{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Princeton",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87073",
      "recalling_firm": "Abbott Point Of Care Inc.",
      "address_1": "400 College Rd E",
      "address_2": "N/A",
      "postal_code": "08540-6607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-0881-2021",
      "product_description": "Abbott i-STAT EG7+ cartridge-IVD.  Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood.  List Number: 03P76-25",
      "product_quantity": "1,521,800 cartridges US and 2,641,053 cartridges OUS",
      "reason_for_recall": "Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above  Where * = lot letter A, K, L, M, N, W or Y.",
      "recall_initiation_date": "20201222",
      "center_classification_date": "20210119",
      "termination_date": "20230901",
      "report_date": "20210127",
      "code_info": "Stored at room temperature (18 to 30¿C) for periods of time in excess of:  3 days for lot numbers *20100 to *20339  7 days for lot numbers *20340 and above  Where * = lot letter A, K, L, M, N, W or Y    Primary GTIN (01) 00054749000163)",
      "more_code_info": ""
    }
  ]
}