{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70043",
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "address_1": "700 Orthopaedic Dr",
      "address_2": "N/A",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including P.R. except AK and RI; and Internationally to: Chile, Canada, Australia, Austria, Belgium, China, France, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, United Arab Emirates (UAE), and United Kingdom (UK).",
      "recall_number": "Z-0905-2015",
      "product_description": "Femoral Impactor ATTUNE INTUITION Impactor.  The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee replacement procedures. The three affected impactors are used at various times within the surgical flow, each impactor is used to impact and drive home its relevant mating part: Fixed Tibial bearing, (254401003) Rotating Tibial Platform, (254401004) and Femoral impactor, (254401006).",
      "product_quantity": "4661",
      "reason_for_recall": "ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.",
      "recall_initiation_date": "20141110",
      "center_classification_date": "20150105",
      "termination_date": "20151208",
      "report_date": "20150114",
      "code_info": "CATALOG NO. : ATTUNE INTUITION Impactors   Femoral Impactor, (254401006)   Barcode GTIN Number:   Femoral Impactor, 10603295130222    ALL LOTS",
      "more_code_info": ""
    }
  ]
}