  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Concord",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64214",
      "recalling_firm": "Siemens Medical Solutions USA,  Inc",
      "address_1": "4040 Nelson Ave",
      "address_2": "N/A",
      "postal_code": "94520-1200",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "5 devices, all located outside of the United States in Germany.",
      "recall_number": "Z-0907-2013",
      "product_description": "SIEMENS brand ARTISTE syngo RT Therapist (RTT) version 4.3, ARTISTE Model, Part No. 08162815, Digital linear accelerators, MOSAIQ OIS system, Distributed by SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Manufactured by SIEMENS AG, MEDICAL SOLUTIONS, GERMANY.",
      "product_quantity": "5 units",
      "reason_for_recall": "Siemens became aware of a potential safety issue when using the \"In-session resumption\" feature of the syngo RT Therapist (RTT) 4.3 software. The RTT will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.",
      "recall_initiation_date": "20130117",
      "center_classification_date": "20130305",
      "termination_date": "20130305",
      "report_date": "20130313",
      "code_info": "All codes/units of ARTISTE model."
    }
  ]
}