  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63822",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AK, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, OA MA, MD MI, MN, MO, MS, NC, D, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; Internationally to Canada and other countries..",
      "recall_number": "Z-0882-2013",
      "product_description": "Philips Medical Systems, GEMINI LXL - Model #882412; GEMINI TF 16 - Model #882470 and Model #882473; GEMINI TF 64 - Model #882471, Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanning systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.",
      "product_quantity": "30 units",
      "reason_for_recall": "Software anomaly.  Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project",
      "recall_initiation_date": "20121024",
      "center_classification_date": "20130226",
      "termination_date": "20150723",
      "report_date": "20130306",
      "code_info": "The affected GEMINI systems are identified with the following Serial Numbers:   GEMINI LXL Model # 882412, Serial Numbers: 14702 and 14707;  GEMINI TF 16  Model #882470, Serial Numbers: 7129, 7121, 7123, 7128, 7096, 7133, 7185, 7187, 7226, and 7220;   GEMINI TF 16, Model #882473, Serial Numbers:  7150, and and 7237;GEMINI TF 64, Model #882471, Serial Numbers: 7103, 7126, 7132, 7158, 7160, 7164, 7175, 7178, 7179, 7218, 7229, 7221, 7245, 7550, 7255, and 7256."
    }
  ]
}