  "results": [
    {
      "status": "Terminated",
      "city": "Birmingham",
      "state": "AL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73088",
      "recalling_firm": "Integrated Medical Systems Inc",
      "address_1": "3200 2nd Ave N",
      "address_2": "N/A",
      "postal_code": "35222-1212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed US (nationwide) and in Australia.",
      "recall_number": "Z-0891-2016",
      "product_description": "Olympus, Model No. CF-H180AL, flexible scope, large diameter, video colonoscope",
      "product_quantity": "232 instruments",
      "reason_for_recall": "If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.",
      "recall_initiation_date": "20160120",
      "center_classification_date": "20160225",
      "termination_date": "20161013",
      "report_date": "20160302",
      "code_info": "Serial Numbers:  2601800,2807942,2104002,2807968,2003519,2003522,2003506,2003533,2003517,2807849,  2807869,2106004,2001560,2602147,2806411,2900360,2001561,2602228,2602229,2309559,  2309556,2602230,2805615,2103724,2103717,2002975,2805586,2806264,2806591,2705038,  2908746,2808043,2704335,2705017,2002983,2106012,2105987,2106009,2106019,2002580,  2002571,2104510,2002603,2900105,2001677,2105162,2909901,2703303,2909972,2601007,  2601084,2104893,2704419,2806374,2806356,2104903,2908799,2601095,2807327,2640009,  2702843,2207664,2207660,2702782,2106428,2806318,2908674,2702730,2704143,2703560,  2106031,2908845,2908846,2908841,2908838,2908852,2908837,2908840,2908843,2106106,  2601525,2807589,2808233,2309810,2805782,2909109,2601535,2104776,2104781,2806163,  2908908,2103829,2104924,2002427,2909718,2002323,2703941,2601588,2601200,2600263,  2309366,2602033,2602034,2900148,2106202,2003379,2000851,2003280,2106797,2003653,  2106800,2807548,2106791,2909657,2003008,2003228,2003232,2106134,2808000,2900241,  2900656,2000774,2106423,2003189,2106410,2309513,2900683,2106413,2807888,2309493,  2106264,2106409,2106422,2107166,2909809,2107149,2808271,2702536,2702344,2704544,  2704568,2704587,2703158,2104281,2703144,2602024,2002832,2805717,2002833,2600272,  2704020,2001762,2207531,2900586,2003128,2208594,2002620,2309683,2001688,2104750,  2104891,2309573,2000881,2309635,2001645,2001650,2106010,2104981,2104993,2104937,  2104890,2900413,2703200,2601190,2309828,2909966,2601218,2001712,2601205,2806343,  2704399,2704716,2105584,2105562,2105559,2104847,2601672,2601642,2602223,2808051,  2703078,2002521,2601938,2601866,2908548,2908499,2105744,2909315,2105988,2106017,  2309497,2309262,2104662,2806423,2105002,2105542,2001423,2309783,2703663,2703654,  2001547,2600215,2001708,2001468,2001514,2001519,2309845,2001311,2309105,2208046,  2208058,2806987,2704187,2208651,2309470,2003116,2003086,2003104,2208215,2001850,  2805888,2309542",
      "more_code_info": ""
    }
  ]
}