      "country": "United States",
      "city": "Mercedita",
      "address_1": "Turpeaux Industrial Park Rd 1 km 123.4",
      "reason_for_recall": "Zimmer received a single complaint wherein a Trilogy&reg; Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter  was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which",
      "address_2": "",
      "product_quantity": "68 units",
      "code_info": "Item No. 00-6200-054-20; Lot 62714305",
      "center_classification_date": "20141224",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: France, Germany, Iran, Romania, Serbia, Singapore, Taiwan, and United Kingdom.",
      "state": "PR",
      "product_description": "Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD    The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.",
      "report_date": "20141231",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Manufacturing B.V.",
      "recall_number": "Z-0875-2015",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "69814",
      "termination_date": "20150618",
      "more_code_info": "",
      "recall_initiation_date": "20141204",
      "postal_code": "00715",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}