  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75851",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US:  AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT,NC, NE NH, NJ, NM, NY, NV, OH, OK, PA, RI, SC, SD, TN, TX,, VA, WA, WO, WY, WV.     OUS:     Algeria  Armenia  Australia  Austria  Brazil  Canada  Chile  China  Denmark  Egypt  France  Germany  Guatemala  Honduras  Hong Kong  Hungary  Iceland  India  Ireland  Italy  Japan  Korea  Libya  Mexico  Netherlands  Netherlands  Peru  Poland  Portugal  Russia  Saudi Arabia  Serbia  Sierra Leone  Slovakia  Spain  Taiwan  Thailand  Turkey  United Arab Emirates  United Kingdom  Venezuela  Vietnam",
      "recall_number": "Z-0909-2017",
      "product_description": "Legacy",
      "product_quantity": "Total for all products 1,598 (USA 824, OUS 774)",
      "reason_for_recall": "GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors - Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems.",
      "recall_initiation_date": "20161121",
      "center_classification_date": "20161223",
      "termination_date": "20170926",
      "report_date": "20170104",
      "code_info": "Mfg. Lot or Serial # System ID  00000118203MO9 256235ORF1  00000118220MO3 256235ORF2  00000100096MO7 205655LRF  00000080072YY2 205783LRF2  00000119875MO3 205930LRF  Not Available 205801KRF4  00000026146YY1 205793FLEG  00000047927YY9 205793FR1  00000089624MO1 205877LRF2  00000047907YY1 205534RM2  00000094699MO6 256517EF2  00000089628MO2 256534RM1  00000113953MO4 205387LRF  00000034291YY5 334293JLD3  00000063480YY8 256386LD1  00000063481YY6 256386LD2  00000119881MO1 205345RF3  00000119916MO5 205554VAL  00000128623MO6 479553RF2  00000118229MO4 501520RF1  00000100110MO6 501622SJR4  00000071209YY1 501751NWX2  00000146291MO0 520432RM1  00000063528YY4 480264RF1  00000119856MO3 480821RM2  00000119873MO8 480821RM2  00000128466MO0 928729RM4  00000128464MO5 928729RM5  00000071297YY6 602588LEGACY  00000128589MO9 602588ROOM5  00000063529YY2 602856LEG  00000063617YY5 602835RM2  00000118196MO5 602835RM3  00000118194MO0 602239RMF  00000063638YY1 602879RM1  00000128507MO1 623445LEG  00000119871MO2 623879RM3  00000071252YY1 602994DRS  00000094664MO0 520742LEGACY  00000071291YY9 520742RM1  00000119898MO5 520872RM3  00000023695YY0 619242RF3  00000100087MO6 209386ATWRF  00000063549YY0 661282XR  00000063581YY3 650696M6  00000071217YY4 619725RM4  00000146295MO1 818884LRF1  00000063461YY8 916537MXF  Not Available 209992CP1  00000026093YY5 818331SFX  00000023850YY1 707443JR5  00000022135YY8 209221TCX  00000100070MO2 831395RF2  00000153568M01 858626RF6  00000153565M07 858626RF7  00000128636MO8 925734RF1  00000113872MO6 RADNETRF130  00000063584YY7 323669ADV  00000047971YY7 209225LEG3  00000052191YY4 336835HCR1  00000031396YY5 510370RFDG2  00000047998YY0 510372RF2  00000052100YY5 510372RF2  00000080068YY0 209577LEG  00000119878MO7 RADNETRF138  00000052169YY0 RADNETRF174  00000026132YY1 619485RM3  00000119892MO8 909788LF3  00000118227MO8 909788LF4  00000026094YY3 916423AVTX2  00000071284YY4 480860RM10  00000022205YY9 1443  00000094667MO3 619487RF1  00000146280MO3 619543RF4  00000063775YY1 415923LEGACYGI  00000071207YY5 415387C1  00000063587YY0 415387C2  00000047912YY1 415821SF1X28  00000063612YY6 415923PR5  00000031389YY0 408947LEGACY  00000113904MO7 408977RF5  00000071242YY2 RADNETRF122  00000031430YY2 707838SFX  00000031377YY5 707423TC  Not Available 707648KVRF1  Not Available 805648RM5  00000071200YY0 916662XF2  00000052168YY2 303695ROOM3  00000052165YY8 303695ROOM4  00000128486MO8 303440BCHRM4  00000089644MO9 303659RF1  00000031365770 719776RM5  00000039662YY2 303778RM3A  00000128577MO4 303398RF  00000022143YY2 303788RM9  00000128582MO4 303699RF2  00000128511MO3 303450RM2  00000063707YY4 860585RM3  00000153573MO1 203797LEG1  00000063670YY4 860358RF2  00000063518YY5 703697RF2  00000063501YY1 407830RF4  00000063503YY7 863533RM1  00000113913MO8 941798RM2  00000146285MO2 941798RM3  00000031454YY2 813681SPEC1  00000031395YY7 407566CH2  00000048093YY9 352394LG1  00000034296YY4 407434RF1  00000100099MO1 352689RF  00000128504M08 850883ER5  00000119904MO1 239931RM2  00000118228MO6 561468RM4  00000071205YY9 850862FR3  00000113905MO4 954985ME11  00000113880MO9 954987MOP4  00000146302MO5 352344RFLEG  00000128526MO1 239369RM3  00000034273YY3 407434RF2  00000063769YY4 850626SR4  00000119860MO5 NCHRFRM3  Not Available 813462RF1  00000071282YY8 352351RM2  00000119940MO5 352861RF1  00000052121YY1 904264LEGACY  00000052185YY6 407896RF5  00000039674YY7 904672LGY1  00000031333YY8 561622RF1  00000031361YY9 561622RF2  00000128472MO8 FHFGRF1  00000113895MO7 850747BMC1  00000113896MO5 850747ER1  Not Available 904769BRA2  00000071226YY5 850111RF1  00000128533MO7 321752WLG1  00000063496YY4 727845RF2  00000048016YY0 813725RF2  00000113874MO2 904819RF2  00000063782YY7 727825RM9  00000128482MO7 RADNETRF144  00000063778YY5 850325RF1  00000063708YY2 850656RF2  00000119888MO6 954724LEG  00000119876MO1 813828RF2  00000128592MO3 813872LEG1  00000071260YY4 941483RM2  00000023844YY4 407646RF6  00000048003YY8 770603LEG1  00000146293MO6 404459MMLEG  00000034230YY3 404851LEG1  00000031331YY2 404851LEG3  00000047923YY8 404851LEG5  00000100088MO4 404851STJOEL1  00000047930YY3 912246MR3  00000031346YY0 770382WCRF  00000031385YY8 404571DL2  00000119920MO7 404321LEG2  00000118209M06 706754HR2  00000119917MO3 478277VAR2  00000080089YY6 478633R3  00000047996YY4 478633R4  00000113914MO6 912757RF2  00000063507YY8 912327LD3  00000039672YY1 770677LEG1  00000034289YY9 912356LEG2  00000031358YY5 912386TR1  Not Available 912386TR5  00000026108YY1 404867BRF1  00000128451MO2 712476LEG  00000022131YY7 708259NWTC  Not Available 312567PR6  00000031341YY1 847297HF3  00000119889MO4 815284RM2  00000063589YY6 847570RF6  00000031335YY3 309343CRF4  00000026110YY7 630858GLD  00000094668MO1 847657GR4  00000034266YY7 618664URF2  00000063691YY0 708216GIRF1  00000026150YY3 708216L12  Not Available 708681WRF1  00000047940YY2 309347PRF1  00000113871MO8 309655FNCRF1  00000089648MO0 309655FXF11  00000063672YY0 217788LEG2  00000022211YY7 217544JRMF  00000031336YY1 217544RMI  00000128639MO2 864560BJWR1  00000071294YY3 217337CXF2  00000031307YY2 217383CRF1  00000119891MO0 847360RF1  00000119909MO0 812256RF2  00000034229YY5 765827LEG  00000094704MO4 219416ADV  00000128568MO3 260425SJRF6  00001285568MO3 260425SJRF6  00000128581MO6 317621LEGACY  00000063559YY9 812482LRF  Not Available 317662RM5  00000094693MO9 219284RM4N  00000100090MO0 219284RM5N  00000119901MO7 812237TH3  00000089620MO9 913684MLEG  00000128456MO1 620223MRF  00000113902MO1 913596PR3  00000146287MO8 316759LEGACY  00000047986YY5 913317SRM2  00000047999YY8 913317SRM3  00000026134YY7 913345MLEG1  00000048048YY3 316672PD1  00000128510MO5 785472RF  00000128542MO8 785743TRF  00000113958MO3 502348RF1  00000023691YY9 502781G2  00000031486YY4 606439LEG  00000031465YY8 606252RF5  00000063582YY1 414785RM3  00000052134YY4 606268FL2  00000128459MO5 859226RF2  00000119913MO2 859255RF1  00000080086YY2 502636AF1  00000118215MO3 502933RF1  00000128477MO7 606285LEGACY  00000113870MO0 606987LEG1  00000128469MO4 606354PRSN  00000080085YY4 502622RF1  00000089612MO6 606451LEG1  00000119899MO3 606365RF1  00000071283YY6 318473RD2  Not Available 318448COP1  00000119852MO2 985748RF  Not Available 225769RF1  00000118207MO0 504898HC2  00000023720YY6 318531FRM4  00000071253YY9 225222LEG  00000146312MO4 985652LEG  00000128611MO1 985882RF1  00000128619MO4 318253LEG  00000119931MO4 504885LEG  Not Available 504885RF1  00000023768YY5 318377MR3  00000063495YY6 318327RM7  00000026096YY8 318327RM8  00000128513MO9 318327RM9  00000026099YY2 318327SFCH2  00000048035YY0 318335RF2  00000094656MO6 318728RF1  Not Available 318255GRM1  00000119859MO7 225635RF1  00000034218YY8 318227SRM5  00000034225YY3 318675LR8  00000022216YY6 318798HR3  00000034268YY3 318798HRM4  00000048047YY5 318539LEG  00000128461MO1 504391RF1  00000128460MO3 504391RF2  00000031309YY8 318329GRM6  00000113915MO3 318435RF1  00000023749YY5 617355RF1  Not Available 617754RM2  00000022197YY8 617499RM3  00000071203YY4 978287ERF1  00000063813YY0 508548TC4  00000052179YY9 978796DEV  00000052123YY7 301981RF5  00000128508MO9 443849RF1  00000119851MO4 301896RF  00000128519MO6 301896RF1  00000063764YY5 410392R2  Not Available 301762RF1  00000022175YY4 301279RF2  00000080099YY5 301681RF  00000094687MO1 207262SRF3  00000094689MO7 207262SRF3  00000118199MO9 207373RF1  00000146279MO5 207255RF2  00000128642MO6 231547ADV  00000113876MO7 810421RF  00000063737YY1 248937LEG  00000146322MO3 313593LEG1  00000153564MO0 313593LEG2  00000113929MO4 313745RF4  00000128547MO7 FHEOLG1  00000113946MO8 517353LEGACY  00000089651MO4 517663LEGACY  00000100081MO9 616391LEGACY  00000100108MO0 989348DRS1  00000128505MO5 517545LEG  00000031393YY2 906524ADV1  00000063777YY7 517353DNR  00000128448MO8 517332LEGACY  00000128550MO1 231845RMB  00000063465YY9 248551RM12  00000113890MO8 517583HRF1  Not Available 734429F1  00000119893MO6 989362DRS1  00000153570MO7 313295LEG  00000071277YY8 507345LEG  00000022203YY4 319352WR2  00000023725YY5 612520N7  00000113942MO7 507332RF  00000063516YY9 651565RF  00000080094YY6 651232SM  00000119894MO4 507372LEG  00000094659MO0 417623FEL  00000023690YY1 816932SR7  00000113954MO2 314625SJW2  00000146320MO7 417235CMRF  00000023742YY0 314577USHRF1  00000146289MO4 417269NX4  00000047984YY0 417885SXL1  00000128574MO1 636239JRF5  00000100111MO4 636327CRL  00000047994YY9 601244OPRF1  00000118223MO7 601933LEG  00000048041YY8 228575XR1  00000071220YY8 228575XR2  00000071248YY9 228575XR3  00000128522MO0 601268W6  00000119862MO1 601364ARF3  00000034235YY2 601376LEG  00000128457MO9 601968COLRF1  00000146296MO9 601968LEG1  00000146282MO9 601968LEG2  00000023696YY8 601684SFX  00000146274MO6 318336LEG  00000128583MO2 228809SR2  00000118240MO1 662324RF1  00000118234MO4 601883RF1  Not Available 601636RF2  00000128524MO6 601735RF2  00000031380YY9 406563A03  00000031476YY5 406543SX2  00000063814YY8 828213MMR1  00000052154YY2 828255SJR5  00000071223YY2 919575XR50  00000063774YY4 704384RF2  00000094663MO2 704543GERF1  00000071240YY6 828452RF1  00000063472YY5 919684RM13  00000063460YY0 919684RM14  Not Available 919684EN3  00000026148YY7 910609HSLEG  00000119872MO0 828322CLEG3  00000071232YY3 919522LRF2  00000039658YY0 910249LRF1  00000113923MO7 919497LEG  00000047950YY1 336719NR151  00000094692M01 910215LEG2  00000094692MO1 910215LEG2  00000023706YY5 919733ADV  Not Available 919783RADV  00000031343YY7 919443NDRF2  00000118221MO1 919934JADV1  00000118218MO7 919934JADV2  00000113879MO1 252641HRLEG  00000118214MO6 910642CRLEG  00000031339YY5 910343NHR8  00000094691MO3 910452CFR2  00000100097MO5 402481RF2  00000063642YY3 603778RM2  Not Available 603332FRFSF  Not Available 801408ENDO  00000128528MO7 732321LEG6  00000039675YY4 908527RF3  00000039670YY5 908527RF4  00000118210MO4 908351RF  00000031398YY1 908821LEG  00000071225YY7 73222215  00000119942MO1 973365RM5  00000119865MO4 973971PED2  00000128498MO3 973972C309  00000080079YY7 973365RM4  00000034295YY6 201754LEG  00000047929YY5 201754RF6  00000047948YY5 201754RF8  00000146300MO9 973977TILTC  00000113948MO4 732442LEG  00000128629MO3 973831RF3  00000052166YY6 505841RM2  00000128470MO2 505722XRM1  00000052183YY1 505287ADVTX  00000128501MO4 505492RF1  00000128489MO2 505522RF1  00000128496MO7 RADNETRF146  00000071288YY5 702564RF1  00000118226MO0 702731SRF13  00000118225MO2 702731SRF7  00000031371YY8 516968LEG1  00000047954YY3 516968LEG2  00000034244YY4 718430RM3  00000048029YY3 718250LEGACY  00000022173YY9 718780RM2  00000023672YY9 718780RM4  00000063531YY8 718940LEGACY  Not Available 718780RM1  00000119926MO4 585393VRF2  00000031392YY4 718738RF  00000063723YY1 718883RF2  00000063724YY9 718883RF3  00000063452YY7 315764MRF1  00000063732YY2 914666DRS  00000063564YY9 212463LEGACY  00000034261YY8 315393RF1  00000023837YY8 516474LEGACY  00000048039YY2 631476RM1  00000089647MO2 516766RF  00000063702YY5 518583RF2  00000128458MO7 740425RF1  00000063800YY7 513686LEG  00000063489YY9 937268RM120  Not Available 614446HX4  00000071259YY6 419227RM2  00000113899M09 330837R4  00000063701YY7 740393LEG2  00000128590MO7 419698RF1  00000128549MO3 419698RF2  00000113945MO0 419734LEGACY  00000128616MO0 330332R18  00000048033YY5 4194475  00000071233YY1 330841R5  00000100084MO3 740454RF3  00000048095YY4 580421VRF1  00000071245YY5 405359ER3  00000039684YY6 918682MRF2  00000128618MO6 405307NRF2  00000048079YY8 405271URF2  00000063509YY4 405636SRF6  00000048064YY0 918294SCRF3  00000063511YY0 918294SCRF2  00000119890MO2 580772RF1  00000113886MO6 580256WRF1  00000118222MO9 405717CVRF1  00000063513YY6 541677MOR3  00000023698YY4 215776LRF3  00000063652YY2 717267RF6  00000026162YY8 610826FL2  00000023838YY6 215221CL1  00000034254YY3 215707LEGACY  00000128471MO0 412624F1  00000034293YY1 717340LD2  00000047955YY0 610378RF5  00000063605YY0 610988RF6  Not Available 215881RF1  00000071201YY8 401274WILEG  00000128565MO9 401737KLEG  00000063710YY8 401596WLEG  00000031399YY9 864716RM2  00000052153YY4 843674RF3  00000052152YY6 843674RF4  00000119877MO9 864487UPR2  00000113906MO2 843527GRF4  00000118233MO6 864255SFOP3  00000118232MO8 864255SFOP4  00000118195MO7 864295GRRM4  00000063544YY1 843293RF  00000048019YY4 605352LEG  00000071228YY1 507537LEG  00000047938YY6 423493PRF3  00000023682YY8 423495MX10  Not Available 423495MX15  00000052192YY2 901853RF1  00000052198YY9 901853RF2  00000031349YY4 901285RA  00000063546YY6 901754G1  00000023769YY3 901757R4  00000128576MO6 423636RF  00000071296YY8 615264H4  00000063773YY6 731425GI  00000094678MO0 901424RF1  00000047997YY2 423837GRF1  00000094698MO8 423857RF5  00000026174YY3 615473RRF  00000034279YY0 901369GE1  00000119849MO8 901384OPRF  00000063610YY0 901523VA5  00000063648YY0 901523VA7  00000031348YY6 901726M21  00000031485YY6 615781LEG  00000071214YY1 615769SKYRF3  00000071212YY5 615769SKYRF5  00000071285YY1 615769SKYRF7  Not Available 615356RM2  00000128584MO0 865481RF3  00000063614YY2 931967SRF  00000063536YY7 512660RF1  00000071289YY3 806355RM2  00000100098MO3 817472LEG1  00000113885MO8 817472LEG2  00000146284MO5 512314AS1  00000146318MO1 512448PORF  00000031327YY0 512476SRM6  00000034267YY5 409838ORF  00000052190YY6 409838ORF2  Not Available 409842BRF  00000128490MO0 903583VADIG  00000034221YY2 903583VAL  00000080065YY6 409680BVPRF1  00000119914MO0 979776SJRF1  00000031475YY7 409764BE6  00000128570MO9 361985RF2  00000094677MO2 903654RF1  00000039677YY0 972376VARF1  00000023673YY7 401354  00000113931MO0 214820RF1  00000080088YY8 214827RF  00000094695MO4 214947RF4  Not Available 401353  00000063735YY5 940898ENDO  00000146315MO7 956388RF1  00000052144YY3 915533RF1  00000052126YY0 915595RF4  00000063675YY3 817882ENDO  00000071263YY8 817551HMLEG4  00000063629YY0 INSIGHTXR041  00000119874MO6 956389RF5  00000022070YY7 713527PRF  00000128480MO1 281618LEG  00000113965MO8 713455LEG1  00000113959MO1 713455LEG4  00000146301MO7 713527PRF1  00000113884MO1 409539LEG2  00000052131YY0 817498LEG  00000031374YY2 972751RF  00000146297MO7 281599LEG  00000128512MO1 806725B1  00000128525MO3 806725B2  00000119854MO8 806792LSRF4  00000119857MO1 806792RF10  00000128638MO4 856346RF  Not Available 956632RM1  00000063809YY8 972548ENDO  00000063810YY6 972548ENDO  00000052141YY9 972548RF2  00000052151YY8 972548RF3  00000089630MO8 214320RF1  00000128637MO6 915685ADV3  00000048075YY6 817255NHLEG  00000026083YY6 915334LEGACY  00000089642MO3 806435RF1  00000063745YY4 972596GI  00000063748YY8 972596GI  00000128569MO1 361552LEGD1  00000047943YY6 210575CRF2  00000146288MO6 210614NRF1  00000119869MO6 210614NW1  00000128532MO9 210614RF1  00000146314MO0 210614SRF1  00000063510YY2 210704SRP1  00000113932MO8 210705SRMC4  00000146305M08 817424GEXR1  00000071281YY0 409938ADTX2  00000128562MO6 936321LEG1  00000128575MO8 936321LEG2  00000128463MO7 907353RM3  00000080064YY9 281338CL5  00000128606MO1 435587RM1  00000023757YY8 801581RM7  00000128599MO8 571223RF  00000048034YY3 540236ARF1  00000071222YY4 804541JR2A  00000113878M03 804861FDOC  00000113878MO3 804861FDOC  00000128596MO4 540862ARF6  00000063603YY5 540743RF  00000063464YY2 757886RM3  00000023818YY8 804466RM10  00000063607YY6 757671RM1  00000031443YY5 540546ARF  00000047947YY7 703478RF  00000071204YY2 703689RM02  00000128453MO8 804289HDF4  00000128631MO9 804747DHP1  00000128585MO7 804747HDP4  00000119919MO9 540462ARF  00000063754YY6 434447CMRF1  000000128520O4 757673RM1  00000119848MO0 757934RM6  00000128626MO9 757363RM3  00000022196YY0 206385RM1  00000063532YY6 509924VL  00000094690MO5 425883FLUORO1  00000094700MO2 509575PYX4  00000063742YY1 864261RM15  00000146321MO5 512454CNTRF  00000113963MO3 310289LEGACYD  00000048057YY4 419354LEGACY  00000052186YY4 781273L2  00000128479MO3 406782SX3  00000094688MO9 717SHCRF1  00000128500MO6 662627RF4  00000039688YY7 214376VARF2  00000034203YY0 214345PRF  00000113968MO2 859236RF2  00000153567MO3 DELANO2LEG1  00000023708YY1 313576RM3611  00000048072YY3 580251DSP  00000031359YY3 610250FL2  00000063550YY8 409772TDC1  00000031367YY6 864675SFWLEG  00000063462YY6 601268W5  00000048070YY7 713442KDRF2  00000063494YY9 713455LEG  00000113899MO9 330837R4  00000146290MO2 414963RF1  00000048088YY9 414266RM5  00000048077YY2 414289GI1  00000146303MO3 414527ADV1  00000100072MO8 510869LEG6  00000119903MO3 732970LEGACY  00000063524YY3 913676SR9  00000047972YY5 941629RF2  00000022100YY2 414636ENDO1  00000100082MO7 262321RF1  00000071234YY9 262687RF112  00000063643YY1 414636LEG6  00000100105MO6 RADNETRF129  00000048020YY2 904825LEGACY  00000052116YY1 210575CRF1  00000100106MO4 805739LG  00000100109MO8 707576RF3  00000113917MO9 707576RF1  00000022110YY1 715346TC1  00000063519YY3 409822TRF  00000071229YY9 580421VRF1  00000071219YY0 281218HSLXXX  00000048046YY7 940764RMCXXX  00000071271YY1 570829RF4VA  00000052140YY1 630682CDH4  00000094673MO1 304256RX1A  00000146294MO4 304256RX3A  00000063450YY1 304348SX2  00000128481MO9 304462LEG  00000119858MO9 304424RF1  00000089613MO4 304235WX1B  00000063711YY6 307332RF  00000113937MO7 307332RF  00000089652MO2 780624RF  00000063705YY8 306778RF1  00000113881MO7 EG1112RX07  00000026208YY9 GOLD001109  00000026123YY0 HC0887XR01  00000063557YY3 XR545633WK0  00000063682YY9 0850080428  00000048038YY4 0850080430  00000052119YY5 0530867LEG  00000048068YY1 RU1556RX03  00000048074YY9 RU1488RX13  00000063538YY3 RU1488RX14  00000048078YY0 RU1488RX17  00000048076YY4 RU1488RX19  00000022086YY3 YU1001VA02  00000031418YY7 AE1009RX01"
    }
  ]
}