      "country": "United States",
      "city": "Carlsbad",
      "address_1": "2271 Cosmos Ct",
      "reason_for_recall": "Potential issues with ventilator navigation ring interface and liquid ingress:  1) The navigation ring may become inoperative or operate intermittently;  2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments.",
      "address_2": "",
      "product_quantity": "3213 systems",
      "code_info": "Model: V60  Part Number: R1076709  V60 Ventilator, JAPAN OPT: CFLEX,AVAPS,AT+    Serial Numbers:  100065352  100065365  100065371  100065487  100065493  100067423  100067425  100067428  100067431  100067451  100068214  100069518  100070301  100070306  100070398  100070476  100070491  100070492  100070768  100070774  100070776  100070777  100070778  100070780  100070781  100070859  100070862  100070865  100070921  100070922  100070924  100070928  100070929  100070930  100070935  100070966  100070969  100070973  100070976  100070981  100071071  100071073  100071077  100071123  100071461  100071510  100071511  100072415  100072417  100072418  100072499  100072570  100072606  100072607  100072673  100072676  100072679  100072684  100072685  100072688  100072690  100072695  100074195  100075099  100075100  100075104  100075189  100075293  100075312  100075381  100078120  100078132  100078140  100078145  100078266  100078302  100078313  100078362  100078365  100078366  100078372  100079387  100080231  100080233  100080236  100080237  100080239  100080271  100080275  100080276  100080277  100080282  100080285  100080290  100080291  100080292  100080294  100080295  100080297  100080298  100080421  100080422  100080423  100080424  100080425  100080427  100080428  100080429  100080430  100080528  100080529  100080530  100080532  100080533  100080535  100080536  100080537  100080540  100080541  100080542  100080543  100080544  100080545  100080547  100080552  100080553  100080563  100080642  100080643  100080644  100080647  100080650  100080651  100080652  100080653  100080654  100080657  100080658  100080659  100080704  100080706  100080708  100080711  100080713  100080716  100080717  100080721  100080722  100080723  100080724  100080725  100080726  100080729  100080731  100080733  100080734  100080736  100080740  100080840  100080841  100080843  100080845  100080846  100080847  100080849  100080850  100080852  100081698  100083532  100083534  100083537  100083538  100083539  100083561  100083563  100083564  100083567  100083569  100083577  100083601  100083603  100083811  100083813  100083816  100083817  100083821  100083822  100083823  100083824  100083825  100083828  100083830  100083832  100083833  100083920  100083921  100083922  100083923  100083924  100083925  100083926  100083927  100083929  100083930  100084955  100084957  100085084  100085086  100085087  100085089  100085090  100085092  100085095  100085242  100085245  100085248  100085249  100085250  100085253  100085322  100085323  100085324  100085325  100085326  100085328  100085329  100085331  100085332  100085334  100085336  100085337  100085338  100085339  100085380  100085381  100085383  100085385  100085386  100085387  100085388  100085395  100085397  100085480  100085482  100085484  100085485  100085486  100085487  100085488  100086236  100086783  100086891  100086896  100086902  100087025  100087028  100087049  100087050  100087052  100087093  100087095  100087097  100087103  100087104  100087172  100087344  100087350  100087351  100087352  100087353  100087354  100087357  100087358  100087462  100087463  100087464  100087465  100087468  100087471  100087472  100087474  100087475  100087477  100087478  100087482  100087484  100087488  100087489  100087602  100087607  100087608  100087609  100087612  100087613  100087616  100087617  100087682  100087683  100087684  100087685  100087691  100087693  100087695  100087696  100087697  100087698  100087699  100088046  100088054  100088055  100088059  100088060  100088429  100088432  100088433  100088434  100088435  100088437  100088438  100088439  100088859  100089023  100089024  100089025  100089026  100089027  100089028  100089029  100089030  100089031  100089032  100089120  100089121  100089123  100089124  100089125  100089126  100089128  100089129  100089130  100089132  100089133  100089134  100089135  100089136  100089138  100089140  100089141  100089142  100089143  100089144  100089145  100089146  100089147  100089148  100089260  100089261  100089262  100089263  100089264  100089266  100089267  100089268  100089270  100089271  100089272  100089274  100089275  100089276  100089277  100089278  100089279  100089280  100089437  100089438  100089440  100089441  100089443  100089444  100089445  100089446  100089447  100089448  100089450  100089451  100089459  100089460  100089462  100089463  100089464  100089802  100089803  100089804  100090135  100090531  100090545  100090547  100090550  100090552  100090553  100090557  100090652  100090654  100090658  100090665  100090668  100090670  100090672  100090674  100090677  100090680  100090681  100090778  100090779  100090781  100090782  100090784  100090786  100090787  100090788  100090789  100090792  100090794  100090796  100090797  100090798  100090799  100090800  100090801  100090803  100090804  100090805  100090914  100090917  100090919  100090922  100090926  100090927  100090928  100090929  100090930  100091265  100091390  100091393  100091398  100091400  100091402  100091404  100091405  100091721  100091725  100091727  100091731  100091733  100091734  100091736  100091740  100091741  100091742  100091744  100091746  100091819  100091840  100091844  100091846  100091850  100091856  100091861  100091862  100091865  100091866  100091867  100091918  100091919  100100747  100100850  100100995  100100997  100100998  100100999  100101056  100101060  100101062  100101063  100101066  100101067  100101074  100101396  100101400  100101402  100101407  100101409  100101410  100101412  100101414  100101421  100101422  100101423  100101425  100101500  100101501  100101504  100102165  100102168  100102170  100103373  100103374  100103375  100103376  100103377  100103378  100103379  100103461  100103464  100103477  100103479  100103480  100103575  100103576  100103577  100103578  100103579  100103581  100103582  100103586  100103592  100103594  100103595  100103596  100103660  100103662  100103664  100103666  100103671  100103672  100103675  100104949  100105246  100105559  100105877  100105878  100105879  100106169  100106170  100106171  100106172  100106175  100106177  100106181  100106186  100106187  100106189  100106190  100106191  100106321  100106322  100106323  100106326  100106328  100106329  100106331  100106332  100106333  100106334  100106335  100106336  100106401  100106404  100106405  100106407  100106408  100106410  100106411  100106413  100106414  100106416  100106419  100106440  100106445  100106498  100106500  100106502  100106503  100106505  100106509  100106510  100106513  100106526  100106605  100106606  100106608  100106609  100106732  100106733  100106735  100106736  100106737  100106738  100110587  100110588  100110590  100110592  100110593  100110594  100110596  100110600  100110601  100110602  100110604  100110649  100110652  100110657  100110658  100110659  100110660  100110661  100110662  100110665  100110666  100110667  100110668  100110681  100110682  100114055  100114058  100114177  100114178  100114439  100114443  100114444  100114445  100114447  100114448  100114449  100114451  100114452  100114453  100114454  100114455  100114456  100114460  100114512  100114514  100114522  100114524  100118045  100118050  100118052  100118053  100118054  100118057  100118059  100118063  100118064  100118068  100118069  100118070  100124806  100124860  100124861  100124866  100124958  100124960  100124964  100125471  100125473  100126846  100126850  100126853  100126856  100126868  100131913  100131914  100132024  100134037  100134286  100134288  100134295  100134349  100134390  100134392  100134395  100136623  100136624  100136633  100136635  100136648  100136650  100136651  100136652  100136657  100136658  100136659  100136711  100136713  100136715  100136760  100136770  100136771  100136772  100136773  100136775  100136801  100136802  100136803  100136806  100136807  100136808  100136887  100136888  100136889  100136890  100136892  100136893  100136895  100136896  100136898  100136899  100137000  100137002  100137009  100137014  100137015  100137017  100137018  100137020  100137024  100137025  100137027  100137029  100137132  100137133  100137134  100137136  100137138  100137139  100137182  100137185  100137186  100137187  100137189  100137191  100137192  100137193  100137195  100137198  100137199  100137200  100137204  100137205  100137206  100137207  100137208  100137209  100137210  100137212  100137213  100137214  100137215  100137216  100137480  100137482  100137483  100137484  100137485  100137487  100137489  100137492  100137493  100137495  100137496  100137497  100137498  100137499  100137500  100137501  100137502  100137504  100137505  100137535  100137537  100137538  100137539  100137635  100137636  100137637  100137638  100137673  100137677  100137679  100137780  100139712  100139713  100139714  100139715  100139716  100139717  100139718  100139800  100139801  100139802  100139803  100139804  100139805  100139807  100139812  100139813  100139814  100139815  100139816  100139817  100139820  100139900  100139903  100139906  100139907  100142368  100159642  100159646  100159648  100159651  100159675  100159677  100159678  100159697  100159762  100159763  100159765  100159767  100159768  100159772  100159774  100159775  100159776  100159779  100159820  100159821  100159822  100159828  100159829  100159877  100159922  100159927  100159928  100160001  100160002  100162794  100162903  100162912  100162951  100162954  100163020  100163021  100163024  100163038  100163044  100163432  100163433  100163436  100163437  100163438  100163439  100163560  100163561  100163562  100163563  100163564  100163565  100163567  100163568  100163569  100163571  100163572  100163573  100163574  100163575  100163576  100163577  100163578  100163579  100163581  100163586  100163587  100163588  100165169  100165688  100169356  100169360  100169361  100169366  100169371  100169372  100169536  100169537  100169539  100169540  100169541  100169544  100169546  100169551  100169552  100169553  100169559  100169769  100172776  100172778  100172779  100172955  100173104  100173106  100173107  100173112  100173114  100173115  100173126  100173130  100173202  100173203  100173206  100173211  100173212  100173216  100173218  100173235  100173440  100173455  100173457  100173460  100173461  100173464  100173467  100173468  100173469  100173471  100173473  100173475  100173477  100173479  100173520  100173521  100173522  100173523  100173524  100173594  100174174  100174309  100174960  100174962  100174965  100174966  100174968  100174969  100174972  100174973  100174975  100174976  100174980  100174982  100174983  100174984  100174985  100174987  100174991  100174992  100174995  100175035  100175543  100175567  100175568  100175571  100175575  100175576  100175577  100175578  100175808  100175809  100175810  100175813  100175814  100175817  100175820  100175822  100175823  100175824  100175826  100175827  100175828  100175829  100175830  100175831  100175832  100175834  100175835  100175836  100175838  100175839  100175840  100175841  100175843  100175844  100175845  100175846  100176003  100176004  100176005  100176006  100176007  100176008  100176009  100176010  100176011  100176012  100176013  100176014  100176015  100176016  100176017  100176018  100176021  100176024  100176026  100176027  100176030  100176033  100176037  100176038  100176039  100176317  100176320  100176326  100176328  100176329  100176334  100176335  100176336  100176340  100176341  100176344  100180766  100180767  100180768  100180769  100180770  100180771  100180775  100180776  100180777  100180778  100180779  100180941  100180943  100180946  100180948  100180949  100180950  100180952  100180954  100180955  100180956  100180958  100180959  100180961  100181205  100181206  100181208  100181209  100181211  100181213  100181216  100181217  100181220  100181221  100181222  100181223  100181225  100181227  100181228  100181229  100181231  100181232  100181233  100181235  100181236  100181237  100181243  100181244  100181247  100181249  100181250  100181251  100182434  100184159  100184180  100184181  100184182  100184184  100184185  100184191  100184193  100184194  100184199  100184329  100184330  100184343  100184344  100184346  100184347  100184348  100184349  100184351  100184359  100184364  100187085  100187086  100187087  100187089  100187114  100187117  100187120  100187121  100187122  100187123  100187124  100187125  100187126  100187127  100187128  100187131  100187132  100187293  100187297  100187298  100187301  100187302  100187305  100187306  100187307  100187308  100187309  100187310  100187311  100187314  100187315  100187316  100187318  100187320  100189726  100189728  100189730  100189732  100189734  100189737  100189738  100189739  100189744  100189820  100189823  100189826  100189834  100189836  100189882  100189883  100189885  100189896  100189906  100189919  100189980  100189984  100189985  100189986  100189987  100190005  100190098  100190174  100190188  100190193  100190194  100190195  100195312  100195444  100195452  100195453  100195465  100195473  100195637  100195640  100195648  100195650  100195653  100195654  100195665  100195670  100195874  100195893  100195900  100196014  100196023  100196133  100196138  100196429  100196431  100196436  100196622  100196623  100221142  100221158  100221162  100221165  100221167  100221168  100221174  100223191  100223192  100223193  100223194  100223196  100223197  100223198  100223199  100223201  100223202  100223203  100223204  100223205  100223206  100223208  100223209  100223210  100223211  100223212  100223213  100223214  100223215  100223216  100223217  100223218  100223219  100223220  100223221  100223222  100223231  100223233  100223235  100223236  100223237  100223238  100223280  100223283  100223284  100223286  100223287  100223288  100223289  100223290  100223293  100223294  100223295  100223296  100223297  100223298  100223299  100223300  100223303  100223304  100223305  100223454  100223456  100223457  100223458  100223459  100223463  100223464  100223465  100223469  100223470  100223471  100223475  100223476  100223479  100223481  100226861  100226877  100226878  100226881  100226883  100226884  100227091  100227094  100227095  100227096  100227099  100227128  100227129  100227130  100227131  100227132  100227133  100227134  100227135  100227136  100227137  100227138  100227139  100227140  100227141  100227142  100227143  100227144  100227145  100227146  100227147  100227148  100227149  100227391  100227392  100227393  100227394  100227395  100227396  100227397  100227398  100227399  100227400  100227402  100227403  100227404  100227405  100227406  100227407  100227408  100227409  100227410  100227416  100227421  100227789  100227799  100227800  100227801  100227803  100227804  100227805  100227806  100227807  100228142  100228144  100228145  100228146  100228147  100228148  100228151  100228152  100228153  100228154  100228155  100228156  100228157  100228158  100228159  100228160  100228161  100228162  100228163  100228164  100228165  100228167  100228169  100228170  100228171  100228173  100228174  100228175  100228177  100228465  100228467  100228468  100228469  100228472  100228479  100228610  100228611  100228612  100228614  100228615  100228618  100228619  100228621  100228622  100228624  100228625  100228626  100228628  100228633  100228635  100228636  100228639  100228727  100228728  100228729  100228730  100228732  100228734  100228823  100228825  100228833  100228921  100228926  100228928  100231106  100231108  100232473  100232488  100232626  100232629  100232631  100232633  100232634  100232638  100232639  100232640  100232642  100232644  100232645  100232647  100232650  100232651  100232653  100232655  100232656  100232657  100232658  100232741  100232743  100232747  100232748  100232749  100232750  100232752  100232754  100232763  100232764  100232768  100232769  100232771  100233055  100233056  100233059  100233061  100233062  100233065  100233067  100233068  100233069  100237319  100237320  100237378  100237622  100237631  100237641  100237642  100237643  100237644  100237645  100237646  100237647  100237648  100237649  100237650  100237651  100237652  100237653  100237654  100237655  100237656  100237657  100237658  100237659  100237660  100238319  100238400  100238401  100238402  100238403  100238404  100238405  100238406  100238407  100238408  100238409  100238410  100238411  100238412  100238413  100238414  100238415  100238417  100238420  100238421  100238422  100238423  100238424  100238425  100238426  100238428  100239248  100239249  100239250  100239251  100239255  100239256  100239257  100239259  100239276  100239277  100239278  100239279  100239280  100239281  100239282  100239283  100239284  100239285  100239286  100239287  100239288  100239289  100239290  100239291  100239292  100239293  100239294  100239296  100239298  100239299  100239300  100239301  100239302  100239305  100239306  100239307  100239308  100239309  100239310  100239311  100239312  100239313  100239314  100239315  100239380  100239381  100239382  100239387  100239389  100239391  100239392  100239393  100239397  100239408  100239526  100239528  100239535  100239581  100239582  100239584  100239585  100239586  100239587  100239588  100239589  100239590  100239591  100239592  100239593  100239594  100239595  100239596  100239597  100239598  100239599  100239600  100239601  100239602  100239603  100239605  100239606  100239607  100239608  100239609  100239610  100239611  100239612  100239613  100239614  100239615  100239616  100239617  100239800  100239801  100239802  100239803  100239805  100239806  100239807  100239808  100239809  100239810  100239812  100239813  100239814  100239815  100239816  100239817  100239818  100239819  100239820  100239822  100239823  100239824  100239825  100239826  100239827  100239828  100239829  100239830  100240145  100240146  100240147  100240148  100240150  100240153  100240154  100240155  100240156  100240157  100240158  100240159  100240160  100240161  100240162  100240163  100240164  100240165  100240166  100240167  100240168  100240169  100240170  100240171  100240172  100240174  100240175  100240176  100243236  100243244  100243514  100243518  100243626  100243636  100243638  100243643  100243647  100243648  100243650  100243653  100243660  100243663  100243710  100243711  100243712  100243713  100243715  100243716  100243717  100243904  100243913  100243922  100243923  100243929  100243932  100244188  100244190  100244191  100244192  100244194  100244196  100244197  100244198  100244202  100244205  100244210  100244211  100244212  100244585  100244589  100244591  100244595  100244597  100244598  100244601  100244602  100244603  100244604  100244605  100244606  100247270  100247418  100247616  100247618  100247764  100247770  100247772  100247773  100247775  100247779  100247781  100247784  100247930  100247933  100247946  100247948  100247949  100248150  100248151  100248155  100248159  100248302  100248303  100248305  100248306  100248307  100248308  100248309  100248310  100248311  100248314  100248315  100248316  100248319  100248320  100248321  100248322  100248323  100248578  100248581  100248582  100248583  100248584  100248585  100248588  100248589  100248590  100248591  100248592  100248597  100249446  100249447  100249448  100249449  100249452  100249455  100249501  100249511  100249743  100249747  100249764  100249907  100249913  100249919  100249920  100249921  100249926  100249927  100249929  100249930  100249938  100249939  100250320  100250321  100250322  100250323  100250324  100250325  100250326  100250327  100250329  100250330  100250332  100250334  100250335  100250336  100250337  100250343  100250345  100250366  100252564  100252570  100252578  100252595  100252604  100252606  100252614  100252622  100252784  100252786  100252791  100252794  100252945  100252946  100252947  100252948  100252949  100252952  100252953  100252954  100252955  100252956  100252957  100252958  100252959  100252960  100252961  100252966  100252967  100252968  100252972  100253475  100253480  100253482  100253483  100253485  100253490  100253499  100253554  100253555  100253558  100253559  100253722  100254409  100254413  100254416  100254417  100254418  100254443  100254444  100254445  100254447  100254453  100254454  100254455  100254457  100254458  100254461  100254462  100254511  100254512  100254514  100254517  100254518  100254519  100254543  100254544  100254545  100254548  100254551  100254554  100254555  100254557  100254558  100254559  100254561  100254562  100254563  100254567  100254654  100254656  100254781  100255549  100255551  100255554  100255555  100255628  100255631  100255632  100255633  100255634  100255995  100255996  100255998  100256003  100256004  100256007  100256010  100256011  100256012  100256014  100256015  100256016  100256019  100256020  100259659  100259728  100259730  100259738  100259795  100259798  100259859  100260041  100260232  100260448  100260969  100261116  100261118  100261277  100261300  100272765  100272769  100272773  100272775  100272778  100272783  100272784  100272785  100272787  100272789  100274091  100274098  100274100  100274101  100274102  100274103  100274104  100274257  100274258  100274259  100274260  100274262  100274265  100274266  100274267  100274271  100274272  100274273  100274275  100274276  100277417  100277418  100277442  100277443  100277444  100277448  100277451  100277452  100277453  100277454  100277456  100277458  100277462  100277464  100277465  100277467  100277470  100277471  100277472  100277473  100277474  100277475  100277476  100277796  100277798  100277801  100277802  100277807  100277809  100277810  100277812  100277813  100277814  100277817  100277818  100277871  100277872  100277873  100277874  100277875  100277878  100277879  100277880  100277881  100277882  100277884  100277885  100277887  100277888  100277891  100277892  100277893  100277895  100277897  100277898  100277899  100277900  100277905  100277908  100277913  100277918  100279010  100279461  100279462  100279465  100279468  100279472  100279473  100279474  100279475  100279476  100279477  100279478  100279479  100279920  100279921  100279922  100279923  100279924  100279925  100279926  100279927  100279928  100279929  100279933  100280227  100280228  100280229  100280230  100280233  100280235  100280237  100280238  100280240  100280241  100280242  100280243  100280246  100280247  100280249  100280250  100280251  100280254  100280255  100280257  100280259  100280262  100280263  100280264  100280265  100280267  100280268  100280270  100280271  100280272  100280275  100280278  100280391  100280392  100280396  100280399  100280402  100280404  100280405  100280406  100280410  100280411  100280413  100280414  100280415  100280417  100280500  100280501  100280502  100280504  100280505  100280506  100280508  100280509  100280581  100280582  100280587  100280588  100280590  100280591  100280592  100280593  100280594  100280595  100280596  100280597  100280598  100280599  100280600  100280601  100280602  100280603  100280604  100280605  100280606  100280607  100280608  100280609  100280610  100280612  100280613  100280615  100280616  100280617  100280618  100280619  100280620  100280748  100280749  100280752  100280753  100280754  100280757  100280976  100280979  100280980  100280985  100280986  100284949  100284950  100284951  100284952  100284953  100284954  100284955  100284956  100284957  100284958  100284959  100285006  100285008  100285009  100285011  100285015  100285016  100285019  100285141  100285142  100285145  100285300  100285784  100285785  100285787  100285790  100285791  100285792  100285793  100285794  100285797  100285798  100286040  100286042  100286044  100286045  100286049  100286211  100286212  100286214  100286280  100286281  100286282  100286283  100286285  100286289  100286290  100286291  100286295  100286302  100286307  100286308  100286310  100286313  100286316  100286318  100286460  100286461  100286468  100286469  100286471  100290649  100290658  100290665  100290667  100290669  100290802  100290803  100290805  100290807  100290808  100290809  100290810  100290811  100290813  100290814  100290815  100290817  100290818  100290819  100290820  100290821  100290822  100290824  100290826  100290827  100290828  100290829  100290830  100290831  100290834  100290835  100290836  100290837  100291179  100291180  100291181  100291182  100291184  100291186  100291188  100291191  100291192  100291193  100291194  100291199  100291200  100291201  100291202  100291203  100291204  100291206  100291210  100291213  100291216  100291217  100291361  100291371  100291376  100291381  100291383  100291384  100291385  100291386  100291387  100291388  100291451  100291452  100291453  100291456  100291458  100291466  100291471  100293604  100293612  100293615  100293616  100293620  100296845  100297840  100302000  100311376  100311377  100311474  100311481  100311485  100311488  100311776  100311777  100311778  100311800  100311802  100311803  100311805  100311811  100311860  100311861  100311862  100311863  100311864  100311865  100311872  100311878  100311880  100312304  100312305  100312306  100312307  100312308  100312311  100312312  100312313  100312314  100312315  100312316  100312318  100312440  100312441  100312442  100312443  100312446  100312447  100312450  100312455  100312456  100312457  100312458  100312521  100312523  100312528  100312530  100312532  100312538  100313062  100313064  100313066  100313067  100313068  100313069  100313075  100313077  100313078  100313100  100346867  100346880  100346887  100346893  100346934  100346936  100346951  100346955  100348091  100348129  100348136  100348148  100350139  100350143  100350161  100350930  100351483  100351493  100351498  100351499  100351542  100351544  100351552  100351687  100352842  100352951  100352956  100353258  201004350  201004356  201004437  201007443  201008026  201008027  201008054  201008076  201008077  201008078  201008079  201008122  201008123  201008126  201008128  201008129  201008133  201008140  201008142  201008144  201008145  201008149  201008150  201008154  201008164  201008166  201008170  201008177  201008203  201008208  201009613  201009617  201009619  201009626  201010133  201010134  201010138  201010161  201010171  201010199  201010222  201010260  201010263  201010308  201010312  201010313  201010315  201010344  201010498  201010524  201010525  201010529  201010533  201010560  201010563  201010564  201010569  201012990  201013084  201013114  201013117  201013121  201013122  201013128  201013150  201013151  201013152  201013153  201013156  201013159  201013164  201013166  201013167  201013169  201013170  201013171  201013173  201013175  201013176  201013177  201013178  201013179  201013726  201013729  201013734  201013735  201013736  201013737  201013754  201013755  201013757  201013758  201013759  201013763  201013765  201013769  201013770  201013781  201013783  201013794  201013795  201013797  201013808  201013810  201013811  201013812  201013813  201014919  201014962  201014963  201014966  201014967  201014968  201014971  201014972  201014973  201014974  201014976  201014977  201014979  201014990  201014992  201014996  201014999  201015003  201015005  201015006  201015007  201015020  201015215  201015230  201015485  201022728  201022732  201022734  201022967  201022983  201023009  201023052  201023055  201023086  201023105  201023109  201023110  201023677  201023678  201023905  201024005  201024033  201024831  201024832  201024834  201024836  201024837  201025282  201025283  201025287  201025298  201025304  201025308  201025310  201025311  201025313  201025317  201025381  201025384  201025386  201025388  201025855  201025856  201025858  201025913  201025920  201025931  201025932  201025940  201025941  201025948  201025949  201025955  201025956  201025960  201025962  201025963  201025964  201025965  201025967  201025968  201025969  201025970  201025971  201025973  201025974  201025975  201025977  201025981  201025982  201025990  201025991  201025999  201026001  201026003  201026350  201026351  201026352  201026353  201026354  201026356  201026357  201026358  201026359  201026380  201026381  201026382  201026383  201026384  201026385  201026386  201026388  201026389  201026391  201026392  201026393  201026394  201026395  201026396  201026397  201026398  201026399  201026401  201026402  201026403  201026404  201026405  201026407  201026408  201026409  201026410  201026411  201026412  201026414  201026415  201026417  201026418  201026419  201026420  201026421  201026422  201026423  201026424  201026425  201026426  201026427  201026429  201026440  201026441  201026442  201026443  201026445  201026446  201026450  201026451  201026452  201026453  201026455  201026456  201026457  201026459  201026462  201026463  201026465  201026466  201026468  201026470  201026471  201026472  201026476  201026477  201026478  201026521  201026522  201026523  201027296  201027299  201027425  201027511  201027515  201027581  201027688  201027790  201027823  201027825  201027880  201027881  201027890  201027892  201027897  201027898  201027900  201027901  201027908  201027914  201027920  201027921  201027923  201027930  201027931  201027937  201027943  201027960  201027961  201027962  201027973  201027977  201027983  201027988  201027989  201027990  201027995  201027999  201028003  201028018  201028026  201028027  201028028  201028031  201028257  201028258  201028259  201028280  201028281  201028282  201028283  201028286  201028288  201028289  201028290  201028483  201028486  201028490  201028495  201028496  201028497  201028521  201028522  201028523  201028527  201028532  201028538  201028539  201028542  201028543  201028544  201028547  201028548  201028549  201028550  201028551  201028553  201028554  201028558  201028559  201028603  201028605  201028606  201028609  201028612  201028613  201028614  201028617  201028618  201028619  201028620  201028621  201028623  201028624  201028625  201028627  201028628  201028630  201028631  201028680  201028682  201028683  201028684  201028798  201028885  201028887  201028888  201028891  201028894  201028895  201028900  201028916  201028977  201028989  201028992  201028994  201028996  201029003  201029004  201029011  201029012  201029014  201029015  201029017  201029018  201029025  201029027  201029028  201029032  201029034  201029038  201029041  201029048  201029103  201029104  201029107  201029110  201029111  201029112  201029113  201029114  201029118  201029121  201029122  201029126  201029127  201029128  201029129  201029132  201029136  201029137  201029139  201029142  201029143  201029144  201029145  201029146  201029147  201029148  201029150  201029151  201029152  201029157  201029158  201029159  201029821  201029842  201029872  201029874  201029906  201029926  201029929  201029951  201030421  201030422  201030424  201030425  201030426  201030427  201030428  201030429  201030430  201030431  201030434  201030435  201030437  201030438  201030439  201030738  201030739  201030749  201030751  201030752  201030753  201030755  201030760  201030761  201030762  201030763  201030765  201030766  201030767  201030768  201030770  201030771  201030772  201030773  201030778  201030782  201030783  201030788  201030790  201030791  201030793  201030796  201030797  201030798  201030823  201030824  201030825  201030826  201030827  201030829  201030831  201030837  201030839  201030840  201030844  201030845  201030851  201030852  201030853  201030854  201030855  201030857  201030858  201030859  201030863  201030866  201030867  201030868  201030869  201030870  201030871  201030872  201030873  201030874  201030875  201030876  201030877  201030878  201030879  201030880  201030886  201030890  201030896  201030897  201030900  201030901  201030902  201030903  201030904  201030906  201030907  201030908  201030909  201030910  201030911  201030912  201030913  201030918  201030920  201030922  201030923  201030924  201030926  201030928  201030949  201030956  201031060  201031064  201031067  201031071  201031073  201031075  201031749  201031750  201031751  201031752  201031755  201031756  201031759  201031933  201031937  201031939  201032010  201032068  201032069  201032071  201032115  201032116  201032117  201032118  201032119  201032157  201032158  201032159  201032163  201032164  201032171  201032174  201032175  201032183  201032184  201032185  20103218",
      "center_classification_date": "20210122",
      "distribution_pattern": "Worldwide distribution - US Nationwide including Guam and Puerto Rico;  Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia.",
      "state": "CA",
      "product_description": "Philips Respironics V60 Ventilator Part Number R1076709 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.",
      "report_date": "20210203",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Respironics California, LLC",
      "recall_number": "Z-0910-2021",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "86697",
      "more_code_info": "6  201032227  201032228  201032235  201032236  201032237  201032238  201032239  201032241  201032242  201032244  201032246  201032247  201032251  201032252  201032253  201032254  201032257  201032320  201032321  201032322  201032323  201032324  201032325  201032326  201032327  201032329  201032330  201032331  201032332  201032333  201032334  201032335  201032336  201032337  201032338  201032339  201032340  201032341  201032342  201032343  201032345  201032346  201032347  201032348  201032349  201032350  201032351  201032360  201032361  201032362  201032363  201032364  201032365  201032366  201032367  201032368  201032369  201032370  201032371  201032373  201032374  201032375  201032376  201032377  201032378  201032379  201032380  201032381  201032382  201032383  201032384  201032400  201032402  201032403  201032405  201032406  201032407  201032408  201032409  201032410  201032411  201032412  201032413  201032414  201032415  201032416  201032417  201032418  201032419  201032420  201032421  201032422  201032423  201032424  201032427  201032428  201032431  201032433  201032434  201032450  201032451  201032453  201032454  201032455  201032456  201032457  201032459  201032460  201032462  201032463  201032465  201032470  201032471  201032483  201032500  201032503  201032504  201032507  201032511  201032512  201032513  201032514  201032515  201032516  201032519  201032520  201032521  201032523  201032524  201032995  201032996  201032997  201032999  201033001  201033009  201033010  201033015  201033016  201033157  201033158  201033159  201033269  201033270  201033272  201033275  201033279  201033283  201033284  201033285  201033286  201033287  201033288  201033289  201033291  201033292  201033300  201033320  201033321  201033322  201033324  201033325  201033326  201033327  201033329  201033330  201033331  201033332  201033333  201033334  201033335  201033336  201033337  201033338  201033339  201033341  201033343  201033344  201033346  201033347  201033348  201033369  201033370  201033372  201033373  201033374  201033376  201033377  201033379  201033380  201033381  201033402  201033404  201033431  201033596  201033625  201033684  201033698  201033729  201033737  201033744  201033745  201033746  201033750  201033752  201033755  201033756  201033757  201033758  201033760  201033761  201033762  201033763  201033765  201033768  201033769  201033771  201033773  201033774  201033775  201033776  201033777  201033779  201033780  201033781  201033782  201033783  201033784  201033786  201033787  201033789  201033790  201033791  201033792  201033800  201033801  201033803  201033804  201033805  201033806  201033807  201033808  201033811  201033812  201033814  201033815  201033816  201033818  201033841  201033842  201033845  201033846  201033863  201033871  201033872  201037822  201037825  201037833  201037834",
      "recall_initiation_date": "20201102",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}