  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lewisville",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64265",
      "recalling_firm": "Orthofix, Inc",
      "address_1": "3451 Plano Pkwy",
      "address_2": "N/A",
      "postal_code": "75056-9453",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwidewide Distribution - USA Nationwide including Puerto Rico and the countries of:  Australia, Lebanon, Germany, Spain, Sweden and Israel.",
      "recall_number": "Z-0896-2013",
      "product_description": "Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700).      Product Usage:  The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.",
      "product_quantity": "565 total (287 PN 52-6800 and 278 PN 52- 6700)",
      "reason_for_recall": "There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening.  Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.",
      "recall_initiation_date": "20130201",
      "center_classification_date": "20130228",
      "termination_date": "20130701",
      "report_date": "20130306",
      "code_info": "Catalog Numbers:  52-6800 & 52-6700;  All Lot Numbers"
    }
  ]
}