      "status": "Terminated",
      "city": "Camarillo",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64294",
      "recalling_firm": "Medacta Usa Inc",
      "address_1": "4725 Calle Quetzal Ste B",
      "address_2": "N/A",
      "postal_code": "93012-8429",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including the states of ID, CO, TX, WI, NY, CT, WA, CA, UT, and AZ.",
      "recall_number": "Z-0878-2013",
      "product_description": "Versafitcup Impacting Ring    Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.10.0139 (58mm), 01.26.10.0140 (60mm), 01.26.10.0141 (62mm), 01.26.10.0142 (64mm).    Designed for cementless use in total hip arthroplasty.",
      "product_quantity": "293 units",
      "reason_for_recall": "Medacta USA is recalling Versafitcup Impacting Ring because the impaction ring may not fit properly in the implant shell in all cases.  This could result in surgical delay potentially increasing possible complications under anesthesia, having to ream up to a larger shell size leaving less bone than planned and repositiong/removal of the shell after initial impaction.",
      "recall_initiation_date": "20100115",
      "center_classification_date": "20130222",
      "termination_date": "20130223",
      "report_date": "20130306",
      "code_info": "Reference Number Lot Number Reference Number Lot Number  01.26.10.0133: 095920, 096050, 096050A;  01.26.10.0138: 095925, 096055, 096055A;  01.26.10.0134: 095921, 096051, 096051A;  01.26.10.0139: 095926, 096056, 096056A;  01.26.10.0135: 095922, 096052, 096052A;  01.26.10.0140: 095927, 096057, 096057A;  01.26.10.0136: 095923, 096053, 096053A;  01.26.10.0141: 095928, 096058, 096058A;  01.26.10.0137: 095924, 096054, 096054A;  01.26.10.0142: 095929, 096059, 096059A."
    }
  ]
}