  HDJS51779  HDJS51780  HDJS51781  HDJS51782  HDJS51783  HDJS51784  HDJS51785  HDJS51786  HDJS51788  HDJS51790  HDJS51791  HDJS51792  HDJS51793  HDJS51794  HDJS51795  HDJS51796  HDJS52102  HDJS52105  HDJS51787  HDJS51789  HDJS52097  HDJS52098  HDJS52099  HDJS52100  HDJS52101  HDJS52103  HDJS52104  HDJS52106  HDJS52107  HDJS52108  HDJS52109  HDJM51929  HDJP52649  HDJN52270  HDJQ50248  HDJQ50874  HDJQ50875  HDJP51988  HDJQ50591  HDJQ50592  HDJQ51582  HDJQ51583  HDJQ51584  HDJQ51585  HDJR50658  HDJR50659  HDJR50660  HDJR50661  HDJS50039  HDJR58216  HDJS50038  HDJS51120  HDJS52264  HDJS52422  HDJQ51267  HDJQ51268  HDJQ51269  HDJQ51270  HDJQ51271  HDJQ51272  HDJQ51273  HDJQ51274  HDJQ51275  HDJQ51276  HDJQ51277  HDJQ51278  HDJQ51279  HDJQ51280  HDJQ51281  HDJQ51282  HDJQ51283  HDJQ51284  HDJQ51285  HDJQ51286  HDJR56109  HDJM50888  HDJM51725  HDJP50931  HDJQ51932  HDJP51323  HDJP52263  HDJQ51720  HDJQ51721  HDJR50787  HDJR56818  HDJS50323  HDJS50324  HDJS50325  HDJP51799  HDJP52227  HDJP52228  HDJR56657  HDJR56681  HDJR56989  HDJR56990  HDJR56991  HDJR56992  HDJS51052  HDJS51053  HDJM52250  HDJS50316  HDJN52225  HDJP51554  HDJS51459  HDJN50272  HDJP51714  HDJP51715  HDJQ50207  HDJQ51128  HDJR56972  HDJR56973  HDJR56974  HDJR56686  HDJQ51298  HDJQ51299  HDJQ51684  HDJQ51685  HDJQ51686  HDJQ51687  HDJQ51688  HDJQ51689  HDJQ51690  HDJQ51691  HDJQ51692  HDJQ51693  HDJQ51694  HDJQ51695  HDJQ51696  HDJQ51697  HDJQ51698  HDJQ52892  HDJQ50526  HDJM50000  HDJM50001  HDJM50002  HDJM50323  HDJM52219  HDJM52220  HDJN50126  HDJN50127  HDJN50128  HDJN50129  HDJN50130  HDJN50131  HDJN50132  HDJN50357  HDJN51670  HDJN51671  HDJP50312  HDJP50313  HDJP50815  HDJP51304  HDJP52515  HDJP52516  HDJP52517  HDJP52518  HDJP52519  HDJP52520  HDJP52521  HDJP52522  HDJQ53019  HDJQ53020  HDJR56098  HDJR56281  HDJR57973  HDJR57974  HDJR57975  HDJS50721  HDJS50722  HDJS51687",
      "center_classification_date": "20140203",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Uruguay, Venezuela, Albania, Algeria, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Canada, Chile, China, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Kazakhstan, Korea, Kuwait, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, and United Kingdom.",
      "state": "MD",
      "product_description": "Giraffe¿ Infant Warmers, part number M1118179, labeled in part *** GE Healthcare 8880 Gorman Road Laurel, MD 20723***www.gehealthcare.com*    Giraffe¿ Infant Warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.",
      "report_date": "20140212",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare",
      "recall_number": "Z-0912-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67225",
      "termination_date": "20150601",
      "more_code_info": "",
      "recall_initiation_date": "20131212",
      "postal_code": "20723-5800",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}