  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70046",
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "address_1": "700 Orthopaedic Dr",
      "address_2": "N/A",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including states of: AZ, CA, CO, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MT, NC, NV, NY, OH, PA, TN, TX, VA, WA, and WI; and countries of: Canada, Australia, Austria, Belgium, Czech Republic, France, Germany, India, Israel, Italy, Japan, Malaysia, Russia, South Africa, Switzerland, and  United Kingdom.",
      "recall_number": "Z-0908-2015",
      "product_description": "I.M. Hole Locator Instrument.  Orthopedic manual surgical instrument.        The IM Hole locator guides the IM drill during IM Hole preparation. The instrument is positioned between the condyles and aligned with the anatomical axis. Pins can be placed in the IM hole locator to assist with additional stability. The drill bush may also be lightly impacted onto the distal femur for additional stability.",
      "product_quantity": "1130",
      "reason_for_recall": "Use of excessive force when impacting may lead to intra-operative femoral fracture.",
      "recall_initiation_date": "20141112",
      "center_classification_date": "20150105",
      "termination_date": "20150224",
      "report_date": "20150114",
      "code_info": "CATALOG NO. : 9505-01-041  Barcode GTIN: 10603295224389  ALL LOTS",
      "more_code_info": ""
    }
  ]
}