{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bethlehem",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89784",
      "recalling_firm": "B Braun Medical Inc",
      "address_1": "824 12th Ave",
      "address_2": "N/A",
      "postal_code": "18018-3524",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0872-2022",
      "product_description": "Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis  Systems. Conductivity Sensor (BIC) - VERSION 2, Article Code Number 3456102A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S  (UDI 04046964285608), 710200U (UDI 04046964285615), 710200K (UDI 04046963686338), 710500L (UDI Not  applicable) and 710500K (UDI Not applicable).",
      "product_quantity": "80",
      "reason_for_recall": "Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.",
      "recall_initiation_date": "20220128",
      "center_classification_date": "20220414",
      "termination_date": "20240515",
      "report_date": "20220420",
      "code_info": "Lots 4/21 and 5/21"
    }
  ]
}