      "status": "Terminated",
      "city": "Ivry Le Temple",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69879",
      "recalling_firm": "PEROUSE MEDICAL",
      "address_1": "ROUTE DU MANOIR",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of AZ and CO.",
      "recall_number": "Z-0876-2015",
      "product_description": "Bard Caliber (CL3030)  Inflation Device    Product Usage:  Inflation Device intended for use during cardiovascular to create and maintain, monitor pressure in the balloon catheter -  balloon inflation device for angioplasty procedures",
      "product_quantity": "24821 units",
      "reason_for_recall": "Additional instructions for use due to inability to raise the pressure above 10 atm during inflation phase may result in an incomplete inflation or delay in procedure",
      "recall_initiation_date": "20141006",
      "center_classification_date": "20141229",
      "termination_date": "20170324",
      "report_date": "20150107",
      "code_info": "All lot numbers beginning with Numbers: 1403,1404,1405,1406,1407"
    }
  ]
}