  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73077",
      "recalling_firm": "Fresenius Medical Care Renal Therapies Group, LLC",
      "address_1": "920 Winter St",
      "address_2": "N/A",
      "postal_code": "02451-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to states of: KS, LA, MS, OK, and TX.",
      "recall_number": "Z-0887-2016",
      "product_description": "Fresenius Naturalyte Liquid Acid Concentrate  Catalog Number: 08-2301-3    Intended Use: Acid Concentrate for Bicarbonate Dialysis",
      "product_quantity": "848 (212 cases)",
      "reason_for_recall": "On the label calcium (Ca)  listed in the right upper corner of the label (red background) incorrectly indicates the product contains 2.5 mEq/L Ca. The concentration indicated in the list of constituents represents the actual calcium concentration of the final dialysate, 3.0 mEq/L.",
      "recall_initiation_date": "20160118",
      "center_classification_date": "20160225",
      "termination_date": "20170814",
      "report_date": "20160302",
      "code_info": "Lot Number: 15STAC072  Exp Date: 2017-12-31",
      "more_code_info": ""
    }
  ]
}