  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76874",
      "recalling_firm": "Bausch & Lomb Inc Irb",
      "address_1": "1400 Goodman St N",
      "address_2": "N/A",
      "postal_code": "14609-3547",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0883-2018",
      "product_description": "LASEREDGE Knives, KNIFE ANGLED BEVELED UP 2.7(6/BOX), REF/PRODUCT CODE E7559A, STERILE, Rx Only     Product Usage:  The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures.",
      "product_quantity": "1,236,749 knives in total",
      "reason_for_recall": "The complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing.",
      "recall_initiation_date": "20170303",
      "center_classification_date": "20180305",
      "termination_date": "20180625",
      "report_date": "20180314",
      "code_info": "none"
    }
  ]
}