      "status": "Terminated",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86646",
      "recalling_firm": "Respironics California, LLC",
      "address_1": "2271 Cosmos Ct",
      "address_2": "",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia.",
      "recall_number": "Z-0901-2021",
      "product_description": "Philips Respironics V680 Ventilator Part Number 850011 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.",
      "product_quantity": "3,453 systems",
      "reason_for_recall": "Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.",
      "recall_initiation_date": "20201030",
      "center_classification_date": "20210122",
      "termination_date": "20240919",
      "report_date": "20210203",
      "code_info": "Model: V680 Ventilator, Part Number 850011  V60 Ventilator, Finished    Serial Numbers:  100138084  100138519  100138520  100138521  100138705  100241765  100241766  100241770  100241771  201014177  201014216  201020752  201020756  201020757  201021264  201021272  100260220  100302057  100302058  100305132  100305134  100305349  100306061  100306064  100306070  201025414  201026091  201026102  201026480  201026481  100142436  100183705  201013216  100148515  100220768  100148513  100180915  100220764  201010433  201028445  201028451  100129149  100219325  100220765  100147302  100138085  100138509  100138511  100138517  100147215  100219037  100219322  100219332  201032777  201032778  201032844  100181643  201028444  100191242  100191243  100191244  100191245  100191246  100191247  100191249  100191250  100191251  100191252  100191253  100191254  100191255  100191256  100191257  100191259  100191321  100191325  100191326  100192522  100218827  100220780  201026491  201026492  100125151  100125515  100125547  201008782  201008765  201008770  201008776  201008805  100305341  100174507  100174510  100175284  100175286  100175291  100175292  100175326  100175327  100175330  100175331  100175332  100175335  100175337  100175339  100138510  100138013  100134076  100146376  100146377  100146440  100146443  100146872  100147218  100147300  100183859  100183860  100290724  100306068  100312192  201014329  201020751  201020753  201020754  201027465  100230813  100302060  100304976  100304977  100305133  100305340  100305342  100291620  100297028  100297029  100292723  100296858  100296860  100296864  100219029  100220779  100221895  100221902  100221910  100222638  100251606  100298531  100263992  100263998  100264021  100264300  100307212  100307542  100307546  100307547  100307557  100308425  100308427  100324465  100329190  100329194  100329195  100329426  100329435  100232323  100232726  100232730  100232731  100232734  201033794  100250079  100297626  100298482  100316029  100316031  100329617  100316229  100324353  100324357  100324461  201032309  201032440  100297630  100264025  100264574  100264576  100244932  100218834  100324468  100328677  100329438  100329607  201032151  201032290  100347363  100347336  100347349  100347335  100255873  100253331  100253328  100253122  100253120  100250480  100227939  100226497  100264301  100264566  100264568  100264569  100264572  100264573  100304942  100296198  100324358  100328678  100328722  100329187  100329420  100329609  100329614  100297026  100302052  100301098  100297024  100257488  100264570  100264575  100291623  100291625  100292460  100292463  100253123  100253127  100255957  100263993  100263994  100264302  100264563  100264571  100222838  100248325  100263999  100264303  100264561  100264562  100265624  100265628  100291629  100291637  100292459  100264564  100264565  100264567  100304601  100304978  100304979  100305345  100305346  100305347  100305348  100305351  100305352  100305353  100308601  100308607  100308608  100308612  100221900  100227960  201033110  201034950  100249212  100247868  201026096  100229267  100231784  100329611  201013717  201014543  100231671  201027860  100153252  201037404  100229440  100229443  100229446  100230807  100230812  100230856  100218821  201031231  201013887  201027852  201028142  100249800  100249812  100249819  100251607  201026103  201026485  201026488  201026490  201027738  201027851  100229266  100265642  100296199  100296865  201030354  201027463  201027846  201027464  201027721  201027726  100244915  100244916  100244919  100244330  100244336  100244914  100244917  100244918  100290729  100291633  100192524  100192525  100192531  100193335  100193336  201027013  201027865  201028146  201028151  201028153  201028163  201028439  201037307  100227962  100229276  201027720  100183319  100183403  201024850  100241769  100230810  100230817  100230821  100147210  100264022  201027021  201028446  100240122  100240123  100227975  201033109  100231785  100232723  100328721  100251605  100253117  100281278  100290715  100290716  100290717  201031239  201032045  100260224  100260625  100279976  100277113  201026099  100183568  100183602  100183719  100183723  100183840  100183853  201026080  201027020  201027728  100307210  100307548  100249208  100218835  100146441  201027028  100251613  201034918  100161111  100175329  100312197  100312198  100193340  100193344  201027024  100220772  100221893  100288584  100289774  100289777  100289778  100289781  201037292  100253341  100253128  100253330  100253333  100253336  100253340  201032152  100222820  100224297  100227965  100176450  100176820  100248342  100248340  100249207  100190529  100190533  100191248  100191320  100192523  201021473  201021474  201021487  201027456  201027457  201027458  100190780  201025426  201026500  201026677  201024841  100222830  100222835  100227978  201026086  201026089  201026496  201032296  201037293  100249808  100249810  100249813  100249814  100249817  100249818  100250064  100250488  100250489  100250490  201028742  201030358  201030600  201030602  201027854  201028145  100253125  100253299  100290727  100194372  100194420  100195382  100219323  100183567  100193338  100195306  100265630  100265634  201023446  201023456  201023458  201023461  201026085  201026088  201026090  201026093  100219036  201032779  201032780  100218839  100218840  100218841  201030357  201030359  201030601  201030604  201030606  201030611  201032153  201032294  201032298  201032311  201032441  201032445  201032447  100329079  201030608  100224291  201033135  201027733  100177146  100190783  100265639  201023061  100240077  100241298  100183852  100230819  100230857  100231131  100260228  100260229  100260617  100263996  100264026  100302050  100302051  201031440  201032044  100183846  201027025  201027027  100225764  100218817  100226494  100297030  100263995  100263997  100264020  100264023  100264024  100264279  201027724  100253332  100265629  100309855  201034946  201035197  201035222  201035226  201027723  201027725  201027729  100264578  100265625  201032297  100219030  100219033  100219122  100219329  100220781  100221889  100231134  100249205  201037271  100146869  100253329  201026487  201028449  201034917  201034919  201035085  100251615  201030614  201031225  100191324  201034949  201035330  201035198  201035199  201035221  201035321  201035015  100277102  100292467  100239797  100240143  100264783  100146870  201027026  201027842  100218836  100276773  100276775  201027867  201033115  100253111  201027730  201035223  100249206  201013713  201013749  100232732  201020758  100251604  100251608  100251610  100251611  100251612  100251614  201031233  201031234  201031236  201031443  201025421  201028458  100147219  100190535  201028598  100244332  201021225  201021267  201037408  100264780  100264785  100170454  100312021  100239261  100306063  100306065  100306066  100306067  100306071  100316036  100316037  100125544  100169380  201014214  100125541  100255874  100257497  100257498  100257501  100258064  100260223  100260627  100260631  100309858  201015021  201028452  100163786  100176444  100232725  201014341  201024840  201025424  201014343  201014344  201008762  201008773  201008781  201008809  201016926  201025428  201026489  201026495  201026674  201026676  100162310  100179819  100160962  100176759  100180008  201014185  201014199  100146867  100155777  100160499  100168869  100168870  100168874  100231416  100264577  100264579  100264781  100296855  100297021  201014384  201014385  201014535  201026092  201026095  201026097  201028586  201028587  201028591  100180023  100139078  100139087  100180012  100180013  100139066  100139068  100139069  100139070  100139071  100139074  100139075  100139077  100139079  100139080  100139082  100139083  100139084  100140238  100140239  100140300  100140301  100140302  100140303  100140304  100140305  100140310  100140311  100168812  100168813  100168815  100168816  100168819  100168866  100168873  100169141  100169143  100169145  100169147  100169149  100169150  100169151  100169152  100169153  100169154  100169155  100169156  100169157  100169158  100169159  100170298  100170301  100170302  100170379  100170440  100170441  100170442  100170447  100277179  100277324  100277617  100277622  100277625  100277626  100277628  100277631  100279971  100279972  100279973  100279975  100279977  100279978  100279979  100279981  100279982  100279983  100279985  100279986  100279989  100279990  100279991  100279997  100279999  100163783  201028140  201028441  100280000  100316221  201027849  201028161  201028443  201028447  100163045  100180014  201028160  100165983  201028143  100289769  100290728  100291621  100291630  100291631  100292461  201013277  100224293  100224296  100225757  201027022  201025418  100142435  100142638  100179320  100179996  100179998  100179999  100180003  100180004  100180005  201013274  100165986  100296853  201028459  100162259  100222636  201010431  201010489  201013276  100180016  100180024  100180323  100222634  100227976  100229270  201016704  201014181  201014184  201014217  100165991  100324351  100324354  201024849  100170460  100308604  100180000  100180002  100180010  100180015  100180017  100180018  100180019  100180020  100180324  100180326  100125513  100142634  100162313  100142640  100194370  201023451  100160663  100276781  100162304  201016719  100142641  100163249  100168818  100219328  100221904  100222635  100222825  100222828  100222829  100222837  100224292  100225755  100225759  100225763  100226492  100227937  100227961  100227963  100227966  100227967  100230820  100230855  100230858  100231132  100231418  100231460  100231461  100231462  100231463  100231464  100231664  100231665  100231666  100231667  100231668  100231669  100231670  100231672  100231783  100231786  100232321  100232324  100232325  100232326  100232327  100232328  100232330  100232720  100232722  100232727  100232728  100232729  100232733  100232736  100238558  100238559  100238580  100238581  100238582  100238583  100238584  100238585  100238586  100238587  100238588  100238589  100238590  100238591  100238592  100238593  100238594  100238595  100238596  100238597  100238761  100238763  100253124  100276785  100276787  100277112  100277114  100277119  100277123  100277124  100277125  100277126  100277128  100277129  100277131  100277164  100277165  100277166  100277167  100277168  100277169  100277170  100277171  100277172  100277173  100277174  100277175  100277176  100277178  100277320  100277321  100277322  100277323  100277327  100277328  100277329  100277330  100277612  100277613  100277614  100277615  100277616  100277618  100277619  201033106  201033107  201033108  201033114  201033116  201033795  201033796  201033797  201033798  201033799  201033907  201033909  201033910  201033911  201033912  201033913  201033914  201033915  201033916  201034040  201034041  201034042  201034049  201034050  201034051  201034052  201034053  201034092  201034093  201034094  201034095  201034096  201034097  201034098  201034099  201034160  201034161  100220767  100220770  201024846  100183304  100183355  100180001  100180007  100281273  100195378  100142428  100154350  100169142  201032444  201032310  100142460  100176825  100312320  201027848  100142637  201033118  201024851  100253121  201021270  100140313  100142426  100142434  201028448  100142432  100227970  100142429  100245757  201021476  201021477  201021481  201021482  201021483  201021484  201021485  201021486  201028141  100176808  100176817  100176822  100154721  100154722  100142431  100183722  100183856  100138083  100176182  100161107  100244931  201024845  201024847  100257496  100257500  100258060  100142437  201013219  100154345  100289768  100289775  100289782  100289783  100290710  100290711  100290712  100290713  100265637  100176812  201011219  100142427  100160496  100134077  100161847  100161848  100162258  100162300  100162306  100162307  100162308  100163248  100324462  100324355  100316230  100316228  100249204  100329619  100346834  100346843  100347201  100347202  100347330  100347331  100347333  100347345  100347346  100347348  100347351  100347352  100347354  100347357  100347358  100347364  100347365  100347366  100347371  100264784  100265636  100276782  100277122  100277127  100279988  100291622  100291628  100291632  100296197  100296851  100296854  100298529  100298532  100298533  100312196  100347206  100347341  100347353  100347355  201035018  201038943  201038945  100298534  100239518  100298500  100298526  100307219  100308518  201037311  100182074  100194368  100194373  100194424  100194426  100194429  100195308  100229314  100229315  100229316  100229317  100229318  100229319  100229441  100229442  100229444  100229445  100298535  100298620  100301143  201023070  201023071  201023072  100347334  201014197  100138518  100180022  100180322  100226493  100170459  100221913  100246953  100248347  100248348  100249203  201037278  201037288  201037310  201037312  201037407  100195385  100229280  100229281  100229312  201013715  201013724  201023062  201033117  201013748  201013893  100137427  201014322  100191258  201023441  100244928  100231130  100222823  201035337  201038950  201038951  100174504  100241295  201028593  201028594  201014179  201014182  201014183  201014211  201014213  201038853  201038855  201038858  100176803  100176811  100156105  100139072  201012734  201012736  100135404  100135633  100162302  100163065  100163244  100163246  100163247  100163792  100292458  100253324  100135648  100154346  100239268  100247867  100135399  100135408  100246954  201026484  201037309  100174503  100175290  100238762  100238766  100238767  100238769  100238770  100238931  100238932  100238933  100238934  100238935  100238936  100238937  100238938  100238939  100239262  100239264  100239265  100239266  100239270  100239272  100248324  100248327  100248329  100248330  100249803  100249804  100249805  100249806  100249807  201009427  201009494  201009495  201009496  201009498  201009499  201009661  201009662  201009663  201009666  201009667  201009672  201009675  201009676  201009677  201035017  100220786  100229271  100229273  100229274  100229275  100229277  100229278  100229279  201022775  201022777  201023444  201023452  201023453  201023455  201023459  201023460  201032295  201014324  201014342  201035080  100220775  100138522  201014176  201025427  201026094  201026101  201028589  201028590  100174264  201016920  100135642  100301140  100169146  100169148  100292465  100292466  100292469  100329429  100329436  100329610  100155779  100222836  100225758  100249815  100249816  100301094  100135637  201013714  100301087  201038859  201013711  201013716  201013744  100253113  100165988  100165989  100165990  100165992  100312191  100298481  100301092  100301141  100301142  100311613  100314607  100314608  100314609  100314611  100314614  100314617  100314862  100314872  201023445  201023450  100291634  100146373  100146873  100244929  100246948  100246950  100246952  100246956  100247849  100247855  100250483  100250486  201035016  201010483  100162314  201028592  201028595  100290723  100169381  100170300  100170444  100298487  100231699  201033908  100161118  100180320  100181644  100304941  100328719  100328780  100328784  100329437  201031232  100231417  100248332  100248335  100248337  100248338  100239267  100240144  100250067  201013230  201013231  100161115  201009674  100227938  100139088  201013213  100298485  100138087  100247858  100229272  100194431  201023440  100138016  100176827  201013710  201013753  100176657  100253112  100163788  100290725  100314610  201013889  201013891  100176813  100135396  100137446  100253115  100253116  100297033  100314868  100328699  100138506  100138516  100292455  100292456  100221908  100328676  100301086  100231133  100247864  100247865  100176802  100180011  100329421  100329608  100312027  201026497  201026498  100257494  100138082  100138507  100168814  100160975  201035334  100324352  100135403  100135640  100135643  100135652  100129142  100129143  100135641  100135647  100153447  100154697  100154699  100156202  100156204  100156205  100160493  100160494  201010254  201010535  201010546  201010674  100135644  100137439  100137444  100241293  100241296  100241762  100308441  201027007  201027009  201027011  201027012  201027014  201027015  201027016  201027017  201027023  100160490  201013890  201014196  201014215  201014218  100193325  100193337  100193339  100193341  100153448  100176824  100137435  100137442  100137443  100137445  100175364  100180328  201022773  201028746  100169500  100175289  100175325  100175368  100175372  100243771  100243775  100243778  100248331  100248333  100246946  100170456  100174262  100190534  100190536  100192526  201017222  100137441  201038953  100138086  100138512  100138513  100138515  100221890  100221896  100222821  100222831  100239517  100240140  100240141  100176807  100170461  100174095  100174096  100174098  100239275  201008784  201023074  100314869  100247857  201009429  201009493  201009660  201009668  201009671  201009678  100221906  100239273  100314615  100163049  100163050  100163051  100163064  100163067  100163068  100163242  100163245  100163250  100163357  100163359  100163420  100163421  100163782  100163784  100163787  100163791  100290714  100290719  100298539  100301144  100125010  100125516  100125517  100126254  100129144  201008807  201010428  201010675  100129141  201010482  201010488  201010551  100183858  100230818  100231124  100137429  100242971  201012739  100154349  100154352  100176809  201015022  201015023  100242962  100242963  100242973  100243770  100247863  100250487  100222824  100222834  100153449  100242969  100242972  100249809  100183309  100160967  100161116  100161117  100134337  201028152  201028162  201028442  201028456  100125518  100129147  201010255  201010360  201010434  201010485  201010538  201010623  100281277  100176445  201021266  100308429  100147211  100170446  100170451  100170453  100170457  100160495  201027732  201027859  201027861  201027862  100297627  100277629  100314759  100137440  100138504  100139085  100140316  100156200  201012732  201013215  201013223  201023065  100176826  100301093  100153451  100154696  100154698  201010439  201010627  201035332  100170452  100174097  100174099  100174261  100176451  100224294  100175365  100137430  100137449  100146369  100147209  201014340  100138018  100146370  100146371  100170443  100170445  100170450  100170455  100175369  100183855  100219031  100220773  100220784  100220785  100220787  100221891  100221894  100221897  100221899  100221901  100221903  100221909  100221911  100221912  100222639  100222817  100222818  100222819  100222826  100222832  100224295  100225761  100226491  100226495  100226496  100226499  100226520  201014321  201014325  201014327  201028157  201028597  201032950  201032951  201032952  201032953  100292719  201032291  201032293  201032442  201032443  201028740  100329430  100125510  201008785  201008801  201009664  201009670  201008771  201008786  100175288  100175363  100161108  100161114  100183527  100183529  100183530  100183532  100183535  100183572  100183608  100183842  100183847  100183849  100183850  100161119  100161120  100218818  100218820  100218822  100218823  100218824  100218825  100218826  100218828  100218829  100219117  100219119  100219120  100219121  201031226  201032049  201032050  201032051  201032052  201032053  201032147  201032148  201032149  201032150  201032289  201032312  201032313  201032315  201032316  100301091  100301096  100292464  100246960  100125504  100125542  100125545  100125551  100179997  100124645  100124646  100125009  100125011  100125012  100125013  100125014  100125148  100125149  100125150  100125152  100125153  100125154  100125155  201008740  201008742  201008774  201008777  201008802  201008804  100174395  100183857  100291599  100292468  100146442  100146875  100170449  201038948  100314871  100301097  100308609  201027727  100232735  100231123  100231126  100241768  100230859  100231121  100231122  100238768  100125552  100177145  100308434  100135397  100135653  100160661  100246957  201026100  201028596  201030610  100241299  100241300  100241301  100241764  100277623  100277624  201035083  201035361  100232329  100232721  201034043  100231419  201014198  100218831  100218832  100218838  100219032  100219035  100219039  100219326  100220776  100220777  100220778  100220782  100220783  100154724  100168867  100190539  201010490  100135398  100291635  201035082  201035363  100324464  100324466  100324467  100324469  100328720  100170458  100328698  100329423  100329615  201038949  201024844  201013712  201013750  100140306  100160664  100174260  100174263  100174392  100170299  201023073  100221898  201027019  201027731  201027734  201027739  201027840  201027843  201027844  201027845  201027847  201027850  201027856  201027857  201028148  201028149  201028150  201028154  201028155  201028159  201028450  201028454  201028457  201028461  201028462  201028463  201028588  100162255  100137447  100307544  100298537  201037406  201037403  100138011  100125505  100308430  100250075  201025432  100302056  100135632  100135649  100129145  201038952  100248336  100248341  100248343  100248344  100248345  100248346  100138508  201031227  201031235  201009428  201009665  201009669  201009679  100175366  201023454  100168872  100168875  201023457  100290718  100146871  100147301  201014537  201014539  100137438  100241292  100241294  201009673  100154725  100249213  100298530  100312194  100314873  201035224  100175362  100183307  100183400  100183401  201027679  201027858  100277621  100183357  100246951  100246955  100292721  100314612  201008800  201034945  201035333  100134184  100137431  100247850  100170448  100296195  100308424  100224290  100227964  100176821  100174396  201035084  100135639  100222833  100222840  100222841  100224287  100224288  201010620  100176449  100176658  100175334  100190785  100244924  100244925  100244926  100244927  100244930  100245753  100245754  100245755  100246958  201035081  201013888  100135650  100138514  100247862  100253109  100253323  100154353  100329613  100162305  201027853  201028156  201014388  100301085  100302055  100169144  100156201  201023442  201016922  201016923  100316224  100324356  201010536  201010622  100138707  100138814  100138816  100138818  100174501  100253321  100135645  100298527  100175367  201038940  100174393  100137450  100250481  100249210  100249802  100307540  100174394  201014195  100240073  100240074  100240075  100240076  100240078  201031237  201031238  201031441  201031442  201032046  100253118  100253320  100231129  100247854  100247866  100277331  100281279  201035323  201035339  201037279  201037286  201037308  201037380  201026098  100226498  100239260  201035320  201028440  100257490  100163054  100163066  100163358  100253126  100155778  100163053  100140315  201012735  201013220  201013228  100241297  201028147  100191323  100290721  100243772  100125543  100134335  100135400  100135401  100135402  100135405  100135406  100135407  100135631  100135634  100135635  100135636  100135638  100162303  100135630  100137448  100244933  100297031  100297628  100297631  100297633  100297634  201037272  201037273  201037285  201037289  201037290  201037291  201037381  201037405  201037409  201037410  100194379  201023066  100291624  100292462  100292720  100297635  100298496  100225756  100225762  100230808  100230809  100230811  100230814  100230815  100230816  100231125  100289780  100290722  100292457  100296869  100302049  100314619  100316030  100316038  100316203  100316204  100316213  100219028  201008780  201008803  100253110  100253322  100174398  100174399  100175333  100175373  100175773  100175774  100175776  100175778  100240079  100240142  100253119  100156203  100183851  100183854  100247852  100247856  100247859  100247861  100190530  100190531  100190532  100190537  100183302  100328715  201013709  100232724  100175361  100183053  100183115  100183353  100312026  201014178  100165984  100165985  100165987  100160666  100160973  100161109  100161110  100161112  100238765  100301099  100296857  100247851  100247853  100277630  201037277  100153450  100154347  100154351  100154354  100156206  100160489  100160492  100160498  100160665  100160668  100160918  100160919  100160960  100160964  100160965  100160966  100160968  100160969  100160970  100160971  100160972  100160974  100160976  100160977  100160978  100160979  100160980  100160981  100160982  100160983  100160984  100161100  100161101  100161102  100161103  100231128  100227977  100246943  100246949  100246961  100125509  100125512  100138012  100160963  100161104  100161106  100161121  100161122  100161845  100165981  100165982  201016703  100229313  100241763  100176805  100302059  100304602  100138010  100246959  100138009  100138019  100290720  100175371  100231415  100250077  100251609  100251603  100253114  100183301  201031229  201031230  100161105  100161113  201014323  201014326  201014328  201014386  201026081  201026084  201008743  201008766  201009497  100195383  201022778  201023076  201023447  201023449  100183713  100253325  100135409  100125550  100277627  201035019  201035360  100308519  100176275  100176276  100176277  100176278  100176279  100176810  100177140  100177141  100177142  100177143  100177144  100177147  100177148  100177149  100179310  100179311  100179312  100179313  100179314  100179315  100179316  100179317  100179318  100179319  100179324  100183052  100183058  100183358  100183361  100183402  100183524  100183525  100183526  100183528  100183531  100183533  100183534  100183536  100183539  100183554  100183555  100183558  100183560  100183565  100183570  100183603  100183701  100183720  100183844  100183848  201021222  201021223  201021224  201021226  201021227  201021228  201021229  201021230  201021231  201021232  201021233  201021234  201021235  201021236  201021237  201021238  201021239  201021260  201021261  201021262  201021263  201021265  201021268  201021269  201021271  201024843  201024854  201024855  201024856  201024857  201024858  201024859  201025408  201025409  201025410  201025411  201025412  201025413  201025415  201025416  201025425  201025429  201025430  201025431  201025433  201025434  201025435  201025436  201025437  201025839  100195376  201023069  100231698  100329191  100329424  100329612  100146378  100146379  100146868  100146874  100147216  100218833  100220788  100221892  100222822  201014330  201014387  201014541  201030603  201030605  201030609  201030615  201031224  201031228  100346823  100346827  100346832  100346833  100329422  100329425  100346822  100346837  100347196  100347199  100347200  100347207  100347328  100347338  100347340  100347343  100347359  100347367  100219330  100220771  100346835  100347197  201008768  201008772  100346820  100346821  100346828  100346829  100346839  100347370  201032782  100142635  100142636  100142642  100174502  100175293  100175338  100175360  100175374  100175775  100175777  100175779  100176180  100176441  100176442  100176443  100218837  100219038  100219327  100220766  100220769  100231697  100231780  100232322  100248326  100248339  100249209  100249211  100249214  100249215  100249216  100249217  100249801  100297022  100298486  100298525  100328679  100329620  100347361  201008744  201008764  201008775  201008783  201008808  201013217  201013218  201013222  201013272  201013275  201028599  201028744  201028745  201030355  201032446  201032781  100176181  201008741  100347368  100219331  100329192  100329428  100346825  100346826  100346830  100346831  100346841  100346842  100347205  100347209  100347337  100347339  100347344  100347356  100347362  100347369  100347208  100329621  100346838  100346840  100346844  100347194  100347198  100347204  100347342  100347347  100347360  201008767  100346824  100347350  201008760  100346836  100139086  100241767  201027461  201013232  100329431  100329618  100347195  100347329  100347332  100183711  100183308  100183556  100183604  100183611  100183700  100183716  100183718  100183721  100183724  100193322  100137432  100138503  100182073  100182297  100182298  100182301  100182302  100182303  100182305  100182306  100182307  100183703  100183116  100183537  100183561  100183562  100183563  100183564  100183569  100183571  100183601  100183609  100182299  100182300  100182304  100183046  100183047  100183049  100183050  100183057  100183059  100183117  100183048  100183055  100183056  100183119  100183305  100183310  100183312  100183359  100183360  100183362  100183051  100183054  100183114  100183303  100183311  100183313  100183317  100183318  100183354  100183356  100139081  100140308  100140317  100193324  201012733  201013221  201013224  201013225  201013226  201013229  100243777  100243840  100244318  100244321  100244322  100244324  100244333  100244334  100244335  100244337  100138008  100138081  100244920  100244921  100244922  100244923  100242970  100243779  100244320  100244323  100244325  100244326  100244327  100244328  100244329  100244331  100183118  100183553  100183557  100183559  100183566  100183707  100183709  100183710  100183715  100183717  100140312  100140314  100161846  201013209  201013210  201013211  201013212  201013227  201013234  201013235  100245756  100245758  100245759  100245760  100245761  100245762  100246942  100246947  100153257  100183538  100183605  100183606  100183607  100183702  100183704  100183706  100183708  100183841  100246944  100246945  201013233  100139076  100183045  100277117  100277121  100181648  100181650  100181651  100182075  100182077  100182081  100182290  100182292  100182294  100182295  100306075  201010548  100138080  100307543  100308438  100308603  100308606  100308610  100308611  100308613  100308614  100309856  100309857  100309859  100310180  100310184  100310186  100310187  100310188  100310189  100310190  100310191  100310192  100311610  100311611  100311612  100311615  100311619  100312022  100312023  100312024  100312025  100138017  100138943  100139067  100146372  100146374  100162311  100162312  100174397  100324463  100329189  100329427  100329616  201013279  201013720  201013725  201013747  201014538  100279994  100297025  201026082  201026087  100147214  100163047  100163052  100163241  100163789  100182078  201026083  100219034  100219320  100219321  201032314  100168868  100250068  100250072  100250076  100195367  100195384  100279987  100279992  100281460  201026494  100180917  100181640  100181642  100193343  201028741  201028750  100260222  100260616  100281272  100311616  100311618  100314606  100314860  100314863  100314864  100314866  100314867  100314870  100314874  201030607  100249811  100180327  100180919  100181647  100181653  100182076  100250074  100182289  100180916  100180918  100181641  100181645  100181646  100181649  100227972  100227974  100250078  100250482  100250484  100250485  201028749  201028748  201028743  100182296  100182293  100182291  100182288  100182082  100182080  100182079  100181654  100180025  100194369  201028144  100248328  100181652  201035362  100125508  100125546  100125549  201021479  100146375  100146876  100147212  100153251  100162257  201024842  201024848  201024852  201024853  100190538  100190781  100190784  100190787  100191322  100192527  100192528  100192529  100192530  100192533  100192535  100193320  100193321  100193323  100193326  100193327  100193329  100193330  100193331  100193332  100193333  100193334  100329434  201028455  100161849  100253327  100314616  201014544  201035225  201035324  201035331  201028453  100194376  100195309  100195375  100314861  100162301  100138819  100138940  100138941  100138942  100139064  100142438  100310183  100250065  100250066  100250069  100250070  100250071  100250073  201013271  201013273  201014219  100142439  100142433  100142430  100140307  201012737  100138817  100138815  100138706  100138015  100138014  100296868  100296867  100296866  100296863  100296862  100296861  100296859  100296856  100296852  100292725  100292724  100292722  100176823  100175287  201017223  201016925  201016924  201016921  100163243  100160491  100153254  100265640  100265643  100265644  100265645  100276770  100276771  100276772  100276774  100276",
      "more_code_info": "776  100276777  100276778  100276779  100276780  100276784  100276786  100276788  100276789  100277103  100277104  100277105  100277106  100277107  100277108  100277109  100277111  100277115  100277116  100277118  100277120  100296194  100306074  201034947  100139065  100139073  100140309  201012738  201013214  201013236  100297027  100168274  100190786  100296196  100301088  100301089  100301090  201014536  201014542  201021478  100147217  201014534  201014600  100168214  100168219  100168263  100168282  100194371  100194430  201023064  201023067  201023068  201023075  201023077  201023439  201023443  201023448  201020755  100229268  100227973  100265632  100265633  100290726  201014180  100260221  100260227  100260618  100260619  100260620  100260621  100260622  100260623  100260624  100260626  100260628  100260629  100260630  100260632  100260633  100260634  100260635  100261843  100261844  100261845  100261846  100261847  100261848  100261849  100261850  100261851  100261852  100261853  100261854  100261855  100261856  100261857  100261900  100261901  100261902  100316202  100316206  100316207  100316208  100316214  100316219  100316220  100316222  100316223  100316225  100316227  100324359  100324460  201038849  201038857  201038941  201038947  201038954  201038955  201038956  201038957  100193342  100277110  100277620  201025417  201025419  201025422  201025423  201026502  201026675  100276783  100277130  100277332  100257491  100279984  100279993  100279996  100279998  100281461  100301095  100302054  100311614  100312020  100312193  100312195  100316028  100316034  100316205  100316212  100316216  100316217  100154348  100153256  100291598  201013746  100175328  100175336  100176447  100176446  100176448  100180009  100222637  100229269  100264782  100265631  100265635  100265626  100265627  100265638  100265641  201028158  201028460  100180021  100180321  100180325  100142639  100160667  100160961  100176800  201035329  100297020  100190782  100314613  100253337  100255875  100257487  100257493  100257495  100257499  100258061  100258063  100258065  100258066  100258067  100258068  100307211  100307213  100307214  100307215  100307216  100307217  100307218  100307541  100307545  100308426  100308428  100308431  100308432  100308433  100308435  100308436  100308437  100308439  100308440  100308517  100308600  100308602  100308605  100347203  201038946  201027459  201027460  201027462  201026482  201026483  100194375  100194377  100195303  100195379  100176814  100195387  201022774  201023060  100195373  100195370  100195372  201022779  100194425  100195300  100183306  100183610  100183843  100190788  100194421  100125553  100163048  100163785  100163790  100163793  100163794  100163795  201016701  201016702  201016705  100135646  100135651  100135654  100135655  100137426  100137428  100137433  100137434  100137436  100137437  100192534  100194374  100194378  100194422  100194423  100194427  100194428  100194432  100195301  100195302  100195304  100195307  100195368  100195369  100195371  100195374  100195377  100195380  100195381  201022776  201025420  201026486  201026493  201026499  201026501  201026503  201026504  201026505  201026506  201026507  201026508  201026509  201027004  201027005  201027006  201027008  201027010  201027018  100129146  100129148  100134073  100134074  100134075  100134078  100134079  100134180  100134181  100134182  100134183  100155776  100160660  100176804  100176806  100176815  100176816  100176818  100176819  201010251  201010252  201010256  201010257  201010258  201010259  201010426  201010427  201010429  201010430  201010438  201010484  201010491  201010537  201010539  201010540  201010544  201010545  201010547  201010550  201010552  201010553  201010624  201010625  201010626  201011216  201011217  201011220  100147208  100147213  201010432  201010549  201010621  201011215  201037287  100310181  100310182  100316200  100316211  201038850  201038851  201038852  100329432  100329193  100174505  100174506  100183845  100192532  100305343  100305344  100305350  100305354  100305939  100306060  100306062  100306069  100306072  100306073  100298488  100298489  100298490  100298491  100311617  100314618  100314865  100315000  100315001  100315002  100315003  100315004  100315005  100316027  100316035  100316209  100316215  100316218  100298484  100298536  100162309  100279974  100281271  100281276  100302053  201014545  201028747  201034951  201026678  201021475  201021480  100239519  100239796  201023063  201032047  201032048  100165980  100239263  100253335  100243774  100243776  100244319  100239271  100240120  100183714  100168817  201037274  201037275  201037284  100277177  100277325  100277326  100277333  100316032  100316033  100329196  100329433  201037276  201037382  201035322  100231127  100238764  100239269  100239274  100239795  100239798  100239799  100240121  100288469  100288471  100288472  100288473  100288474  100288475  100288476  100288477  100288478  100288479  100288580  100288582  100288583  100288585  100288586  100288587  100288588  100289770  100289771  100289772  100289773  100289776  100289779  201034952  201034954  201035328  201035335  201035336  201035338  201035365  201037313  100291626  100183612  100291627  201010486  100253338  201011218  201010487  100180006  100162256  201038854  100324470  100328674  100328675  100328716  100328717  100328718  100328761  100328781  100328782  100328783  100329188"
    }
  ]
}