  "results": [
    {
      "status": "Terminated",
      "city": "Shenyang",
      "state": "N/A",
      "country": "China",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75383",
      "recalling_firm": "Neusoft Medical Systems Co., Ltd.",
      "address_1": "NO. 16 Shiji Road",
      "address_2": "Hunnan Industrial Area",
      "postal_code": "N/A",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution",
      "recall_number": "Z-0876-2017",
      "product_description": "NeuViz 128 Multi-slice CT Scanner System with  software version 1.0.7.4038+P06 or previous version",
      "product_quantity": "23 units",
      "reason_for_recall": "The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the interruption error before scanning can resume.",
      "recall_initiation_date": "20160831",
      "center_classification_date": "20161223",
      "termination_date": "20180104",
      "report_date": "20170104",
      "code_info": "NeuViz 128 Multi-slice CT Scanner System with  software version 1.0.7.4038+P06 or previous version"
    }
  ]
}