  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "South Jordan",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89815",
      "recalling_firm": "Merit Medical Systems, Inc.",
      "address_1": "1600 W Merit Pkwy",
      "address_2": "N/A",
      "postal_code": "84095-2416",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: MN, PA, AZ, FL, IL, AK, NY, CA, TX, KY, WI, NC, NJ, VA, CO, IN, AR, MO, LA, OR, MT, MA, MI, NE, WV, KS, MS, HI, IA, NH, ME, AL, DC, DE, NM, OH, GA, MD, TN, UT, RI, SC, OK, WA, ID, ND, SD, NV, VT.  OUS: Poland, Germany, France, United Kingdom, Belgium, Korea, Republic of, Austria, Canada, Spain, Luxembourg, Italy, Switzerland, Czech Republic, Netherlands, Japan, Honduras, Colombia, Greece, Ireland",
      "recall_number": "Z-0903-2022",
      "product_description": "Access-9 Large Bore Hemostasis Valves, Catalog Code: 580775001, IN8112/C, K05-00053H, K05-00109G, K05-00137K, K05-00410G, K05-00438G, K05-00502F, K05-00587D, K05-01052, K05-01706C, K05-01880K, K05-02130, K05-02595, K09-12599A, K10-05487P, K12-00264, MAP111/F, MAP112/B, MAP112/F, K08-YP085A, K08-YS017B, K08-YS055A, K09-YS029A, K09-YS040F, K12-YP002A, K12-YP030A, K12-YP060A, K12-YP061A, K12-YP063C, K12-YP072, K12-YS018A, K12-YS019A, K12-YS026A, K05-01140D, K05-01706C***, K05-02595, MAP112, K05-00564L, K12-YP002A,",
      "product_quantity": "36445",
      "reason_for_recall": "A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire through the device. This defect may result in a delay of procedure.",
      "recall_initiation_date": "20220301",
      "center_classification_date": "20220411",
      "report_date": "20220420",
      "code_info": "Catalog Code/Lots:  580775001/I2219771, I2278741, I2226790; IN8112/C/H2292357, H2287440, H2281404, H2298626, H2305047, H2305048, H2309356; K05-00053H/H2305217; K05-00109G/H2304656; K05-00137K/H2304653; K05-00410G/H2305216; K05-00438G/H2309130; K05-00502F/H2309122; K05-00587D/H2309211; K05-01052/H2304607; K05-01706C/H2309133; K05-01880K/H2306962;  K05-02130/H2304639; K05-02595/H2304614; K09-12599A/H2295450; K10-05487P/H2279857; K12-00264/H2285040; MAP111/F/I2259670, I2279884, I2259657, I2289610; MAP112/B/P1908667; MAP112/F/I2259661, I2279915, I2279894, I2176725, I2215955, I2289611, I2279895, I2289613, I2289612; K08-YP085A/H2249051, H2302394; K08-YS017B/H2249052; K08-YS055A/H2275990; K09-YS029A/H2266470, H2302474; K09-YS040F/H2286601; K12-YP002A/H2269321, H2278279, H2283742, H2295661; K12-YP030A/H2263653; K12-YP060A/H2272013; K12-YP061A/H2249088; K12-YP063C/H2272382, H2295556; K12-YP072/H2266224; K12-YS018A/H2284956; K12-YS019A/H2302506; K12-YS026A/H2276014, H2302507; K05-01140D/H2304611; K05-01706C/ (H2304610, H2309133)***; K05-02595/H2304614; MAP112/I2279895; K05-00564L/H2304617; K12-YP002A/H2295661;"
    }
  ]
}