  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Best",
      "state": "",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82038",
      "recalling_firm": "Philips Medical Systems Nederlands",
      "address_1": "Veenpluis 4-6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. Nationwide distribution.",
      "recall_number": "Z-0901-2019",
      "product_description": "Ingenia 1.5T CX, Model Number 781261.   Nuclear Magnetic Resonance Imaging System",
      "product_quantity": "11226 total",
      "reason_for_recall": "There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.",
      "recall_initiation_date": "20181221",
      "center_classification_date": "20190222",
      "termination_date": "20201119",
      "report_date": "20190306",
      "code_info": "All systems",
      "more_code_info": ""
    }
  ]
}