  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marietta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75858",
      "recalling_firm": "Amendia, Inc",
      "address_1": "1755 W Oak Pkwy",
      "address_2": "N/A",
      "postal_code": "30062-2260",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution to AL, AZ, AR, CA, CO, FL, GA, IA, KS, KY, LA, MA, MI, MN, MO, NV, NJ, NC, SC, TN, TX, VA, WA, WV, WI and Puerto Rico",
      "recall_number": "Z-0883-2017",
      "product_description": "Savannah-T Straight Rod, 200mm, Part Number 10704-200; Savannah-T Straight Rod, 250mm, Part Number 10704-250; Savannah T-Straight Rod 300mm, Part Number 10704-300; Savannah-T Straight Rod, 400mm, Part Number 10704-400; Savannah-T Straight Rod, 500mm; Savannah-T High Top Straight Rod, 200mm; Savannah-T High Top Straight Rod, 220mm, Part Number 1705-220; Savannah-T High Top Straight Rod, 240mm, Product Code 10705-240; Savannah-T Straight Rod, 260mm, Part Number 10705-260; Savannah-T Straight Rod, 280mm, Part Number 10705-280; Savannah-T Straight Rod, 300mm, Part Number 10705-300; Savannah-T Notched Curved Rod, 200mm, Part Number 10706-200; Savannah-T Curve Rod, 200mm, Part Number 10707-200     The guide wire is passed through the cannulated handle of the Dual Stylet Needles and pressed into the bone",
      "product_quantity": "240 units",
      "reason_for_recall": "The rods are packaged in packages with unacceptable seals.  In some cases, the integrity of the seal is compromised by channels or voids.",
      "recall_initiation_date": "20161122",
      "center_classification_date": "20161222",
      "termination_date": "20240201",
      "report_date": "20161228",
      "code_info": "Part Number 10704-200  Lot Number 124422, 129766, 130739, 132430; Part Number 10704-250  Lot Number 129790, 130738, 131211, 131309, 131310, 131509, 132811; Part Number 10704-300  Lot Number 129862, 134229; Part Number 10704-400  Lot Number 129861, 130145, 130147, 130502, 131499, 131646; Part Number 10704-500  Lot Number 127934, 127937, 129770, 129863, 130130, 130143, 130781, 134228; Part Number 10705-200  Lot Number 122654, 123583, 129773, 130131, 132153, 132154, 132762; Part Number 10705-220  Lot Number 128818, 131311, 132881; Part Number 10705-240  Lot Number 130500; Part Number 10705-260  Lot Number 121521, 131500, 132199; Part Number 10705-280  Lot Number 121519, 130740; Part Number  10705-300  Lot Number 130140, 130332, 134299; Part Number 10706-200  Lot Number  128522, 129567, 129781, 129891, 130741; Part Number 10707-200  Lot Number 128815, 129782, 129783, 129784, 130182, 131410, 132880."
    }
  ]
}