  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76874",
      "recalling_firm": "Bausch & Lomb Inc Irb",
      "address_1": "1400 Goodman St N",
      "address_2": "N/A",
      "postal_code": "14609-3547",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0881-2018",
      "product_description": "LASEREDGE Knives, 3.2MM STRAIGHT SLIT KNIFE (6/BOX), REF/PRODUCT CODE E7557, STERILE, Rx Only     Product Usage:  The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures.",
      "product_quantity": "1,236,749 knives in total",
      "reason_for_recall": "The complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing.",
      "recall_initiation_date": "20170303",
      "center_classification_date": "20180305",
      "termination_date": "20180625",
      "report_date": "20180314",
      "code_info": "none"
    }
  ]
}