  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Parsippany",
      "address_1": "399 Jefferson Rd",
      "reason_for_recall": "A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (\u001cthe  Sleeves\u001d) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.",
      "address_2": "",
      "product_quantity": "193 units",
      "code_info": "Part number 14-531593  Part number 14-531594    Catalog number         Lot Number  14-531593                   N23597   14-531594                   N23599",
      "center_classification_date": "20141224",
      "distribution_pattern": "US Nationwide Distribution.",
      "state": "NJ",
      "product_description": "Solitaire-C Cervical Spacer System    a) 14-531593 Drill/Awl Sleeve  b) 14-531594 Spring-Loaded Drill/Awl Sleeve    Intended to stabilize cervical spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft with or without supplemental fixation.",
      "report_date": "20141231",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Ebi, Llc",
      "recall_number": "Z-0874-2015",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "69981",
      "termination_date": "20150415",
      "more_code_info": "",
      "recall_initiation_date": "20141111",
      "postal_code": "07054-3707",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}