{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Birmingham",
      "state": "AL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73088",
      "recalling_firm": "Integrated Medical Systems Inc",
      "address_1": "3200 2nd Ave N",
      "address_2": "N/A",
      "postal_code": "35222-1212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed US (nationwide) and in Australia.",
      "recall_number": "Z-0896-2016",
      "product_description": "Olympus, Model No. CF-Q180AL, flexible scope, large diameter, video colonoscope",
      "product_quantity": "82 instruments",
      "reason_for_recall": "If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.",
      "recall_initiation_date": "20160120",
      "center_classification_date": "20160225",
      "termination_date": "20161013",
      "report_date": "20160302",
      "code_info": "Serial Numbers:  2807123,2706168,2704697,2704696,2706205,2603823,2500832,2806683,2706253,  2706256,2704313,2704314,2008421,2602202,2602873,2806818,2704323,2603826,  2806676,2807264,2603013,2602S15,2501708,2806563,2501622,2602650,2501251,  2704932,2501637,2705315,2208812,2603725,2806595,2601980,2500885,2603935,  2705615,2705705,2705305,2500522,2907865,2705498,2603941,2807163,2807143,  2501819,2704886,2501553,2501744,2501768,2603178,2603044,2704755,2501372,  2807593,2705425,2208838,2208833,2706176,2008309,2704655,2705847,2706206,  2603321,2705005,2501169,2704832,2704839,2602446,2806333,2705983,2500287,  2601941,2501912,2604229,2501770,2806741,2806744,2603057,2704336,2806592,  2500217",
      "more_code_info": ""
    }
  ]
}