{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Braintree",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84183",
      "recalling_firm": "Haemonetics Corporation",
      "address_1": "400 Wood Rd",
      "address_2": "N/A",
      "postal_code": "02184-2412",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB,  IT, CH.",
      "recall_number": "Z-0905-2020",
      "product_description": "Japan DISP SET, 263J, 205J, 208 - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma",
      "product_quantity": "14271",
      "reason_for_recall": "Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.",
      "recall_initiation_date": "20191021",
      "center_classification_date": "20200129",
      "termination_date": "20201106",
      "report_date": "20200205",
      "code_info": "Catalog number/FG Item Number: 263J-SET    1. Specific Lots with Bowl Cracks (1382 Distributed)  FP Lot number: 0718027-0119029-190124H2; 0718027-0119029-190125H1; 0718027-0818085-181030H1; 0718027-0818086-181030H1; 0718027-1018006-181224H1; 0718027-1018007-181224H1; 0718027-1018008-181225H1; 0718027-1018009-181226H1; 0718027-1118031-181226H1; 0718027-1118032-181226H1; 0718027-1118081-190124H1; 0718027-1118083-190124H2    2. Missing instructions to check for cracks in IFU  All products/lost manufactured between  October 2014 and October 2019."
    }
  ]
}