  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Savannah",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84545",
      "recalling_firm": "Brasseler USA I Lp",
      "address_1": "1 Brasseler Blvd",
      "address_2": "N/A",
      "postal_code": "31419-9576",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US: AZ, AR, FL",
      "recall_number": "Z-0884-2020",
      "product_description": "Brasseler USA KM169-39-76 Steinmann Pin Single Diamond Smooth, Diameter 2.8 mm, Length 228 mm    Product Usage:  A Steinmann Pin is used for internal fixation of large bone fractures.",
      "product_quantity": "56 packs of 6",
      "reason_for_recall": "103 packaged and labeled tubes (103 product tubes with 6 pins per tube) of KM168-39-76 Single Diamond Smooth Steinmann Pins were labeled with the incorrect product labels.  The tubes were incorrectly labeled as KM169-39-76 Single Diamond Threaded Steinmann Pins.",
      "recall_initiation_date": "20191211",
      "center_classification_date": "20200129",
      "termination_date": "20210212",
      "report_date": "20200205",
      "code_info": "Lot Number NR8ZR"
    }
  ]
}