{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86646",
      "recalling_firm": "Respironics California, LLC",
      "address_1": "2271 Cosmos Ct",
      "address_2": "",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia.",
      "recall_number": "Z-0885-2021",
      "product_description": "Philips Respironics V60 Ventilator Part Number 1053613 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.",
      "product_quantity": "2,144 Systems",
      "reason_for_recall": "Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.",
      "recall_initiation_date": "20201030",
      "center_classification_date": "20210122",
      "termination_date": "20240919",
      "report_date": "20210203",
      "code_info": "Model: V60 Ventilator, Part Number 1053613  V60 Ventilator, Intl Opt: CFLEX,AVAPS,PP    Serial Numbers:  100029460  100086827  100106522  100125955  100173096  100173294  100173478  100173526  100173591  100173593  100173595  100174299  100174305  100174308  100190730  100252781  100252782  100252785  100304095  100304097  100304263  100304264  100304446  100304447  100304448  100304451  100304462  201022963  100190253  201024196  201025984  201025987  201025988  100176847  100176856  100252776  100315180  100315377  100315598  100315602  100315607  100315610  100315889  100350701  100361426  100363749  100370895  100375246  100044066  100074200  100074206  100074207  100085241  100085243  100085327  100102145  100102150  100102152  100102155  100114031  100118553  100118676  100118684  100119491  100120774  100138585  100138589  100146847  100146849  100147395  201008175  201008180  201008192  201008218  201008225  201008226  201013645  201013649  201013650  201014018  201014041  201014042  201014047  201031153  201031164  201031195  201031200  201031214  201031215  201031218  100039855  100039944  100040070  100045713  100045728  100046139  100064525  100064561  100064565  100064566  100064717  100071592  100071593  100081759  100082460  100083215  100086784  100086887  100087022  100091384  100091385  100101055  100106443  100106507  100106523  100114030  100125962  100136656  100137203  100180416  100181591  100181614  100189483  100189758  100229116  100230021  100252575  100252788  201004061  201004067  201004363  201004374  201004375  201004687  100087603  100129448  100129779  100129782  100129783  100129784  100129785  100173086  100173089  100173090  100173095  100173097  100173219  100173241  100173246  100173252  100173597  100243515  100243631  100243632  100243640  100243912  201007865  201008232  201011112  100118546  100118694  100260182  100260183  201013158  100189466  201029733  100036150  100036155  100067450  100107006  100160829  100190082  100219300  201026534  100026664  100026678  100029057  100029058  100029066  100029068  100029075  100029077  100029078  100029222  100029240  100037555  100045715  100045723  100059156  100084841  100084845  100084847  100084849  100084850  100085135  100085240  100086454  100114020  100137503  100146859  100181615  100181741  100190804  100229633  201002728  201004021  201004051  201004055  201004056  201004057  201004058  201004060  201004063  201004064  201008173  201029651  201029729  201029762  201029807  100116680  100116681  100116697  100116830  100116831  100116832  100118079  100118080  100118264  100124863  100124968  100147573  201008060  100026672  100026686  100026755  100026760  100029335  100031914  100031916  100037540  100037545  100037546  100037550  100037558  100038202  100038212  100038214  100045670  100045676  100045678  100045681  100045682  100045711  100045714  100045717  100045719  100045721  100045722  100045724  100045727  100045731  100046143  100073255  100085247  100091381  100091383  100104055  100104065  100106520  100114024  100118693  100137022  100137481  100147282  100147563  100174990  100181612  100230020  100237374  100237623  100237625  100237626  100237628  100237629  100237630  100237632  100237635  100237638  100261124  201001547  201001568  201001570  201001576  201002273  201002274  201002276  201002394  201002401  201002576  201002579  201002729  201002740  201002741  201002744  201002746  201002747  201002748  201002749  201002750  201002751  201002754  100038208  100045726  100118695  100125750  201001573  201004062  201004371  100037539  100046136  100046138  100046140  100046142  100046152  100086773  100086886  100091389  100098477  100100310  100103663  100106511  100106517  100118557  100118697  100163749  100164087  100176028  100181587  100181611  100181619  100189468  100189471  100189472  100189480  100189486  100229635  100229636  100229639  100230023  100244201  100251169  100254446  100254791  201002725  201029886  201029948  100029459  100036094  100036143  100036153  100036154  100036156  100036161  100036290  100039849  100040072  100040073  100045725  100045732  100046145  100046149  100046900  100050086  100050097  100064567  100064571  100064572  100064574  100064575  100064576  100064669  100064670  100064962  100071513  100071514  100073256  100075378  100075407  100081757  100082431  100082435  100082436  100082462  100087174  100091098  100091099  100091380  100091386  100091387  100091388  100091392  100091396  100100859  100101061  100101068  100101081  100103676  100104050  100104053  100104057  100104068  100104071  100104072  100106524  100106734  100114023  100114026  100114029  100118349  100118687  100118690  100118691  100118692  100125175  100125178  100125721  100125753  100125757  100125758  100125760  100125762  100125852  100125953  100125954  100125956  100125957  100125958  100136805  100136886  100137131  100146858  100147384  100147393  100147566  100147577  100163934  100174175  100174964  100174970  100174974  100174986  100176029  100176846  100180942  100181592  100181600  100189463  100189467  100189476  100189491  100189748  100190008  100190080  100190190  100190739  100190801  100190802  100195440  100195643  100195647  100195651  100195657  100195676  100195867  100195876  100195897  100196405  100196432  100219288  100219294  100219296  100219297  100219298  100219299  100219301  100219305  100219306  100228999  100229101  100230024  100230074  100231171  100231173  100231175  100231176  100231177  100231178  100231202  100231203  100243644  100252774  100252783  100252792  100253495  100292297  100315391  100315604  100350686  100350697  100350757  100354059  100354431  100354434  100354443  100354449  100354457  100361203  100361208  100361587  100362571  100362577  100362711  100362938  100363006  100363023  100363081  100363753  100363784  100367301  100370788  100371117  100371646  100373208  100373510  100373515  100373517  100373542  100373544  100373549  100373558  100373559  100373603  100373604  100373608  100373609  100373611  100375002  100375005  100375006  100375011  100375012  100375015  100375016  100375018  100375144  100375146  100375218  100375221  100375234  100375235  100375237  100375242  100375247  100375263  100377893  100377913  100382184  100382195  100382197  201002575  201002682  201002689  201002714  201002727  201004065  201004354  201004376  201004380  201004681  201004684  201022721  201023046  201023103  201025264  201025274  201026475  201026552  201026707  201027791  201029718  201029730  201029731  201029732  201029735  201029738  201029739  201029743  201029747  201029790  201029935  100136654  100137135  100147398  100147562  100163751  100181584  100181590  100026671  100253473  100315036  100315390  100361577  100362560  100363765  100363774  201040666  100071594  100253478  201002753  201004398  100029074  100029457  100029467  100029468  100029473  100037544  100037556  100231168  100091391  100147293  100147388  100146854  100147197  100147286  100147381  100147570  100160825  100160826  100160833  100160836  100181581  100181601  100181606  100252970  100253489  100036151  100071507  100071512  100071515  100071519  100071595  100073206  100073249  100073261  100073262  100084757  100085246  100038199  100038204  100046399  100047067  100055806  100055807  100089641  100089644  100089646  100089649  100089650  100089652  100089653  100089655  100089656  100089657  100094138  100094216  100094394  100094425  100097039  100100832  100100836  100100839  100100840  100100843  100100844  100100845  100100847  100100848  100100849  100107145  100107147  100107148  100107149  100107293  100107294  100107297  100107300  100107302  100107304  100107305  100137839  100138042  100138050  100138053  100138056  100138058  100138286  100138289  100138417  100138581  100138590  100138592  100138593  100138594  100145966  100346938  100346940  100346946  100346948  100346949  100346950  100346953  100346957  100346959  100346962  100346963  100346976  100348093  100348094  100348095  100348096  100348097  100348099  100348101  100348103  100348104  100348105  100348106  100348107  100348108  100348115  100348117  100348134  100348160  100348161  100348162  100348164  100348165  100348166  100348167  100348168  100348169  100348170  100348171  100348172  100359658  100359876  100359951  100360190  100360191  100360211  100360212  100360282  100360342  100360350  100360351  100360354  100360355  100360357  100360359  100360361  100360363  100360380  100360381  100361198  100362046  100362443  100362518  100363078  100363651  100363655  100363742  100363757  100363791  100364520  100364603  100365116  100365175  100365178  100365188  100365192  100365649  100365651  100365652  100366646  100366657  100366658  100366701  100366702  100366705  100366707  100366711  100366712  100366714  100366914  100366917  100366941  100367018  100367020  100367030  100367032  100367270  100367271  100367273  100367274  100369412  100369423  100369429  100369438  100369439  100369442  100369443  100369448  100370288  100370289  100370756  100370804  100370806  100370815  100370837  100370884  100370999  100371100  100371202  201007526  201008056  201008189  201008221  201008224  201008229  201008231  100028670  100029465  100189489  100189659  100026094  100291733  100037543  100037547  100046154  201001571  201002277  100064960  100064988  100071522  100073263  100103585  100103677  100104054  100104059  100104070  201004358  201004369  100120756  100120768  100120856  100120860  100120863  100243639  100237624  100237633  100237634  100237637  100237639  100237640  201034072  201002752  100118345  100125748  100026121  100040071  100118348  100118544  100118558  100118680  100125173  100125177  100125179  100125759  100189759  100190736  100190803  100219302  100219303  100229115  100229631  100229632  100229634  100229637  100229638  100252579  100252778  100252790  100291846  100291849  100291850  100291853  100292298  201004351  201004362  201004366  201004372  201004381  100036096  100036097  100036147  100036157  100036159  100039844  100040074  100040075  100040076  100040078  100040079  100040080  100040081  100040082  100040083  100040084  100040085  100040086  100040087  100040089  100045716  100045718  100045720  100045729  100045733  100046141  100046150  100046151  100064563  100064568  100064570  100064961  100081492  100081691  100083222  100083227  100083228  100083232  100083233  100083234  100083235  100083236  100087013  100091395  100187424  100350146  100350152  100350156  100350157  100350158  100350163  100350667  100350674  100350683  100350688  100350691  100350692  100350699  100350733  100350742  100350743  100350751  100350758  100350932  100351347  100351348  100351351  100351352  100351354  100351357  100351358  100351547  100351562  100351565  100351572  100351575  100351576  100351582  100351584  100351681  100351683  100352744  100353246  100353256  100353308  100353310  100353313  100353317  100353318  100353319  100353586  100353587  100353590  100353591  100353593  100353594  100353595  100353596  100353919  100354440  100354442  100354461  100354468  100354710  100354854  100354887  100354891  100355153  100355165  100355168  100355391  100355398  100355668  100355678  100355800  100355844  100355847  100355848  100355859  100355862  100355866  100355869  100355870  100355871  100355875  100355877  100355881  100355889  100355892  100355910  100355918  100355922  100355924  100355925  100355928  100356019  100356083  100356085  100356086  100356088  100356209  100356246  100356253  100356254  100356275  100356276  100356284  100356289  100356294  100356295  100356299  100356300  100356304  100356765  100356770  100356774  100357766  100359603  100359654  100359768  100359769  100359871  100359877  100359879  100359947  100359948  100359955  100360083  100360086  100360144  100360148  100360204  100360258  100360259  100360280  100360348  100360366  100360382  100360384  100361235  100361399  100361405  100361435  100361441  100361444  100361489  100361495  100361529  100361532  100361541  100361544  100361554  100361572  100361589  100361767  100361770  100361778  100361960  100362000  100362002  100362016  100362029  100362041  100362062  100362383  100362385  100362386  100362394  100362446  100362448  100362519  100362541  100362550  100362552  100362562  100362572  100362715  100362720  100362724  100362730  100362734  100362744  100363009  100363010  100363019  100363031  100363058  100363652  100363654  100363658  100363752  100363762  100363770  100363782  100363789  100363790  100363796  100363803  100363805  100363806  100364253  100364254  100364255  100364257  100364259  100364500  100364501  100364502  100364504  100364505  100364506  100364507  100364508  100364509  100364510  100364513  100364535  100364536  100364612  100364614  100364615  100365104  100365106  100365114  100365118  100365646  100365647  100365736  100366649  100366652  100366654  100366659  100366704  100366920  100366922  100366936  100366940  100366943  100366945  100366948  100367019  100367022  100367024  100367027  100367031  100367142  100368045  100368047  100368059  100368074  100368552  100368720  100368726  100370786  100370789  100370791  100370792  100370795  100370802  100370805  100370808  100370809  100370811  100370812  100370813  100370822  100370828  100370830  100370834  100370840  100370841  100370842  100370892  100370893  100370896  100370960  100370985  100370986  100370990  100370992  100370994  100370998  100371002  100371105  100371123  100371127  100371141  100371143  100371172  100371211  100371212  100371409  100371411  100371415  100371419  100371613  100371640  100371642  100371648  100371652  100371657  100371658  100371659  100371680  100371687  100371689  100371692  100371695  100371699  100371701  100371704  100372281  100372283  100372286  100372287  100372289  100372291  100372292  100372293  100372896  100372898  100373204  100373209  100373211  100373224  100373228  100373232  100373234  100373235  100373433  100373449  100373459  100373462  100373463  100373464  100373465  100373507  100373512  100373513  100373516  100373518  100373606  100373610  100373680  100373682  100373683  100373684  100373941  100373942  100373943  100373945  100373946  100375003  100375017  100375150  100375151  100375229  100375231  100375245  100375264  100375277  100375408  100377558  100377885  100377892  100377894  100377896  100377906  100377909  100377912  100377914  100377915  100377916  100377917  100377918  100377922  100377923  100377924  100377925  100377926  100377928  100377930  100377932  100377936  100377939  100378009  100378017  100378061  100378073  100378074  100378076  100378079  100378082  100378201  100378205  100378228  100378229  100378232  100378234  100378235  100378285  100378286  100378287  100378289  100378299  100378846  100378861  100378929  100380818  100381372  100381375  100381376  100381734  100381828  100381833  100381836  100381837  100381838  100381846  100381849  100381851  100381855  100381858  100381859  100381860  100381863  100381865  100381869  100381900  100381902  100381907  100381913  100381915  100381920  100381927  100381929  100382180  100382183  100382185  100382188  100382192  100382196  100383931  100385164  100385316  100385318  100385319  100385320  100385321  100385324  100385325  100385326  100385329  100385339  100385347  100385351  100385353  100385360  100385361  100385363  100387273  100388723  100388735  100389237  100389382  100389386  100389392  100389398  100389408  100389409  100389432  100389442  100389444  100389445  100389446  100389447  100389458  100389459  100389465  100389466  100389473  100389481  100389485  100389486  100389489  100389491  100389496  100389499  100389500  100389501  100389502  100389506  100389507  100389510  100389511  100389513  100389514  100389517  100389518  100389540  100389541  100389542  100389543  100389545  100389546  100389547  100389548  100389992  100390092  100390129  100390161  100390182  100390414  100390484  100390492  100390505  100391041  100391043  100391050  100391051  100391052  100391053  100391056  100391057  100391059  100391071  100391072  100391073  100391074  100391075  100391076  100391077  100391078  100391079  100391081  100391083  100391085  100391086  100391089  100391091  100391092  100391097  100391098  100391099  100391100  100391102  100391103  100391104  100391105  100391109  100391110  100391111  100391116  100391119  100391200  100391201  100391202  100391204  100391205  100391206  100391207  100391208  100391209  100391210  100391211  100391213  100391215  100391216  100391217  100391219  100391363  100391364  100391365  100391366  100391367  100391368  100391370  100391371  100391372  100391375  100391376  100391464  100391472  100391474  100391478  100391480  100391483  100391489  100391490  100391492  100391494  100391496  100391497  100391498  100391499  100391501  100391503  100391505  100391507  100391510  100391511  100391513  100391516  100391517  100391518  100391519  100391521  100391522  100391523  100391525  100391526  100391527  100391528  100391531  100391532  100391533  100391534  100391536  100391538  100391539  100391540  100391541  100391543  100391544  100391547  100391548  100391549  100391552  100391839  100391861  100391862  100391867  100391868  100391869  100391870  100391871  100391874  100391877  100391883  100391884  100391889  100391894  100391900  100391905  100391908  100391909  100391910  100391911  100391912  100391913  100391914  100391916  100391917  100391960  100391962  100391975  100391976  100391980  100391981  100392130  100392134  100392135  100392136  100392137  100392138  100392240  100392241  100392242  100392243  100392244  100392245  100392248  100392250  100392252  100392253  100392254  100392257  100392258  100392268  100392276  100392287  100392292  100392293  100392295  100392296  100392299  100392613  100392624  100392634  100392635  100392636  100392637  100392638  100392640  100392691  100392693  100392694  100392696  100392698  100392699  100393577  100393579  100393641  100393642  100393643  100393649  100393650  100393651  100393653  100393656  100393657  100393658  100393700  100393701  100393702  100393703  100393704  100393705  100393707  100393708  100393709  100393710  100393711  100393715  100393861  100393863  100393868  100393869  100393870  100393873  100393874  100393875  100393877  100393878  100393879  100393880  100393881  100393882  100393883  100393884  100393886  100393888  100393889  100393891  100393894  100393895  100393897  100393898  100393899  100394361  100394363  100394364  100394365  100394366  100394367  100394368  100394370  100394371  100394372  100394373  100394374  100394376  100394377  100394379  100394381  100394384  100394385  100394386  100394387  100394388  100394392  100394394  100394395  100394397  100394399  100394454  100394455  100394456  100394458  100394459  100394461  100394463  100394464  100394465  100394466  100394467  100394469  100394472  100394473  100394474  100394475  100394478  100394482  100394483  100394485  100394489  100394490  100394491  100394492  100394494  100394497  100394505  100394506  100394507  100394510  100394512  100394514  100394515  100394517  100394523  100394525  100394531  100394547  100394548  100394550  100394552  100394668  100394669  100394670  100394672  100394673  100394674  100394675  100394676  100394679  100394948  100394949  100394950  100394951  100394954  100394964  100394965  100394968  100394978  100394980  100394981  100394983  100394984  100394985  100395000  100395001  100395002  100395003  100395005  100395006  100395068  100395069  100395074  100395075  100395320  100395321  100395323  100395329  100395330  100395331  100395342  100395343  100395344  100395345  100395346  100395347  100395351  100395353  100395355  100395358  100395416  100395418  100395560  100395567  100395570  100395571  100395572  201002683  100037537  100037538  100037554  100060417  100060477  100060478  100060479  100060489  100060580  100060581  100060582  100060583  100060584  100060586  100060587  100060588  100060589  100060590  100060592  100060595  100060596  100060598  100060601  100060602  100060603  100061650  100061651  100061775  100061780  100061781  100061782  100061783  100061784  100061785  100061786  100061787  100061788  100061789  100061790  100061791  100061793  100061795  100061799  100029469  100125959  100125965  100180970  100181586  100181605  100189473  100026086  100266744  100045730  100049073  100049074  100049192  100057214  100057353  100064559  100064838  100065594  100065597  100067440  100067531  100067535  100073203  100084823  100084831  100091401  100174176  100174178  100174961  201015523  100083223  100085474  100098474  100100132  100100173  100114021  100114022  100114040  100137026  201002217  201002386  201004059  100394470  100394508  100051959  100051964  100051967  100051968  100094219  100094345  100095647  100095648  100095649  100097381  100181618  201002380  100072480  100072481  100072482  100072484  100073253  100115401  100115402  100115404  100115405  100115406  100115409  100115412  100115461  100115464  100115798  100115802  100115803  100115806  100115809  100115810  100115811  100118343  100118550  100119490  100120751  100120759  100120854  100120855  100120857  100120861  100120865  100120868  100122342  100122343  100122346  100122582  100122585  100132201  100132205  100132208  100136647  100136649  100136653  100136809  100147107  100147205  100147283  100147285  100147383  100147386  100147396  100147397  100147564  100154610  100154611  100159226  100159228  100159236  100159238  100159560  100159561  100159563  100159566  100159568  100159920  100164088  100174307  100189485  100190806  100190818  100190821  100228824  100230070  100230076  100230167  100231170  100256987  100257092  100257809  100257811  100257818  100278754  100278785  100278794  100278799  100278809  100350165  100350169  100350171  100350174  100350179  100350180  100350187  100350700  100350745  100354033  100354426  100354430  100354433  100354436  100354441  100354446  100354447  100354450  100354451  100354455  100354456  100354460  100354474  100355919  100359648  100359872  100360358  100363028  100363029  100363055  100363783  100363795  100364619  100366700  100367141  100367300  100367306  100367308  100368377  100368522  100370807  100370814  100370889  100371176  100371433  100371708  100373552  100373553  100373554  100373605  100373944  100375013  100375140  100375143  100375228  100375230  100375239  100375244  100375276  100375287  100375288  100375297  100375380  100377880  100377887  100378217  100378222  100378226  100378230  100378231  100378233  100378282  100378284  100378288  100378292  100378304  100381374  100381834  100381910  100382187  100387153  100387243  100387249  100391915  100392695  201008038  201008185  201008230  201009337  201009339  201010599  201014978  201015059  201015061  201023059  201028002  201029728  201030728  201030884  201030940  201030942  201030947  201030952  201030961  201031063  201031068  201031074  201031161  201031193  201031196  201031201  201031204  201031205  201031206  201031212  201031213  201031217  201031444  201031446  201031447  201031448  201031449  201031450  201031452  201031454  201031455  201031458  201031520  201031527  201031528  201031531  201031532  201031535  201031538",
      "more_code_info": ""
    }
  ]
}