  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63674",
      "recalling_firm": "Philips Healthcare Inc.",
      "address_1": "3000 Minuteman Road",
      "address_2": "N/A",
      "postal_code": "01810",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA Nationwide and the Foreign countries of: Austria, Canada, Belgium, China,Denmark, France, germany, Italy, Netherlands,New Zealand, Norway, Philippines, Portugual, SOuth Korea, Spain, Switzerland, Thailand, and UK.",
      "recall_number": "Z-0884-2013",
      "product_description": "Philips DigitalDiagnost X-Ray System with Eleva Software 2.1.3  Model numbers 712020, 712022, and 712082      Product Usage:      Digital radiographic system",
      "product_quantity": "551",
      "reason_for_recall": "When the operator for a wall stand view selects an \"image rotation\" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction.  When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.",
      "recall_initiation_date": "20121025",
      "center_classification_date": "20130226",
      "termination_date": "20140331",
      "report_date": "20130306",
      "code_info": "Systems with Eleva SW version 2.1.3"
    }
  ]
}