  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84655",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.",
      "recall_number": "Z-0877-2020",
      "product_description": "SIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.",
      "product_quantity": "62 (30 US, 32 OUS) total devices",
      "reason_for_recall": "It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.",
      "recall_initiation_date": "20191223",
      "center_classification_date": "20200128",
      "termination_date": "20201211",
      "report_date": "20200205",
      "code_info": "Mfg. Lot or Serial #   System ID  22D-4-0868           901BMHCRITMR  HDGOC1900010SC  209521SGMR  HDGOC1900015SC  4577711MR11  002776WHF            954916MR1  022D-4-1175            808243MR  857                            LONGWOOD30MR2  00000290815MR0    508580BMR3  002801WHF            734246MR  002930WHF            845279MR  00000269040MR2    412749MR1  022D-4-1125            801RIVMR2  HDGOC1900003SC   206386SMR  H05114                    4815915MR11  00000277443MR8    304526SMR1  To be provided    GON4796599  To be provided    424881MR01  To be provided    083027829345912  To be provided    083027862100408"
    }
  ]
}