  "results": [
    {
      "status": "Terminated",
      "city": "Gainesville",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84532",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 NW 66th Ct",
      "address_2": "N/A",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "International distribution in the countries of Australia; Austria; Brazil; Canada; China; Colombia: France; Germany; Guatemala, India; Italy; Japan; Luxembourg; Netherlands; Singapore; South Korea; Spain; Sweden: United Kingdom.",
      "recall_number": "Z-0873-2020",
      "product_description": "Exactech equinoxe PLATFORM FRACTURE, HUMERAL STEM, CEMENTED, Right, 10.5mm, REF 304-22-11 - Product Usage: The Equinoxe Shoulder System is indicated for use and in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint work total or hemi- arthroplasty is determined by the surgeon to be the preferred method of treatment.",
      "product_quantity": "N/A",
      "reason_for_recall": "Labels with incorrect translations were identified by Exactech is a distributor in the Netherlands.  The translations on the outer labels include the word \" long\"  in all languages other than the original English version.  The implants are standard length fracture stems.",
      "recall_initiation_date": "20191203",
      "center_classification_date": "20200128",
      "termination_date": "20221011",
      "report_date": "20200205",
      "code_info": "All lot numbers in OUS distribution."
    }
  ]
}