  "results": [
    {
      "status": "Terminated",
      "city": "New York",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89735",
      "recalling_firm": "Kaneka America Corp",
      "address_1": "546 5th Ave Fl 21",
      "address_2": "N/A",
      "postal_code": "10036-5000",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US only: CA, NH, CT, MA, FL, DE, NY, NJ, PA, OH, MD, VA, NC, SC, GA, MS, TN, TX, IN, AZ, MI, WI, MN, IL, MO, NE, CO, WA, OR, UT, KS, OK, AL, ME",
      "recall_number": "Z-0900-2022",
      "product_description": "LIPOSORBER System, Model: LA-15, consisting of:     Apheresis machine, product# 101445 model: MA-03, and     disposables (product#, model)   [#101447, KP-05 Sulflux Plasma Separator],   [#101473, HDE KP-05 Sulflux Plasma Separator],   [#101785, LA-15 (AU) LDL Adsorption Column (Luer Lock Type)],   [#101786, NK-M3R(UL) Tubing (Luer Lock Type)],   [#101788, HDE LA-15(AU) LDL Adsorption Column (Luer Lock Type)],   [#101789, HDE NK-M3R(UL) Tubing (Luer Lock Type)].     Associated labeling:        1) Instructions for Use (6): IFU LIPOSORBER LA-15 LDL Adsorption Columns FH, IFU SULFLUX KP-05 PLASMA SE-PARATOR FH, IFU TUBING SYSTEM FOR PLASMAPHERESIS NK-M3R(UL) FH, IFU LIPOSORBER LA-15 LDL Adsorption Columns FSGS, IFU SULFLUX KP-05 PLASMA SE-PARATOR FSGS, IFU TUBING SYSTEM FOR PLASMAPHERESIS NK-M3R(UL) FSGS,  2) Operators Manuals for the following (2): LIPOSORBER LA-15 SYSTEM OPERATOR's MANUAL FH, LIPOSORBER LA-15 SYSTEM OPERATOR's MANUAL FSGS     3) Patients Guide (1): A PATIENT GUIDE TO THE LIPOSORBER LA-15 SYSTEM FSGS",
      "product_quantity": "138 Systems, 54703 disposables",
      "reason_for_recall": "Labeling change to due to the risk of severe anaphylactoid reactions including shock in patients treated with the Liposorber LA-15 LDL Adsorption Column under concomitant ACE inhibitor medication treatment.",
      "recall_initiation_date": "20220217",
      "center_classification_date": "20220408",
      "termination_date": "20240506",
      "report_date": "20220420",
      "code_info": "All product lot and UDI numbers are affected."
    }
  ]
}